1. Catalogs
  2. BRAND
  3. PLT unit_2014

PLT unit_2014

PLT unit_2014

PLT unit_2014

Product catalog summary
Introduction
The document discusses the BRAND Pipette Leak Testing Unit (PLT unit), designed to improve the reliability of air-displacement pipettes by detecting leaks that cause inaccuracies. Leaks often result from damage to seals, pistons, or tip cones and are not easily visible.

Specifications
The PLT unit is pre-programmed with limit values for single- and multichannel pipettes ranging from 1 µl to 10 ml. It provides test results in seconds and can test pipettes with or without tips. The unit is patent pending.

Testing Procedures
Regular checks are necessary as leaks can occur anytime, affecting pipette accuracy. The PLT unit complements regular gravimetric testing by providing daily checks. It uses differential pressure measurement to detect leaks, calculating the leak rate (QL) based on pressure and volume changes over time.

Dynamic and Static Tests
The dynamic test involves moving the pipette piston to identify leaks caused by piston defects. The static test checks for leaks without piston movement, providing a general leak detection.

Limit Values and Standards
Testing adheres to ISO 8655-2 standards, with limit values indicating potential volume loss. The PLT unit's calculations consider factors like dead volume and pipette type, ensuring compliance with tolerance limits.

Ordering Information
The document provides ordering details for various components, including single- and multichannel adapters, PE filters, and a universal AC adapter.

Conclusion
The PLT unit enhances process reliability by detecting even minor leaks, ensuring pipettes remain within specified tolerances between calibrations.
See more

Catalog excerpts

PLT unit_2014-1

Improve process reliability with functional validation of air-displacement pipettes! Pipette Leak Testing Unit F I R S T The most frequent cause of inaccuracy in air-displacement pipettes is leakage. This arises from damage either to the seals, pistons, or tip cones. Often not detectable by the naked eye, leaks lead to significant volume errors. The BRAND pipette leak tester (PLT unit) for air-displacement pipettes finds even the smallest leaks within seconds. I Limit values for the commercially available singleand multichannel pipettes in the volume range 1 µl to 10 ml are pre-programmed. I Testing with and without tips I Test results in seconds I Patent pending

 Open the catalog to page 1
PLT unit_2014-2

Pipette Leak Testing Unit (PLT unit) Leak rates and their detection According to monitoring of measuring instruments, air-displacement pipettes must be checked at regular intervals and the results must be compared with the ISO 8655-2 error limits. However, a calibration certificate only reflects the results at the time of testing. The time between these calibrations is crucial, since leaks can occur at any time. Well over 80% of pipettes sent in for repair have leaks and are outside their volume tolerances, even if they don’t drip. The leak rate is a measure of the quantity of material that flows...

 Open the catalog to page 2
PLT unit_2014-3

Main Menu A wide variety of submenus can be selected from the main menu, e.g., pipette type, volume range, self-test, and settings (language, shut-off time, pressure units, etc.) Main Menu Pipette Type Self Test Switch OFF Settings... Auswahl select zurück back Testing with and without tip To check the overall pipette system, the test is conducted with mounted, unused tip. When a leak has been identified, the test can be repeated without a tip to determine whether the leak arises from the tip cone/tip coupling region. Dynamic or static test? The dynamic test can rapidly determine whether a defective...

 Open the catalog to page 3
PLT unit_2014-4

Ordering Information PLT unit (Pipette Leak Testing Unit) Including one 1-channel pipette adapter* each for testing single-channel air-displacement pipettes with tip (mounted) and without tip, 2 plugs, 3 replacement PE filters for the pipette adapters, universal AC adapter, quality certificate and operating manual. Pack of 1. Cat. No. * 4-channel pipette adapter optional 1- channel pipette adapter for testing of single-channel air-displacement pipettes with tip mounted, including 1 plug. Pack of 1. Cat. No. 4- channel pipette adapter for testing of single-channel air-displacement pipettes without...

 Open the catalog to page 4

Archived catalogs

  1. PLT unit

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.