Battery test chambers
22Pages

{{requestButtons}}

Catalog excerpts

Battery test chambers - 1

Battery test chambers LIT MK series Safety packages for aging, performance, and stress tests

Open the catalog to page 1
Battery test chambers - 2

EUCAR hazard levels & standards What classifications apply to handling energy storage devices? The EUCAR hazard levels are used to assess the level of danger associated with handling batteries. They have been defined by EUCAR (the European Council for Automotive R&D) by classifying the hazards presented to batteries and describing the consequences of them. BINDER offers various packages with additional safety measures based on these effects on batteries. Passive protection activated No defect; no leakage; no venting, fire, or flame; no rupture; no explosion; no exothermic reaction or...

Open the catalog to page 2
Battery test chambers - 3

Battery test chambers Battery test chambers Safety packages A good decision: You are in safe hands with BINDER Battery test chambers from BINDER are suitable for tests performed on lithium-ion cells and modules. Handling lithium-ion batteries can present a variety of potential hazards. System operators need to evaluate the level of risk and counteract it using an appropriate safety concept. The LIT MK series of temperature test chambers from BINDER provides a controllable temperature range of -40°C to +110°C. The units come with extensive safety equipment as standard so that tests on...

Open the catalog to page 3
Battery test chambers - 4

Product range | Battery test chambers PRODUCT RANGE FOR BATTERY TEST CHAMBERS Features General data Temperature range [°C] Ethernet interface Forced convection Controller and timer functions Controller display Ramp function Programming option Class 1 temperature safety device Class 2 independent adjustable temperature safety device Temperature alarm, acoustic Temperature alarm, visual

Open the catalog to page 4
Battery test chambers - 5

Battery test chambers | Advantages at a glance EVEN TEMPERATURE DISTRIBUTION OCCUPATIONAL SAFETY Thanks to BINDER APT.line™ preheating chamber technology. Long-term operational stability even under full load. Extensive safety equipment allows tests to be performed safely on lithium-ion batteries. PROGRAMMING AND DATA COLLECTION EASY ASSEMBLY With the program controller featuring a color display and Multi Management Software APT-COM™ for remote programming through a PC. Thanks to large access area. The large heated viewing window with LED illumination provides a view of the test object.

Open the catalog to page 5
Battery test chambers - 6

LIT MK series | Battery test chambers LIT MK series | Battery test chambers with safety equipment for rapid temperature changes Battery test chambers from BINDER are ideal for aging, performance, and stress testing in quality assurance and battery production. To ensure maximum safety, all the chambers in the LIT MK series are fitted with detection and fire suppression equipment as standard. IMPORTANT FEATURES • Temperature range: -40°C to +110°C • Class 2 independent temperature safety device when temperature is set to +120°C • Reversible pressure relief flap made of stainless steel...

Open the catalog to page 6
Battery test chambers - 7

Battery test chambers | LIT MK series OPTIONAL EQUIPMENT Electromechanical door-locking mechanism Electromechanical door-locking mechanism controlled in a time program or manually. Access port with silicone plug For introducing external measuring instruments. Large selection of different Notch-type access port in door Notch-type access port in the door, 100 x 35 mm, makes it easy to guide measuring diameters and positions (left or top). devices to the test object. You can access our installation, maintenance, calibration, validation, or warranty services by taking out a maintenance contract...

Open the catalog to page 7
Battery test chambers - 8

LIT MK series | Battery test chambers TECHNICAL DATA Description Net weight of the unit (empty) [kg] Permitted load of individual shelf [kg] Wall clearance, rear [mm] Wall clearance, side [mm] Temperature uniformity dependent on set value [± K] Temperature fluctuation dependent on set value [± K] Average heating-up rate acc. to IEC 60068-3-5 [K/min] Average cooling down rate acc. to IEC 60068-3-5 [K/min] Nominal power (400 V version) [kW] Nominal power (480 V version) [kW] Interior dimensions Housing dimensions not incl. fittings and connections Technical data: temperature Electrical data...

Open the catalog to page 8
Battery test chambers - 9

Battery test chambers | LIT MK series DIMENSIONS incl. fittings and connections [mm] LIT MK 240 ORDERING INFORMATION FOR OPTIONS Designation For temperature values (output not adjustable) Dry-air purge, connection To enable connection to the on-site compressed air network Controlled via control contact in time program or manually Measurement in center of chamber at specified temperature Temperature measurement incl. certificate, 15–18 measuring points at specified test temperature Temperature measurement incl. certificate, 27 measuring points at specified test temperature Temperature...

Open the catalog to page 9
Battery test chambers - 10

LIT MK series | Battery test chambers Additional serial interface can be used parallel to Ethernet, for Multi Management Software APT-COM™ Additional flexible Pt 100, interior, for displaying the temperature on the unit display Door lock Lockable door handle Reinforced chamber base For increasing the load capacity of the interior base ORDERING INFORMATION FOR ACCESSORIES Designation APT-COM™ 4 PROFESSIONAL Edition Description For working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11. Version 4,...

Open the catalog to page 10
Battery test chambers - 11

Battery test chambers | LIT MK series

Open the catalog to page 11
Battery test chambers - 12

BINDER INDIVIDUAL Safety packages Package A | Package P | Package P Plus | Package S BINDER offers test chambers for aging, performance, and stress tests. Our BINDER INDIVIDUAL department can equip the following models with package A, P, P Plus, and S:

Open the catalog to page 12
Battery test chambers - 13

Battery test chambers | Package A Package A | For aging tests This safety package is ideal for tests that present a low potential hazard level and is exactly in line with EUCAR hazard level 2. AGING TESTS: These involve testing at a certain temperature without the battery load cycle. They are performed within a safe temperature range for the battery. MK 56 model with package A SOLUTION • Class 2 independent temperature safety device when temperature is set to 120°C • Temperature range limited to 110°C on the controller 1 TEMPERATURE RANGE LIMITATION Class 2 independent temperature safety...

Open the catalog to page 13

All BINDER catalogs and technical brochures