1. Catalogs
  2. BD|SENSORS GmbH
  3. data sheet LMK 382 H

data sheet LMK 382 H

data sheet LMK 382 H

data sheet LMK 382 H

Product catalog summary
Overview: The LMK 382H is a stainless steel probe designed for continuous level measurement in various fluids, featuring HART® communication and a ceramic sensor. It offers high accuracy and stability, suitable for environments with sewage, polluted, and high-viscosity fluids.
Specifications:
  • Accuracy: 0.1% FSO according to IEC 60770.
  • Nominal Pressure: Ranges from 0.06 to 20 bar, equivalent to 0.6 to 200 mH2O.
  • Output Signal: Standard 2-wire, 4-20 mA with HART® communication.
  • Temperature Range: Up to 85°C permissible.
  • Materials: Housing made of stainless steel 1.4404 (316L), diaphragm options include ceramics Al2O3 96% or 99.9%.
Features:
  • High overpressure resistance and long-term stability.
  • Optional versions include intrinsically safe (IS) for gas and dust, mounting with stainless steel pipe, and flange version.
  • Adjustable parameters such as electronic damping, offset, and span.
Applications:
  • Water: Groundwater level measurement, rain spillway basins.
  • Sewage: Wastewater treatment, water recycling.
  • Fuel and Oil: Level monitoring in open tanks, fuel storage, tank farms, biogas plants.
Performance:
  • Accuracy varies with pressure range and turn-down ratio (TD).
  • Long-term stability is ≤ ± (0.1 x turn-down) % FSO/year.
  • Thermal effects are within a tolerance band of ≤ ± 1% FSO in the compensated range of -20 to 80°C.
Mechanical and Electrical Details:
  • Vibration resistance up to 4 g according to DIN EN 60068-2-6.
  • Ingress protection rated at IP 68.
  • Electrical protection includes short-circuit and reverse polarity protection.
Explosion Protection: Approved for use in explosive atmospheres with specific safety technical maximum values and permissible media temperatures for different zones.
Accessories: Includes options for cable protection, mounting flanges, and terminal clamps, with various materials and configurations available.
See more

Catalog excerpts

data sheet LMK 382 H-1

LMK 382H Stainless Steel Probe with HART®-communication Ceramic Sensor accuracy according to IEC 60770: 0.1 % FSO Nominal pressure The stainless steel probe LMK 382H has been designed for continuous level measurement in sewage, polluted and higher viscosity fluids. Basic element is a robust and high overpressure capable capacitive ceramic sensor e.g. for low levels. Output signals 2-wire: 4 ... 20 mA others on request HART® communication (setting of offset, span and damping) high overpressure resistance Water ground water level measurement rain spillway basins high long-term stability Sewage waste water treatment water recycling IS-version Ex ia = intrinsically safe for gas and dust mounting with stainless steel pipe flange version Fuel and oil level monitoring in open tanks with low filling heights fuel storage tank farms biogas plants accessories e.g. transmitter and mounting flanges and terminal clamp BD SENSORS GmbH BD-Sensors-Straße 1 D - 95199 Thierstein

 Open the catalog to page 1
data sheet LMK 382 H-2

LMK 382H Stainless Steel Probe Technical Data Pressure ranges 1 Nominal pressure [bar] 0.06 Level [mH2O] 0.6 Overpressure [bar] 2 Max. ambient pressure (housing): 40 bar 1 Permissible load Long term stability Influence effects Turn-on time Mean response time Max. response time Adjustability 12 ... 36 VDC with HART® communication 14 ... 28 VDC with HART® communication TDmax = 1:10 ≤ ± 0.2 % FSO ≤ ± [0.2 + 0.03 x TD] % FSO pN < 160 mbar TDmax = 1:3 ≤ ± [0.2 + 0.1 x TD] % FSO pN ≥ 1 bar TDmax = 1:10 TD ≤ 1:5 ≤ ± 0.1 % FSO TD > 1:5 ≤ ± [0.1 + 0.02 x TD] % FSO Rmax = [(VS - VS min) / 0.02 A] Ω load...

 Open the catalog to page 2
data sheet LMK 382 H-3

LMK 382H Stainless Steel Probe Miscellaneous Option cable protection for probes Ingress protection Current consumption Weight CE-conformity ATEX Directive Wiring diagram Technical Data prepared for mounting with stainless steel pipe IP 68 max. 21 mA approx. 400 g (without cable) EMC Directive: 2014/30/EU 2014/34/EU 2-wire-system (current) HART® supply + Pin configuration Electrical connection Supply + Supply – Shield cable colours (IEC 60757) WH (white) BN (brown) GNYE (green-yellow) flange version prepared for mounting with stainless steel pipe  transmitter flange is not part of supply and...

 Open the catalog to page 3
data sheet LMK 382 H-4

LMK 382H Stainless Steel Probe Technical Data Accessories Transmitter flange for flange version Technical data Suitable for Flange material Hole pattern Ordering type Transmitter flange DN25 / PN40 Transmitter flange DN50 / PN40 Transmitter flange DN80 / PN16 LMK 382, LMK 382H, LMK 458, LMK 458H stainless steel 1.4404 (316L) according to DIN 2507 Ordering code ZSF2540 ZSF5040 ZSF8016 Mounting flange with cable gland cable gland M16x1.5 with seal insert (for cable-∅ 4 ... 11 mm) size b D d2 d4 f k n Technical data Suitable for all probes Flange material stainless steel 1.4404 (316L) Material of...

 Open the catalog to page 4
data sheet LMK 382 H-5

stainless steel 1.4404 (316L) customer HART®-communication 4 … 20 mA / 2-wire HART®-communication intrinsic safety 4 … 20 mA / 2-wire customer FKM EPDM FFKM customer Electrical connection PVC-cable (grey, Ø 7.4 mm) PUR-cable (black, Ø 7.4 mm) FEP-cable (black, Ø 7.4 mm) TPE-U-cable (blue, Ø 7.4 mm) customer Accuracy pN ≥ 1 bar: 0.1 % FSO pN < 1 bar: 0.2 % FSO customer Cable length in m Special version standard prepared for mounting with stainless steel pipe flange version customer stainless steel pipe is not part of the supply shielded cable with integrated ventilation tube for atmospheric pressure...

 Open the catalog to page 5

All BD|SENSORS GmbH catalogs and technical brochures