1. Catalogs
  2. Bigneat Containment Technology
  3. UV Sterilisation and PCR Workstation

UV Sterilisation and PCR Workstation

UV Sterilisation and PCR Workstation

UV Sterilisation and PCR Workstation

Product catalog summary
Introduction
When conducting PCR reactions, it is crucial to prevent contamination from DNA. This requires thorough decontamination of the cabinet, reagents, and equipment. UV irradiation is an effective sterilizer, breaking down DNA sequences to prevent replication. UV exposure times range from 5 to 30 minutes. The Bigneat UV Sterilisation Cabinet and PCR Workstation utilize UV effectively, with the latter also providing sterile air and positive pressure to prevent contamination.

Materials of Construction
The fan housing and shelving are made of epoxy-coated steel, the support structure is anodised aluminium, windows are 10 mm thick acrylic, and the base tray is PVC.

Filtration
The workstation uses a HEPA filter system with a pre-filter efficiency of 90% to EFV grade EU7 and a HEPA filter efficiency of 99.997% for particulates as small as 0.3µm.

Technical Information & Specification
The cabinets require a 230V/50 Hz electrical supply. Options for a stand or mobile trolley are available. The PCR® process is patented by Hoffmann-La Roche Inc.

Features of UV Sterilisation Cabinet
  • Constructed from high-quality, corrosion-resistant materials that are easy to clean.
  • Equipped with powerful UV tube lights and optical reflectors for uniform UV distribution.
  • Acrylic windows provide clear vision and block UV transmission.
  • Control system allows UV exposure timing from 5 to 30 minutes.
  • Door sensors turn off UV lamps when the door is opened for safety.
  • Acrylic protects against beta emissions from 32P labelled compounds.
  • Includes corner shelves and door flaps for easy access.
  • 15W white lights illuminate the interior, operated by an external switch.
  • Side panels have ports for cable access.

Features of PCR Workstation
  • Includes all features of the UV Sterilisation Cabinet.
  • Filtration system delivers air to Class 100 quality.
  • Auto speed increase on door opening maintains air velocity at 0.45 m/s.
  • Electronic filter saturation indicator.
  • Fan noise level is below 50 dB(a).
See more