Overview
The Bionetics Corporation offers Rheotherm duct flow meters designed for measuring mass flow in air and non-condensing gases, including corrosive and explosive types. These meters are suitable for industrial environments due to their stainless steel construction and lack of moving parts, ensuring durability and minimal maintenance.

Specifications
  • Output Units: Mass flow, standard volumetric flow, or velocity.
  • Process Connection: Options include NPT, flange, bore-through compression, or sanitary fittings.
  • Turndown Ratio: Standard 10:1, with options up to 200:1.
  • Wetted Surface Materials: 316 SS, Hastelloy C-276®, Monel®.
  • Output: 4-20 mA, 0-10 or 0-5 Vdc, frequency, open collector, 1 or 2 SPDT relays.
  • Input Power: 24 Vdc (9-36 Vdc) or 85-250 Vac for remote electronics.
  • Enclosure: NEMA 4X, NEMA 7.

Features
  • Thermal Mass Method: Utilizes two RTDs to measure temperature differential, providing accurate flow signals.
  • Installation: Can be installed in horizontal or vertical pipes, with electronics optionally mounted remotely up to 200 feet.
  • Rheovec™ Sensor Heads: Equipped with a honeycomb structure to eliminate off-axis errors, ensuring accurate mass flow rate measurement.

Applications
  • Suitable for air ducts and process gas pipes with line sizes from 1” to 36”.
  • Flow Rate Range: 50 to 25,000 ft/min.
  • Temperature Limits: Standard 0 to 212°F, optional -40 to 500°F.
  • Pressure Limits: Varies with process connection rating.
  • Repeatability: ±0.5% of reading.
  • Calibration Accuracy: ±2% of reading typical.

Options for Hazardous Locations
FM approved options are available for hazardous locations, with specific classes, divisions, and groups detailed in the product summary.

Contact Information
For further assistance, contact Bionetics at 888-LOW-FLOW (569-3569) or visit their website at www.bioneticsflow.com.
See more

Catalog excerpts

Rheotherm Duct Flow Meters-1

The Bionetics Corporation: CESG 770 Brooksedge Plaza Dr. Westerville, OH 43081 Rheotherm mass flow sensing insertion probes are available for air and most other non-condensing gases, including corrosive and explosive gases, are easy to install and provide long-term service with little or no maintenance. Common uses include vent air, stack gas, and digester gas. The output units can be in mass flow, standard volumetric flow, or velocity. Options for hazardous (classified) locations are available. The stainless steel construction and lack of moving parts make it ideal for rugged industrial environments. • Accurate mass flow No moving parts Wide rangeability • Easy installation Rheotherm duct flowmeter electronics can be installed remotely from sensors Thermal Mass Method Our insertion flow sensors have two stainless steel flow sensing tips welded to the probe shaft. One tip has a heated RTD in it and the other an unheated RTD. The temperature differential between the two RTDs provides the primary flow signal: at high flow rates the differential is lower as flow removes more heat, and at low flow rates the differential is higher, as flow removes less heat. Nothing touches the gas but the metal probe and tips. Installed through the pipe or duct wall, connections can be made with NPT, flange, bore-through compression, or sanitary fittings. The probes can be installed in horizontal or vertical pipes. Electronics with optional display may be integral to the sensor or remotely mounted up to 200 feet from the sensor. Typical calibration ranges are 10:1 (options up to 200:1). Over-ranging cannot damage the sensors.

 Open the catalog to page 1
Rheotherm Duct Flow Meters-2

Duct Flow Meter Features and Options What if there is little or no straight run available? Rheovec™ Sensor Heads The turbulence and swirling of air and gas in short-run and curved ducts create accuracy problems for ordinary flow sensors due to fluctuating velocities. The Rheovec™ sensor head, with its unique honeycomb structure around the probe tips, eliminates off-axis error, enabling highly repeatable true mass flow rate measurement. See Product Summary Rheovec for further information. The Bionetics Corporation: CESG 770 Brooksedge Plaza Dr. Westerville, OH 43081 • Air ducts and process gas...