Determination of the magnetic properties of electrical steel in all common quality grades, thicknesses and geometries and of other soft magnetic materials Quality control of strip steel, punched parts, stators, transformer or ring cores Measuring categories Specific hysteresis loss Maximum polarization Effective polarization Maximum field strength Fully digitized measuring unit Effective field strength Three coil connectors for different measuring requirements Module for the measurement of rings, ring strip cores, stacked punched parts etc. Remanence Coercive field strength Permeability Possibility to extend the modules to include Epstein measurement Specific apparent power User commands and control via PC under Windows Hysteresis display MPG Expert software for measurement, display and integration into QM systems Form factor Measuring Technology for Soft Magnetic Materials J(H) values for graphic display
Open the catalog to page 1Measuring Instrument C 510 Operating principle The sample is exposed to a defined magnetic field within a measuring coil. A magnetic flux is created inside the material. The required current is supplied by a power amplifier. The current is measured by means of a temperature-stable and induction-free precision resistor (shunt) or by field coils. Polarization is determined by measuring the induced voltage, and then conversion and integration by means of a 16 bit processor. Parallel recording of H and J sizes in separate recording systems guarantees absolutely parallel measurement. Errors in measurement...