video corpo

ISOMETER® isoMED427P
6Pages

{{requestButtons}}

Catalog excerpts

ISOMETER® isoMED427P - 1

ISOMETER® isoMED427P Insulation monitoring device with integrated load and temperature monitoring and locating current injector for insulation fault location systems for medical IT systems

Open the catalog to page 1
ISOMETER® isoMED427P - 2

Insulation monitoring device with integrated load and temperature monitoring and locating current injector for insulation fault location systems for medical IT systems Product description The ISOMETER® isoMED427P monitors the insulation resistance of unearthed AC circuits (medical “IT systems”). At the same time, the load current and temperature of the IT system transformer is monitored. In combination with EDS series insulation fault locators and the appropriate measuring current transformers, the isoMED427P is designed to set up the respective equipment for insulation fault location....

Open the catalog to page 2
ISOMETER® isoMED427P - 3

Alarm messages LEDs Operating elements isoMED427P “ON” Operation System fault* Insulation fault Overcurrent Overtemperature * Detailed alarm information on LCD Test function/connection monitoring A self test is carried out once supply voltage is fed and later at hourly intervals. During the self test, the internal device functions, the connections to PE (E/KE) and the connections to the current transformer are monitored for interrruption and short-circuit. In the event of a fault, the alarm relay K1 switches and the LEDs ON/AL1/AL2 flash. The respective error code appears on the LC display....

Open the catalog to page 3
ISOMETER® isoMED427P - 4

Wiring diagram isoMED427P with CMS460-D4 Application example Technical room SEB = Supplementary equipotential bonding

Open the catalog to page 4
ISOMETER® isoMED427P - 5

Technical data Insulation coordination acc. to IEC 60664-1/IEC 60664-3 Rated insulation voltage 250 V Rated impulse voltage/pollution degree 4 kV/3 Protective separation (reinforced insulation) between (L1, L2, E, KE, T1, T2, A, B, Z, Z/k, l) - (11, 12, 14) Voltage test acc. to IEC 61010-1 2.21 kV Voltage supply Supply voltage US = Un Power consumption < 6,5 VA Interface Interface/protocol Baud rate Cable length Cable (twisted in pairs, one end of shield connected to PE) Terminating resistor Device address, BMS bus IT system being monitored acc. to IEC 60364-7-710 Nominal...

Open the catalog to page 5
ISOMETER® isoMED427P - 6

Ordering information BENDER The Power in Electrical Safety6 o3 X _Q E u ai ~o c ai CQ © x Bender GmbH & Co. KG P.O. Box 1161 • 35301 Gruenberg • Germany Londorfer Strasse 65 • 35305 Gruenberg • Germany Tel.: +49 6401 807-0 • Fax: +49 6401 807-259 E-Mail: info@bender.de • www.bender.de g

Open the catalog to page 6

All BENDER catalogs and technical brochures

  1. CTBC17

    2 Pages

  2. POWERSCOUT®

    8 Pages

  3. Energy meter

    6 Pages

  4. RDC121

    7 Pages

  5. RCMB123

    7 Pages

  6. RCMB 121-2

    7 Pages

  7. RCMB121-1

    7 Pages

  8. RDC104-4

    6 Pages

  9. 300ST

    4 Pages

  10. 400ST

    4 Pages

  11. USC710D4

    6 Pages

  12. DI-1DL

    4 Pages

  13. DI-2

    4 Pages