Introduction
Fieldbus technology is increasingly prevalent in modern instrumentation, necessitating the calibration of fieldbus transmitters for accuracy, efficiency, and safety. Beamex introduces a solution for calibrating these instruments in hazardous areas using intrinsically safe calibrators.

Fieldbus Technology Overview
Fieldbus transmitters offer significant advantages over conventional transmitters, such as delivering extensive information via a two-way bus and reducing installation costs. The Foundation Fieldbus and Profibus standards dominate the market, providing benefits like better return on investment and improved system performance.

Calibration of Fieldbus Transmitters
Fieldbus transmitters require calibration similar to conventional transmitters, involving physical input and output measurement. Calibration cannot be performed remotely or solely with configuration software, necessitating practical solutions for hazardous environments.

Intrinsically Safe Calibration
Intrinsic safety is a technique for operating electronic equipment safely in explosive environments by ensuring low electrical and thermal energy levels. Intrinsically safe calibrators, or Ex calibrators, are essential in industries like oil refineries and chemical plants, offering safety and functionality.

Benefits and Technical Differences
Intrinsically safe calibrators provide the safest calibration method and maintain performance similar to regular devices. They feature conductive surfaces, specific usage constraints, and require charging in non-explosive areas. They are certified under ATEX and IECEx standards, ensuring safety and compliance.

Beamex MC5-IS Calibrator
The Beamex MC5-IS is a multifunctional, intrinsically safe calibrator certified for use in hazardous environments. It supports calibration for pressure, temperature, electrical, and frequency signals, and integrates with calibration software to enhance process efficiency and quality.

See more

Catalog excerpts

White Paper- The Safest Way to Calibrate Fieldbus Instruments-1

110680_Wo_KO_05_2011_Um_frz.qxp:100988_Wo_KO_05_2010_Um table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:7.00pt "Arial", sans-serif; } .font1 { font:17.00pt "Arial", sans-serif; } .font2 { font:26.00pt "Arial", sans-serif; } .font3 { font:32.00pt "Arial", sans-serif; } .font4 { font:33.00pt "Arial", sans-serif; } .font5 { font:4.50pt "Microsoft Sans Serif", sans-serif; } Beamex Calibration White Pape [email protected] The safest way to calibrate fieldbus instruments beamex WORLD-CLASS CALIBRATION SOLUTIONS

 Open the catalog to page 1
White Paper- The Safest Way to Calibrate Fieldbus Instruments-2

110680_Wo_KO_05_2011_Um_frz.qxp:100988_Wo_KO_05_2010_Um table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:8.00pt "Arial", sans-serif; } .font1 { font:9.00pt "Arial", sans-serif; } .font2 { font:11.00pt "Arial", sans-serif; } .font3 { font:13.00pt "Arial", sans-serif; } .font4 { font:28.00pt "Arial", sans-serif; } .font5 { font:9.00pt "Times New Roman", serif; } BEAMEX Calibration White Paper The safest way to calibrate fieldbus instruments Fieldbus is becoming more and more common in today's instrumentation. Fieldbus transmitters must also be calibrated just like...

 Open the catalog to page 2
White Paper- The Safest Way to Calibrate Fieldbus Instruments-3

110680_Wo_KO_05_2011_Um_frz.qxp:100988_Wo_KO_05_2010_Um table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:8.00pt "Arial", sans-serif; } .font1 { font:9.00pt "Arial", sans-serif; } .font2 { font:11.00pt "Arial", sans-serif; } .font3 { font:13.00pt "Arial", sans-serif; } .font4 { font:9.00pt "Times New Roman", serif; } BEAMEX Calibration White Paper word in metrology. In fieldbus terminology, "calibration" is often used to mean the configuration of a transmitter. In terminology pertaining to metrology, "calibration" means that you compare the transmitter to a traceable...

 Open the catalog to page 3
White Paper- The Safest Way to Calibrate Fieldbus Instruments-4

110680_Wo_KO_05_2011_Um_frz.qxp:100988_Wo_KO_05_2010_Um table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:8.00pt "Arial", sans-serif; } .font1 { font:9.00pt "Arial", sans-serif; } .font2 { font:11.00pt "Arial", sans-serif; } .font3 { font:13.00pt "Arial", sans-serif; } .font4 { font:9.00pt "Times New Roman", serif; } BEAMEX Calibration White Paper mixture is continuously prsent or prsent for a long time). Performance and functionality. Multifunctional intrinsically safe calibrators provide the functionality and performance of regular industrial calibration devices,...

 Open the catalog to page 4
White Paper- The Safest Way to Calibrate Fieldbus Instruments-5

110680_Wo_KO_05_2011_Um_frz.qxp:100988_Wo_KO_05_2010_Um table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:6.00pt "Arial", sans-serif; } .font1 { font:7.00pt "Arial", sans-serif; } .font2 { font:8.00pt "Arial", sans-serif; } .font3 { font:9.00pt "Arial", sans-serif; } .font4 { font:11.00pt "Arial", sans-serif; } .font5 { font:13.00pt "Arial", sans-serif; } .font6 { font:9.00pt "Times New Roman", serif; } .font7 { font:13.00pt "Times New Roman", serif; } BEAMEX Calibration White Paper HARFVr amm* A hazardous atmosphre is an area that contains l詩ments that may cause...

 Open the catalog to page 5

All Beamex catalogs and technical brochures

  1. MC6-Ex

    16  Pages

  2. Beamex MC2-IS

    12  Pages

  3. MCS200

    18  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.