Overview: The DTMc300 is a double monochromator designed for applications requiring superior scattered light performance. It combines two TMc300 monochromators in tandem, offering flexibility and a wide spectral range within a single scan. It is compatible with Bentham's range of sources, detectors, and accessories.
Key Features:- Double monochromator with two 300mm focal length monochromators in Czerny-Turner configuration for additive or subtractive dispersion.
- Each monochromator can hold up to three diffraction gratings, allowing single-scan measurements over a spectral range of 200nm-50µm.
- Supports up to two entrance ports and one exit port in single mode, and up to two exits in double mode, with automated mirror-based port selection.
- Fully automated via USB 2.0 interface.
Optical Layout:- Pair of symmetrical Czerny-Turner monochromators for additive or subtractive dispersion.
- Total focal length of 600mm (additive), with individual units at 300mm.
- Triple grating turret with 1-3 gratings, grating size of 68 x 84mm, and aperture ratio of f/4.1.
Optical Performance:- Spectral range from 200nm to 50µm.
- Linear dispersion of 1.35nm/mm (additive).
- Wavelength accuracy of ±0.1nm and reproducibility of ±0.025nm.
- Resolution of 0.15/0.05nm (full/reduced slit height).
- Stray light rejection at 2.5 FWHM of 10-8.
- Wavelength acquisition speed of 2000nm/s.
Automation and Control:- Controlled via USB with BenWin+/SDK software.
Physical Specifications:- Dimensions: 634W x 360D x 211H mm.
- Optical height: 115mm.
- Weight: 28 kg.
- Power supply: 110/220V 50/60Hz.
Options:- Diffraction gratings and order sorting filters available separately.
- Filter wheel options: 252-6 or 252-8 (includes shutter).
Contact Information: Bentham Instruments Ltd, 2 Boulton Road, Reading, Berkshire, RG2 0NH, United Kingdom. Phone: +44 (0) 118 975 1355, Email:
[email protected], Website: www.bentham.co.uk