Genova Nano

Genova Nano

Genova Nano

Product catalog summary
Introduction
The Jenway Genova Nano spectrophotometer integrates a micro volume accessory with life science measurement modes, allowing for sample volumes as low as 0.5µl to be measured with high accuracy and speed. This reduces sample waste and eliminates the need for cuvettes, with easy cleaning between samples.

Specifications
The Genova Nano is specified for an absorbance accuracy of +/-2% at 260nm and a precision of < 0.005A between 0 and 1A at a 0.5mm path length. This application note evaluates these specifications across eight units and examines concentration reproducibility in dsDNA mode.

Materials and Methods
Calibration was performed using a certified standard solution. Absorbance accuracy was tested at 260nm and 330nm, with measurements taken at 0.5mm and 0.2mm path lengths. Precision was assessed using green food coloring, measuring at 430nm and 520nm. Concentration repeatability was tested using Calf thymus DNA, with measurements at multiple wavelengths.

Results
All units met the absorbance accuracy specification within +/-2% of expected values. Precision tests showed a range of ≤0.005Abs for 10 consecutive readings. Concentration measurements demonstrated a variation of less than +/-2µg/ml in DNA samples, highlighting the instrument's reproducibility.

Conclusions
The Genova Nano provides accurate and precise absorbance measurements, offering significant cost savings in time, sample, and consumables by allowing direct sample application onto the read head.
See more

Catalog excerpts

Genova Nano-1

GENOVA NANO Accuracy and precision of the Genova Nano Introduction The Jenway Genova Nano spectrophotometer combines a micro volume accessory with the dedicated life science measurement modes of the Genova Plus as well as those of a standard spectrophotometer. The Genova Nano is able to measure sample volumes as low as 0.5µl with a high degree of accuracy, reproducibility and speed. This ability conserves precious samples, reduces the need for dilution and eliminates the requirement for cuvettes. Cleaning is quick and simple: wiping the read heads with a lintfree cloth removes all trace of the sample, allowing faster change over between samples and therefore increasing sample throughput. Fig. 1: Applying samples to the read head of the micro volume accessory. and precision is a measurement of repeatability or reproducibility. The Genova Nano is specified to an absorbance accuracy of +/-2% at 260nm and a precision of < 0.005A between 0 and 1A (at 0.5mm path length). This application note demonstrates this specification on eight individual units and also shows concentration reproducibility when measuring in dsDNA mode. Materials and Methods Each of the Genova Nano units was calibrated using the certified calibration standard solution provided following the instructions detailed in the user manual. To determine absorbance accuracy, the unit was set up in multiwavelength mode to measure at 260nm and 330nm. The unit was zeroed using 2µl of the calibration blank solution then 2µl aliquots of the calibration standard solution were measured. Following each measurement, the sample was wiped off the read head using a lint-free wipe. This test was performed at both at 0.5mm and 0.2mm path lengths and on each individual unit. 10 successive readings were used for data analysis. Fig. 2: A droplet of sample is shown on the Genova Nano read head. In order to measure absorbance precision or repeatability, a sample of green food colouring (Supercook, Leeds, UK) was used. This contains a mix of tartrazine (E102) and Green S (E142). The absorbance spectrum [email protected] www.jenway.com Tel: +44 (0)1785 810433 An important requirement for measurement of micro volumes is accuracy and precision. Accuracy is defined as how close the measured parameter is to a defined value

 Open the catalog to page 1
Genova Nano-2

of the food colouring shows a peak at 430nm and a trough at 520nm (Fig. 3), therefore these wavelengths were chosen for measurement. The food colouring was diluted with water to give an absorbance between 0 and 1 in a path length of 0.5mm. measurement, the sampls was wiped off the read head using a lint-free wipe. This test was performed at 0.5mm path length on each individual unit. 10 successive readings were used for data analysis. Fig. 3: Visible spectrum of green food colouring measured using the microvolume accessory at 0.5mm path length showing a peak at 430nm and trough at 520nm. In a...

 Open the catalog to page 2
Genova Nano-3

The expected absorbance values for the calibration standard solution were 0.490 at 0.5mm path length and 0.196 at 0.2mm path length, according to the certificate supplied with the reagents. All units were well within the specification set at +/-2% of the expected values at both 0.5 and 0.2mm path lengths. In addition there was less than 2% variation between instruments at 0.5mm path length. further the reproducibility of the Jenway Genova Nano. To determine repeatability of measurement, a series of food colouring samples were measured and the maximum and minimum reading of the range determined....

 Open the catalog to page 3
Genova Nano-4

Bibby Scientific [email protected] www.ienwav.com Tel: +44 (0)1785 810433

 Open the catalog to page 4

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.