Blower C65MS1

Blower C65MS1

Blower C65MS1

Product catalog summary
Specifications:
The C65MS1-24PR-04-Lx is a miniature centrifugal blower designed with medical-grade thermoplastics. It features a three-phase brushless motor without a Hall sensor and an integrated driver. The blower operates within a temperature range of -20 to +50°C (-4 to +122°F) and weighs 235 grams (8.29 oz). It is available in a 12 V version upon request.

Technical Data:
The blower operates at a nominal voltage of 24 VDC, with a voltage range of 10 to 28 VDC. Performance metrics include zero flow pressure of 4,570 Pa and 7,200 Pa for different models, with corresponding current and speed values. The working point pressure is 3,390 Pa and 5,050 Pa, with flow rates of 220 lpm and 255 lpm, respectively. The blower's acoustic emission level is below 52.5 dB(A), and it has a lifespan of over 20,000 hours at 25°C.

Electrical Connections:
The blower uses AWG22 wires for power supply, ground, speed control voltage (VSP), and frequency generator (FG). The speed is calculated as 20 times the FG value. Speed control is achieved through a voltage range of 0 to 3.3 V, with specific voltage thresholds for stopping and full speed.

Power Supply Requirements:
The blower requires a voltage of 10 to 28 VDC with a ripple voltage of less than 5% and a current of 3.5 A. It includes over-current protection, limiting the current below 3.5 A.

Blower Limits and Warnings:
The blower should not be hot-swapped, and the maximum temperature on the heat sink should not exceed 75°C. Continuous speed should not exceed 52,000 rpm, and blocking the impeller for more than 20 seconds may cause damage. A minimum airflow is necessary at voltages higher than nominal to prevent overheating.

Manufacturing Standards:
The blowers are manufactured by an ISO9001-2015 certified manufacturer.

Typical Hookup:
The document provides guidance on measuring generator frequency to calculate blower speed, using external power supplies and potentiometers for speed control, and ensuring proper connections for optimal performance.
See more

Catalog excerpts

Blower C65MS1-1

E-mail: [email protected], URL: www.boreasa.com Miniature Centrifugal Blower C65MS1-24PR-04-Lx 65.5 x 63 x 47 mm (2.58 x 2.48 x 1.85 in) ∗ Housing / Impeller Materials: Medical Grade Thermoplastics ∗ Lead Wire: 285 mm (AWG22 & AWG22) ∗ Weight: 235 gr. (8.29 oz) ∗ Operational Temperature: -20 to +50°C / -4 to +122°F ∗ Three Phase Brushless Motor w/o Hall Sensor ∗ Note: This blower has an integrated driver ∗ Available in 12 V on demand Note: - Performances are dependent on the driver used and the test conditions. The data in this spec. sheet was obtained using Boreasa test chamber at 20°C (±2°C) and sea level. It may be different than data measured in other conditions. - When used as a vacuum, the blower performance might be less than shown herein, depending on the operating point. - Specifications are subject to change without notice. - Boreasa blowers have been designed and are manufactured to meet the medical device industry requirements. However, it is not possible to test blowers in all conceivable customer application configurations. Consequently, to be used in life support equipment, blowers must be life tested and validated by customers, in their actual application conditions, in a statistically meaningful quantity.

 Open the catalog to page 1
Blower C65MS1-2

E-mail: [email protected], URL: www.boreasa.com #1 At given working points and cycle. #2 Measured at distance of 1 meter from inlet, with open inlet, outlet connected to breathing tube and 4 mm orifice in sound chamber at 10 cmH2O. N ote : - Working point is an example close to 50% of max. flow. - Tolerances based on specified speed data according to ISO 13348, grade 4: pressure ±10%, power ±16%, speed ±5% and flow rate ±5%.

 Open the catalog to page 2
Blower C65MS1-3

E-mail: [email protected], URL: www.boreasa.com Electrical Connections Speed Output Speed Calculation VSP (Speed Control Voltage) 22 Input Voltage Blower speed [rpm] = 20 * FG FG = Frequency Generator 0 - 1.0 V O Blower stopped Power Supply Requirements VSP not connected Vsp to GND Voltage Ripple Voltage Current Blower speed 100% Blower stopped 10 ~ 28 VDC < 5% Blower Limits and Warning Manufacturing System * Max. allowed temp. on heat-sink: 75°C * Blocking the impeller more than 20 seconds may result in damages or destruction of the blower. * A minimum air flow is required at VCc higher than...

 Open the catalog to page 3
Blower C65MS1-4

E-mail: [email protected], URL: www.boreasa.com Typical Hookup 1. Measuring the Generator Frequency to calculate the blower speed. 2. Using an external +3.3 V power supply voltage and a potentiometer to control the blower speed. 3. Using a potentiometer to control the blower speed. 4. Using external analog control signal (0 - 3.3 V) to control the blower speed (note: Do not use DC power supply voltage).

 Open the catalog to page 4