FB/MB a smarter way to calibrate temperature
1 / 20Pages

Catalog excerpts

FB/MB a smarter way to calibrate temperature - 1

1 a smarter way to calibrate temperature Products and Services: • Beamex® Field Temperature Block (FB Series) • Beamex® Metrology Temperature Block (MB Series) • Beamex® Smart Reference Probes • Temperature Calibration Laboratory Services

Open the catalog to page 1
FB/MB a smarter way to calibrate temperature - 2

2 Introducing a smarter way to calibrate temperature Beamex introduces a smarter, more efficient and accurate solution for calibrating temperature. It is a complete solution for temperature calibration with various different products and services, such as a series of high-quality dry blocks for fi eld and laboratory use, smart reference probes and temperature calibration laboratory services. High-quality dry blocks Beamex has two different dry block series: the Beamex® FB Series Field Temperature Blocks and the Beamex® MB Series Metrology Temperature Blocks. The dry blocks in the FB Series...

Open the catalog to page 2
FB/MB a smarter way to calibrate temperature - 3

3 Beamex® MB Series Portable temperature dry block that provides bath-level accuracy for industrial applications. Main features: • High accuracy: a dry block that provides bath-level accuracy • Built-in high-accuracy reference probe input (in R model) • Immersion depth up to 200 mm • Wide temperature range from -45 °C to +700 °C • Accredited calibration certifi cate included as standard • Part of Beamex® Integrated Calibration Solution (ICS) • Warranty: 1 year Available models: MB Model Range MB140 / MB140R -45 °C … +140 °C MB155 / MB155R -30 °C … +155 °C MB425 / MB425R +35 °C … +425 °C...

Open the catalog to page 3
FB/MB a smarter way to calibrate temperature - 5

5 Introducing an integrated and automated temperature calibration solution. Seamless connectivity! The Beamex® dry blocks communicate with the selected Beamex® MC-calibrators enabling fully automated temperature calibration and documentation! About Beamex® Integrated Calibration Solution The Beamex® Integrated Calibration Solution improves the quality and efficiency of the entire calibration system through faster, smarter and more accurate management of all calibration assets and procedures. Beamex calibrators, workstations, calibration software and professional services form the most...

Open the catalog to page 5
FB/MB a smarter way to calibrate temperature - 7

7 Main Features of the Beamex® MB The Beamex® Metrology Temperature Block (MB) is a high accuracy temperature dry block. It delivers bath-level accuracy in a convenient dry block. It enables you to take laboratory-level accuracy performance with you out to the fi eld. The unique dual zone control techniques enable excellent stability and uniformity. Immersion depth to 203 mm and temperature ranges from -45 °C … +700 °C. High accuracy and stability With a conventional dry block, you typically needed an external reference sensor if you wanted better accuracy. The Beamex® MB has accurate...

Open the catalog to page 7
FB/MB a smarter way to calibrate temperature - 9

9 Main Features of the Beamex® FB Beamex Field Temperature Block The Beamex® Field Temperature Block (FB) is an ideal temperature block for industrial fi eld use. It is lightweight and easy to carry. It is a very fast dry block, yet it provides excellent accuracy. Lightweight, portable The Beamex® FB Field Temperature Block is ideal for industrial fi eld use. It only weights about 8 kg, and it is small enough to carry around. Speed The Beamex® FB is extremely quick to reach various temperatures, i.e. it cools down to -25 °C in 15 minutes and heats up to +660 °C in 15 minutes. This saves...

Open the catalog to page 9
FB/MB a smarter way to calibrate temperature - 10

10 Beamex® FB Series Specifi cations FB150 FB350 FB660 Temperature range at 23 °C –25 °C to 150 °C (–13 °F to 302 °F) 33 °C to 350 °C (91 °F to 662 °F) 50 °C to 660 °C (122 °F to 1220 °F) Display accuracy ±0.2 °C Full range ±0.2 °C Full range ±0.35 °C at 50 °C ±0.35 °C at 420 °C ±0.5 °C at 660 °C Stability ±0.01 °C Full range ±0.02 °C at 33 °C ±0.02 °C at 200 °C ±0.03 °C at 350 °C ±0.03 °C at 50 °C ±0.05 °C at 420 °C ±0.05 °C at 660 °C Axial uniformity at 40 mm (1.6 in) ±0.05 °C Full range ±0.04 °C at 33 °C ±0.1 °C at 200 °C ±0.2 °C at 350 °C ±0.05 °C at 50 °C ±0.35 °C at 420 °C ±0.5 °C at...

Open the catalog to page 10
FB/MB a smarter way to calibrate temperature - 11

11 Beamex® MB Series Specifi cations MB140 MB155 MB425 MB700 Temperature range at 23 °C –45 °C to 140 °C (–49 °F to 284 °F) –30 °C to 155 °C (–22 °F to 311 °F) 35 °C to 425 °C (95 °F to 797 °F) 50 °C to 700 °C3) (122 °F to 1292 °F) Display accuracy ±0.1 °C Full range ±0.1 °C Full range ±0.1 °C to 100 °C ±0.15 °C to 225 °C ±0.2 °C to 425 °C ±0.2 °C to 425 °C ±0.25 °C to 660 °C Stability ±0.005 °C Full range ±0.005 °C Full range ±0.005 °C to 100 °C ±0.008 °C to 225 °C ±0.01 °C to 425 °C ±0.005 °C to 100 °C ±0.01 °C to 425 °C ±0.03 °C to 700 °C Axial uniformity 40 mm (1.6 in) ±0.08 °C: to –35...

Open the catalog to page 11
FB/MB a smarter way to calibrate temperature - 12

12 Inserts Inserts for FB models: Insert Model Description MH1 FB150 Multihole, metric / reference; ¼”, 3 mm, 4 mm, 6 mm, 8 mm, 10 mm MH1 FB350, FB660 Multihole, metric / reference; ¼”, 4 mm, 6 mm, 8 mm, 10 mm MH2 All models Multihole, metric / reference; ¼”, 2x3 mm, 2x4 mm, 6 mm B All models Blank Special All models Special Please contact Beamex for custom inserts. 10 mm 8 mm 6 mm 1/4 in 4 mm 3 mm FB150-MH1 MH1 FB350-MH1, FB660-MH1 FB350-MH2, FB660-MH2 MH2 10 mm 6 mm 8 mm 4 mm 1/4 in Inserts for MB models: Insert Model Description MH1 All models Multihole, metric / reference; ¼”, 3 mm, 4...

Open the catalog to page 12
FB/MB a smarter way to calibrate temperature - 13

13 Beamex® Smart Reference Probes The Beamex® Smart Reference Probe is a high-quality and extremely stable PRT probe with an integrated memory that stores the individual sensor coefficients. The sensor works as plug-and-play with Beamex® FB series of temperature blocks (R model). The temperature block automatically reads the sensor coefficients from the sensor and makes the necessary adjustments. This eliminates the need to enter the coefficients manually. The sensor can also be used with the Beamex® MB series of temperature blocks (R model). The sensor coefficients can be manually entered...

Open the catalog to page 13