The MOVISTROB® Series 600 is a high-power stroboscope designed for analyzing rapid periodic motions. It undergoes stringent quality and safety tests to meet high standards. Users are advised to read the operating instructions thoroughly for safe and effective use.
General Description
The stroboscope emits short light flashes to make rapid motions appear slow or stationary, facilitating observation and speed measurement without mechanical interference. The MOVISTROB® 600.00 is noted for its high light intensity, stability, accuracy, and user-friendliness, making it suitable for large objects and high-speed photography.
Components
The device includes a control unit and a flashlamp with a high-output Xenon flash tube. It supports both internal and external flash frequency control, featuring digital readout and phase-shift capabilities.
Controls and Indicators
The stroboscope features various controls and indicators, including an output socket, display window, selector switch, adjustment knob, phase shifter, power pushbutton, range selector pushbuttons, external control pushbutton, line synchronization pushbutton, and trigger input for external signal sources.
General Instructions
To operate, connect the flashlamp to the control unit and power it on. For internal control, select the frequency range and adjust with the generator knob. For external control, connect a pulse generator and select the appropriate frequency range.
Stroboscopic Principle
Stroboscopy involves using brief light flashes to observe high-speed periodic motion, creating the illusion of a stopped or slow-motion image for detailed observation.
Flash Frequency and Stopped Images
The relationship between flash frequency and stopped images is explained. The flash period must be a whole-number multiple of the rotation period to create a stopped image. Adjusting flash frequency can change the observed phase of rotation.
Measurement of RPM and Frequencies
RPM can be measured by identifying the highest flash frequency that results in a stopped image or by calculating rotational frequency from two neighboring flash frequencies, with formulas provided.
Slow-Motion Cycle
If the flash period slightly deviates from a whole-number multiple of the rotation time, the object appears to rotate slowly. Formulas are provided to calculate the apparent angular speed.
Stopped Images of Phantom Objects
Stopped images can occur at other flash frequencies, representing phantom objects. Conditions and characteristics of these images are explained.
Objects with Finite Rotational Symmetry
For symmetrical objects, stopped images result from specific flash periods and frequencies, with formulas provided.
Replacing Flash Tube or Fuses
Instructions for safely replacing the flash tube and fuses emphasize disconnecting the power source and checking for residual charge before servicing.
Maintenance and Repair
If the instrument is unsafe, it should be removed from operation. Repairs should be conducted by trained personnel or returned to the manufacturer.
Available Accessories
Available accessories include a carrying case and tripod base plate.
Technical Specifications
Technical specifications for Model 600.00 include power supply, power consumption, flash tube type, light intensity, flash duration, frequency range, digital readout, accuracy, and control features.
Mechanical Data
Specifications for the flashlamp and control unit include dimensions, weight, and connection details.
See more

Catalog excerpts

Series 600-1

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:10.00pt "Times New Roman", serif; } .font1 { font:15.00pt "Times New Roman", serif; } .font2 { font:23.00pt "Times New Roman", serif; } .font3 { font:14.00pt "Verdana", sans-serif; } Bamberg + Bormann ELECTRONIC GmbH INSTRUCTION MANUAL MOVISTROB® Sries 600 High-power Stroboscope Type 600.00 BBE Bamberg + Bormann - Electronic GmbH WiebelsheidestraBe 45 D-59757 Arnsberg / Neheim-Hsten Tel.: 0049 (0)2932-547760 Fax: 0049 (0)2932-34675 Internet: http://www.bbe-electronic.de e-mail: [email protected]

 Open the catalog to page 1
Series 600-2

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:6.00pt "Arial", sans-serif; } .font1 { font:9.00pt "Times New Roman", serif; } .font2 { font:10.00pt "Times New Roman", serif; } Bamberg + Bormann ELECTRONIC GmbH L Introduction Each MOVISTROB® product has to pass through various controls during its production phases and must also undergo very strict and conscientious function and quality tests before leaving the factory for delivery to our clients. We can assure you that the MOVISTROB® product you received is in strict conformity with our...

