Blower C65MS1-24PS-01-L5

Blower C65MS1-24PS-01-L5

Blower C65MS1-24PS-01-L5

Product catalog summary
Product Overview:
The C65MS1-24PR-0B-L5 is a miniature centrifugal blower designed for medical applications. It features a three-phase brushless motor with a Hall sensor and requires a driver for operation. The blower is constructed from medical-grade thermoplastics and has a compact size of 65.5 x 63 x 47 mm, weighing 235 grams.
Specifications:
  • Nominal Voltage: 24 VDC, with a range of 10 to 28 VDC.
  • Operational Temperature: -20 to +50°C (-4 to +122°F).
  • Storage Temperature: -20 to +80°C (-4 to +176°F).
  • Relative Humidity: 10 to 85% (non-condensing).
  • Acoustic Emission Level: < 52.5 dB(A).
  • Maximum Continuous Speed: 60,000 rpm.
Performance Data:
  • Zero Flow Pressure: 8,220 Pa at nominal voltage.
  • Working Point Pressure: 5,770 Pa with a flow rate of 286 lpm.
  • Free Flow Rate: 510 lpm.
  • Power Consumption: 51.84 W at working point.
Electrical Connections:
  • Motor windings and Hall sensors are connected via specific pins with designated wire colors (e.g., Motorwinding U - Brown, Hall sensor A - Blue).
Warnings and Limitations:
  • No hot-swapping is allowed.
  • Maximum temperature on NTC: 85°C.
  • Blocking the impeller for more than 20 seconds can cause damage.
  • A minimum airflow is required at voltages higher than nominal to prevent overheating.
Manufacturing and Quality Assurance:
The blowers are manufactured by an ISO9001-2015 certified manufacturer, ensuring compliance with industry standards. The internal temperature of the motor is monitored via a 10 KΩ NTC thermistor.
See more

Catalog excerpts

Blower C65MS1-24PS-01-L5-1

E-mail: [email protected], URL: www.boreasa.com Miniature Centrifugal Blower * Housing / Impeller Materials: Medical Grade Thermoplastics * Three Phase Brushless Motor w/ Hall Sensor * Standard Motor Internal Temperature Signal * Note: This blower uses a brushless motor and needs a driver to operate Note: - Performances are dependent on the driver used and the test conditions. The data in this spec. sheet was obtained using Boreasa test chamber at 20°C (±2°C) and sea level, and using a Boreasa D30/5/4Q-E2 driver. It may be different than data measured in other conditions or with other drivers. - When used as a vacuum, the blower performance might be less than shown herein, depending on the operating point. - Specifications are subject to change without notice. - Boreasa blowers have been designed and are manufactured to meet the medical device industry requirements. However, it is not possible to test blowers in all conceivable customer application configurations. Consequently, to be used in life support equipment, blowers must be life tested and validated by customers, in their actual application conditions, in a statistically meaningful quantity.

 Open the catalog to page 1
Blower C65MS1-24PS-01-L5-2

E-mail: [email protected], URL: www.boreasa.com #1 At given working points and cycle. #2 Measured at distance of 1 meter from inlet, with open inlet, outlet connected to breathing tube and 4 mm orifice in sound chamber at 10 cmH2O. N ote : - Working point is an example close to 50% of max. flow. - The phase inductance and impedance are measured by LCR at 1 kHz, including the connector. - Tolerances based on specified speed data according to ISO 13348, grade 4: pressure ±10%, power ±16%, speed ±5% and flow rate ±5%.

 Open the catalog to page 2
Blower C65MS1-24PS-01-L5-3

E-mail: [email protected], URL: www.boreasa.com Electrical Connections Rear side view shown (Mating connector side) Description Motorwinding U Motorwinding V Motorwinding W Hall sensor A Hall sensor B Hall sensor C VHall 5 ... 24 VDC GND NTC NTC Wire Brown Blower Limits and Warning Manufacturing System Motor Internal Temperature Signal (IT) * Blocking the impeller more than 20 seconds may result in damages or destruction of the blower. * A minimum air flow is required at VCc higher than Nominal Voltage to avoid the air in the blower to overheat by compressing and mixing. Blowers manufactured...

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.