Overview: The document provides technical specifications and details for the Ultrasonic Thickness Gauge TT130, which is designed for measuring the thickness of various materials, including both metallic and non-metallic substances.
Specifications:- Measuring Range: For steel, the range is 1.20mm to 225.0mm using a 5MHz transducer. For steel pipes, the minimum thickness is 3.0mm with a diameter of 20mm.
- Transducer Frequency: Standard frequency is 5 MHz with a 10mm diameter.
- Display Resolution: 0.01mm.
- Calibration: Integrated 4.0mm steel base plate.
- Measurement Accuracy: ±(1%H+0.1) mm.
- Measuring Units: mm/inch.
- Sound Velocity: Range from 1000m/s to 9999m/s.
- Display: 4-Digital LCD with backlight.
- Memory: Storage for 10 thickness readings.
- Surface Temperature Range: -10 to +60°C.
- Battery Indicator: Low battery voltage indicator.
- Power Supply: 2 AA batteries (1.5V each).
- Operation Time: 250 hours continuously.
- Dimensions: 126mm x 68mm x 23mm.
- Weight: Approximately 250g including batteries.
Probe Characteristics:- 5P 10: 5MHz, measuring range 1.2-225mm, diameter 10mm, standard for steel pipe size 20mm with 3mm thickness.
- SZ2.5P: 2.5MHz, measuring range 3-300mm, diameter 14mm, suitable for thick material/rough surfaces.
- 7P 6: 7MHz, measuring range 0.75-60mm, diameter 6mm, suitable for thin material.
Standard Delivery Includes:- Main unit.
- Standard 5 MHz transducers.
- Integrated steel calibration plate (4.0mm).
- AA batteries (1.5V).
- Ultrasonic couplant.
- TIME certificate.
- Instruction manual.
- Warranty card.
- Carrying case.
Optional Accessories:- Transducer 5P 10/90°.
- Transducer SZ2.5P.
- Transducer 7P 6.
See more