MET ONE 3411 Portable Airborne Particle Counter
Open the catalog to page 1MET ONE 3411 PORTABLE AIRBORNE PARTICLE COUNTER Does your particle counter contaminate your cleanroom? How much time do you spend recording and analyzing the data?
Open the catalog to page 2Introducing the MET ONE 3411 0.1 micron sensitivity portable airborne particle counter. Unlike many particle counters the MET ONE 3411 fan remains off when operating in a standard semiconductor cleanroom environment thus particle generation is virtually eliminated. Cleanroom friendly monitoring Reduced particle impact to cleanroom environment Unit-to-unit accuracy and reproducibility Assured through ISO 21501 compliance Flexible communications Wireless, Ethernet, Serial and USB Long continuous operation Dual hot-swappable batteries Intuitive touch-screen user interface Easy area, location,...
Open the catalog to page 3Number of Size Channels Particle Sizes Flow Rate Zero Count Coincidence Loss Counting Efficiency Light Source Pump Type 6 0.1, 0.2, 0.3, 0.5, 1.0, 5.0 1.0 CFM (28.3 LPM) ISO 21501-4 and JIS B 9921: 1 count or less in 5 minutes, 95% UCL 5% at 50,000 /ft 3 per ISO 21501-4 method 50% at 0.1 μm; 100% for particles > 0.15 μm per ISO 21501-4 and JIS B 9921 Helium Neon Laser, 5 mW Max Power at 632.8 nm Patented multi-lobe closed loop controlled rated for continuous use ¼ VGA Color TFT touch screen High speed thermal English, French, German, Italian, Spanish, Korean*, Japanese*, Chinese (Simplified...