• Connect to DART®-MS and Visualize Hidden Compounds •Temperature Gradient System Visualizes Compounds in Real Time • Eliminates Sample Preparation Steps ion Rocket DART® -MS System 2. Linear temperalure gradients from ambient to 600°C. Vaporized sample rides into the DART® gas stream ionRocket generates a temperature gradient from ambient up to as high as 600°C over several minutes. This allows compounds in your samples to be sublimated, vaporized, or pyrolized according to their volatility, and then introduced into the DART® gas stream.
Open the catalog to page 2Purpose Analysis of polyethylene terephthalate (PET) used in PET bottles using a time-of-flight mass spectrometer as a detector MS measurement of small pieces of PET bottles (A, B) using the ionRocket while raising the temperature from room temperature to 500°C(A; hard PET B; soft PET) Using Spectra Scope and Data Analysis software, PET repeating structures were extracted from mass spectra, and A and B PET were compared. Fig.l shows the analysis results (bubble plot) of the Polymer engine. It was found that A had more PET repeating structures (Ci0H8O4) than B. Figure 2 shows the TIC and mass...
Open the catalog to page 3Main body Heater control software Product name https://www.biochromato.com/ E-mail:[email protected]