DMP 331i DMP 333i Precision Pressure Transmitter Stainless Steel Sensor accuracy according to IEC 60770: 0.1 % FSO Nominal pressure from 0 ... 400 mbar up to 0 ... 600 bar The precision pressure transmitter DMP 331i and DMP 333i demonstrate the further development of our industrial pressure transmitters. Output signal 2-wire: 4 ... 20 mA 3-wire: 0 … 10 V others on request Product characteristics ► thermal error in compensated range -20 ... 80 °C: 0.2 % FSO TC 0.02 % FSO / 10K The signal processing of sensor signal is done by digital electronics with 16-bit analogue digital converter. Consequently, it is possible to conduct an active compensation and the transmitters with excellent measurements and exceptionally attractive price to offer on the market. Preferred areas of use are excellent long term stability Laboratory techniques IS-versions Ex ia = intrinsically safe for gases and dusts welded pressure sensor pressure port G1/2" flush customer specific versions BD SENSORS GmbH BD-Sensors-Straße 1 D - 95199 Thierstein Energy production (gas consumption and thermal energy measurement) www.bdsensors.de info

Open the catalog to page 1
data sheet DMP i - 2

DMP 331i / DMP 333i Precision Pressure Transmitter Technical Data Pressure ranges DMP 331i Nominal pressure gauge / absolute Overpressure Nominal pressure gauge Pressure ranges DMP 333i Nominal pressure gauge / absolute Overpressure Vacuum ranges Output signal / Supply Standard Option IS-version Options analogue signal Long term stability ≤ ± 0.1 % FSO current 2-wire: Rmax = [(VS – VS min) / 0.02 A] Ω voltage 3-wire: Rmin = 10 kΩ supply: 0.05 % FSO / 10 V load: 0.05 % FSO / kΩ ≤ ± 0.1 % FSO / year at reference conditions Response time Permissible load Influence effects accuracy according to...

Open the catalog to page 2
data sheet DMP i - 3

DMP 331i / DMP 333i Precision Pressure Transmitter Technical Data Mechanical stability Vibration 10 g RMS (20 ... 2000 Hz) according to DIN EN 60068-2-6 Shock 100 g / 11 msec. according to DIN EN 60068-2-27 Explosion protection (only for 4 ... 20 mA / 2-wire) Approvals DX19-DMP 331i IBExU 10 ATEX 1068 X / IECEx IBE 12.0027X DX19-DMP 333i zone 0: II 1G Ex ia IIC T4 Ga zone 20: II 1D Ex ia IIIC T135 °C Da Safety technical max. values Ui = 28 V, Ii = 93 mA, Pi = 660 mW, Ci ≈ 0 nF, Li ≈ 0 µH, the supply connections have an inner capacity of max. 27 nF to the housing Permissible temperatures for...

Open the catalog to page 3
data sheet DMP i - 4

DMP 331i / DMP 333i Precision Pressure Transmitter Technical Data Electrical connections (dimensions mm / in) cable outlet with PVC cable (IP 67) 5 cable outlet, cable with ventilation tube (IP 68) 6 compact field housing (IP 67) universal-field housing stainless steel 316L with cable gland M20x1.5 (ordering code 880) and other versions on request standard: 2 m PVC cable without ventilation tube (permissible temperature: -5 ... 70 °C) different cable types and lengths available, permissible temperature depends on kind of cable

Open the catalog to page 4
data sheet DMP i - 5

DMP 331i / DMP 333i Precision Pressure Transmitter Technical Data with electrical connection Bayonet MIL-C-26482 (10-6) increases the length of devices by 5 mm for nominal pressure pN > 400 bar increases the length without IS-version by 19 mm and with IS-version by 39 mm Mechanical connections (dimensions mm / in) metric threads and others on request www.bdsensors.de info@bdsensors.de © 2023 BD|SENSORS GmbH – The specifications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modifications to the specifications and materials.

Open the catalog to page 5
data sheet DMP i - 6

gauge absolute Electrical connection male and female plug ISO 4400 male plug Binder series 723 (5-pin) male plug M12x1 (4-pin) / metal - for analog output male plug M12x1 (4-pin) / metal - for digital output Bayonet MIL-C-26482 (10-6); 2 wire Bayonet MIL-C-26482 (10-6); 3 wire cable outlet with PVC cable (IP67) 3 cable outlet, cable with ventilation tube (IP68) 4 compact field housing stainless steel 1.4301 (304) customer Mechanical connection G1/2" DIN 3852 G1/2" EN 837 G1/4" DIN 3852 G1/4" EN 837 G1/2" DIN 3852 with flush sensor G1/2" DIN 3852 open pressure port 1/2" NPT 1/4" NPT customer...

Open the catalog to page 6