TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials
6Pages

{{requestButtons}}

Catalog excerpts

TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 1

Features » A-scan waveform can be displayed for echo analysis and measurement of complex workpiece; » Compatible with many types of transducers, single and dual element transducers are suitable; » With a variety of measurement methods, users can set blanks to shield aftershocks or clutter; » Echo-echo measures the true metal thickness with ignoring the thickness of coating layer. Thru-coat technology measures metal and nonmetallic coating thickness. » Signal auto-amplification function (centered display of the detected echo); » The resolution is 0.001, 0.01, 0.1 mm optional in any mode (or 0.0001, 0.001, 0.01 inch optional) » Gain adjustment range 0-99dB; » Adjustable voltage variable pulse width square wave pulse generator; » Single value B-scan display function; » Users can turn on fast measurement mode up to 20 times per second; » Alarm function, user can set the upper and lower limits of the alarm; » Differential, maximum and minimum display mode; » Internal data storage, data can be output to a removable MicroSD memory card. Can store up to 500,000 measured values and waveforms;

Open the catalog to page 1
TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 2

Standard Configuration Main unit 1 5MHz dual element wideband transducer 1 Couplants 1 AA battery 1 User manual 1 Optional Configuration Standard block 1 MHz single element contact transducer TSTU32 2MHz Dual element transducer 5MHz Single element contact transducer 15MHz single element with delay lines transducer ZW5P high temperature transducer

Open the catalog to page 2
TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 3

Resolution: Sound velocity adjustment range: Display screen: Pulse Generator: Emission voltage: Emission pulse width: Gain range: Frequency range: Measurement rate: Transducer settings: Data Storage: Working environment temperature: POWER Power: Power consumption: 0.001mm or 0.01 or 0.1mm optional 508 m/s~18699m/s Color TFT LCD, 320x240 pixels Adjustable Square Wave Pulse Generator 60V, 110V, 150V, 200V optional varies with transducer frequency 0-99dB, 1dB step 0.5 Mhz~20Mhz standard (4Hz), fast (20Hz) 10 sets of fixed transducer setting and 22 sets of custom transducer setting 500 data...

Open the catalog to page 3
TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 4

♦ The standard echo detection mode measures the thickness based on the time interval between the excitation pulse and the first back-wall echo. User can measure uncoated materials in this mode. Measure in standard echo detection mode ♦ Automatic echo-echo detection mode allows thickness measurement of materials with paint or coating because the time interval between two successive back-wall echoes eliminate paint or coating thickness Measurement in manual E-E detection mode Measurement in automatic E-E detection mode ♦ Paint thickness measurement can simultaneously display layer thickness...

Open the catalog to page 4
TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 5

♦ The instrument includes three detection modes (Mode 1, Mode 2, and Mode 3): Mode 1: Measures the time interval between the main pulse signal and the first back-wall echo with direct contact transducer. Mode 2: Measure the time interval between the interface echo (or delay line echo) and the first back-wall echo with a delay line or immersion transducer. Mode 3: Measure the time interval between two successive back-wall echoes with a delay line or a immersion transducer. Measuring Mode

Open the catalog to page 5
TIME2190 Ultrasonic Thickness Gauge with A/B Scan for Complicated Materials - 6

The TIME®2190 gage B-scan feature convers live thickness reading into cross-sectional images drawn on the display. This standard feature is very helpful in viewing the changes in thickness measurements over a distance. The B-scan is activated as soon as the transducer makes contact with the surface of the material. ♦ The ZW5P transducer can be used to measure the thickness of steel with a surface temperature of up to 300 °C. For every 100 °C increase in steel temperature, the sound velocity of the material drops by about 1%, so the measured value should be corrected. Example: H0 is defined...

Open the catalog to page 6