MC2-Practical tools for field calibration - 1

Practical tools for field calibration Intrinsically Safe MC2-IS

Open the catalog to page 1
MC2-Practical tools for field calibration - 2

Practicality in calibration. The Beamex® MC2 series includes three different hand-held calibrators for fi eld use: the MC2 Pressure/Electrical Calibrator, the MC2 Temperature/Electrical Calibrator and the MC2 Multifunction Calibrator. The MC2 is a compactsized and easy-to-use hand-held calibrator. It has a large graphical display, a menu-based interface and full numerical keyboard. Being a Beamex calibrator, MC2 represents the high, uncompromised quality standards evident in other Beamex calibration equipment. Beamex® MC2-IS: Practical tool for calibration in hazardous environments Beamex®...

Open the catalog to page 2
MC2-Practical tools for field calibration - 3

3 The MC2 series includes four different portable calibrators MC2 Feature comparison Features MC2-PE Pressure / Electrical MC2-TE Temperature / Electrical MC2-MF Multifunction MC2-IS Intrinsically Safe Internal pressure module • - • (barometric only) Connection for external pressure modules • • • • Current measurement (with internal and external supply) • • • • Voltage measurement • • • • Frequency measurement • • • • Pulse counting • • • • Switch sensing • • • • Internal HART® compatible 24 VDC loop supply • • • •(1 Current generation (with internal and external supply) - • • •(2 Voltage...

Open the catalog to page 3
MC2-Practical tools for field calibration - 4

4 BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS Accuracy guaranteed The Beamex® MC2 and Beamex® MC2-IS calibrators are extremely accurate process calibrators. As proof of this, each MC2 and MC2-IS calibrator is delivered with a traceable, accredited calibration certifi cate. Compact and user-friendly The MC2 and MC2-IS calibrators are compact-sized, lightweight portable calibrators with a large graphical display and multilingual interface. They have a full numerical keyboard. Using the MC2 and MC2-IS is quick and easy. Robust field calibrator The MC2 and MC2-IS are robust and made for tough...

Open the catalog to page 4
MC2-Practical tools for field calibration - 5

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS 5 General Specifi cations General MC2 MC2-IS Display 60 mm x 60 mm (2.36” x 2.36”), 160 x 160 pixels LCD 60 mm x 60 mm (2.36” x 2.36”), 160 x 160 pixels LCD Weight 720 ... 830 g (1.59 ... 1.83 lbs) 1070 g (2.3 lbs) Dimensions 215 mm (8.5”) x 102 mm (4”) x 49 mm (1.9”) (d/w/h) 215 mm (8.5”) x 102 mm (4”) x 49 mm (1.9”) (d/w/h) Keyboard Membrane keyboard Membrane keyboard Battery type Rechargeable NiMH, 4000 mAh, 3.6V DC Rechargeable NiMH, 1700 mAh, 4.8V DC Charging time 5 hours 16 hours Charger supply 100...240 VAC, 50-60 Hz 100...240 VAC, 50-60 Hz...

Open the catalog to page 5
MC2-Practical tools for field calibration - 6

6 BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS Frequency measurement 0.0027 … 50 000 Hz Range Resolution 1 Year Uncertainty(±)(1 0.0027 ... 0.5 Hz 0.000001 Hz 0.01% RDG 0.5 ... 5 Hz 0.00001 Hz 0.01% RDG 5 ... 50 Hz 0.0001 Hz 0.01% RDG 50 ... 500 Hz 0.001 Hz 0.01% RDG 500 ... 5000 Hz 0.01 Hz 0.01% RDG 5000 ... 50000 Hz 0.1 Hz 0.01% RDG Feature Specification Temperature coefficient Specification valid from -10 to 50°C (14 ... 122°F) Input impedance > 1 MƒÇ Trigger level -1…14 V in 1 V steps and open collector inputs Minimum signal amplitude 2 Vpp (< 10 kHz), 3 Vpp (10...50 kHz) Supported units...

Open the catalog to page 6
MC2-Practical tools for field calibration - 7

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS 7 Internal Pressure Modules (IPM) Internal Module MC2 (3 Internal Module MC2-IS (4 Unit Range(2 Resolution 1 Year Uncertainty(±)(1 IPM200mC - kPa mbar iwc ±20 ±200 ±80 0.001 0.01 0.01 0.05% RDG + 0.05% FS IPM2C - kPa bar psi -100 to 200 -1 to 2 -14.5 to 30 0.01 0.0001 0.001 0.05% FS IPM20C - kPa bar psi -100 to 2000 -1 to 20 -14.5 to 300 0.1 0.001 0.01 0.05% FS IPM160 - MPa bar psi 0 ... 16 0 ... 160 0 ... 2400 0.001 0.01 0.1 0.05% FS Barometric option Barometric option -IS Also enables absolute pressure measurement for the above pressure inputs....

Open the catalog to page 7
MC2-Practical tools for field calibration - 8

8 BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS mV measurement (T/C-terminals) -25 ... 150 mV Range Resolution 1 Year Uncertainty(±)(1 -25 ... 150 mV 0.001 mV 0.02% RDG + 4 ìV Feature Specification Temperature coefficient < ±0.0015% RDG / °C outside of 18 ... 28°C < ±0.0008% RDG / °F outside of 64.4 ... 82.4°F Input impedance > 10 MƒÇ Supported units V, mV, ìV Display update rate 3 / second mV generation (T/C-terminals) -25 ... 150 mV Range Resolution 1 Year Uncertainty(±)(1 -25 ... 150 mV 0.001 mV 0.02 % RDG + 4 ìV Feature Specification Temperature coefficient < ±0.0015% RDG / °C outside of...

Open the catalog to page 8
MC2-Practical tools for field calibration - 9

BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS 9 Resistance measurement 0 ... 4000 ƒÇ Range Resolution 1 Year Uncertainty(±)(1 0 ... 250ƒnƒÇ 1 mƒÇ 4-wire connection: 0.02 % RDG + 3.5 mƒÇ 3-wire connection: 0.02% RDG + 13.5 mƒÇ 250 ... 2650ƒnƒÇ 10 mƒÇ 2650 ... 4000ƒnƒÇ 100 mƒÇƒn Feature Specification Temperature coefficient < ±0.0015% RDG / °C outside of 18 ... 28°C < ±0.0008% RDG / °F outside of 64.4 ... 82.4°F Measurement current Pulsed, bi-directional 1 mA (0..500 ƒÇ), 0.2 mA (>500 ƒÇ). Supported units ƒÇ, kƒÇ Display update rate 3 / second Resistance simulation 0 ... 4000 ƒÇ Range Resolution...

Open the catalog to page 9
MC2-Practical tools for field calibration - 10

10 BEAMEX - WORLD-CLASS CALIBRATION SOLUTIONS Thermocouple measurement and simulation Thermocouple types available as standard Type Range (°C) Range (°C) 1 Year Uncertainty (±)(1 B(2 0 ... 1820 0...200 200...400 400...1820 (3 2.0 °C 1.0 °C R(2 -50 ... 1768 -50...0 0...100 100...1768 1.0 °C 0.8 °C 0.6 °C S(2 -50 ... 1768 -50...0 0...1768 1.0 °C 0.7 °C E(2 -270 ... 1000 -270...-200 -200..1000 (3 0.25 °C J(2 -210 ... 1200 -210...1200 0.3 °C K(2 -270 ... 1372 -270...-200 -200...1000 1000...1372 (3 0.3 °C 0.4 °C N(2 -270 ... 1300 -270...-200 -200...1300 (3 0.4 °C T(2 -270 ... 400 -270...-200...