 Open the catalog to page 2
Series 600-3

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:10.00pt "Times New Roman", serif; } .font2 { font:18.00pt "Times New Roman", serif; } If you desire to study line-synchronous cycles, such as slip measurements, you can control the unit directly through the line frequency by pressing the LINE" button switch. You can also use the phase-shift control knob to change the phase position in this mode. The flash rate is continuously adjustable from 120 to 19200 flashes/minute, equivalent to 2 to 320...

 Open the catalog to page 3
Series 600-5

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:10.00pt "Times New Roman", serif; } positively-directed edge of an impulse. The impulse current (maximum) should not exceed 100 V. The response cycle lies at 2,5 V (TTL). Caution: Always press the proper "Frequency Range Selector Button " (8) for the flash frequency range in which the external synchronization frequency lies. In any case if the control impulse frequency exceeds the preselected low range the next higher range should be selected...

 Open the catalog to page 5
Series 600-6

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:10.00pt "Times New Roman", serif; } 5. Stroboscopic Principle With stroboscopy, high-speed periodic motion which cannot be followed by unassisted eyes can be made accessible for observation and its frequency measured. For this purpose the oscillating or rotating object is illuminated in a periodic series of light impulses (flashes) which are as brief as possible. The object then appears (at the appropriate flash frequency) to be motionless...

 Open the catalog to page 6
Series 600-7

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:10.00pt "Times New Roman", serif; } The opposite holds for e < 0. With increasing /e/ the angular speed w' of the apparent rotation rises. Finally the angle 2pe becomes so large that the mark on the rotating dise appears at two different places during two succeeding flashes. Other phenomena (described below) also occur. 5.4 Stopped Images of Phantom Objects Stopped images of rotating objects results from flash frequency periods Tn = nr, and...

 Open the catalog to page 7
Series 600-8

500 - Art.Nr.1023, 1025_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:6.00pt "Arial", sans-serif; } .font1 { font:12.00pt "Arial", sans-serif; } .font2 { font:9.00pt "Times New Roman", serif; } .font3 { font:10.00pt "Times New Roman", serif; } Bamberg + Bormann ELECTRONIC GmbH Two fuse-holders are located in the control unit at 90° to each other; these are marked with the appropriate operating current. For 230V WE operation a 2.5A T fuse is installed; for 115V WE, a 4A T fuse. The unit is always delivered set up for either 230V/50-60 Hz operating current or...

 Open the catalog to page 8
Series 600-9

600 - Art.No. 1136, 1137_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:10.00pt "Times New Roman", serif; } Technical Spcifications Power supply: Power consumption: Flash Tube: Light intensity: Model 600.00 230 / 115 V AC 50-60 Hz approx. 135 VA high energy, socket mounted Xenon quarz rod shapped flash bulb depending of frequency range Range I 5000 Lux Range II 4000 Lux Range III 2500 Lux at a distance of 0.5 m from light source. depending on frequency range 1 20^s II 15^s III 10^s 2 - 320 Hz / 120 - 19.200 rpm,...

 Open the catalog to page 9
Series 600-10

600 - Art.No. 1136, 1137_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:5.00pt "Arial", sans-serif; } .font1 { font:9.00pt "Times New Roman", serif; } .font2 { font:10.00pt "Times New Roman", serif; } Bamberg + Bormann ELECTRONIC GmbH Flashlamps: 13 Flashlamp Type GS Flashlamp Type KS Dimensions: 270 x 220 x 160 (GS) 180 x 160 x 130 (KS) w x h x d Weight: 2,00 kg (GS) 1,40 kg (KS) length of connection 3,50 m standard cable: up to 10,0 m at request 13: Linear, high-output Xenon flash bulb B11038 (GS) or B11058 (KS) 14: " ON - OFF " - switch Right of technical...

 Open the catalog to page 10
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.