Introduction
The MOVISTROB Series 1000 stroboscope is designed for observing rapid periodic processes by using short light flashes to make objects appear stationary. It features a microprocessor control and an LCD screen for displaying information and commands.
Operating the Stroboscope
  • Front Board Elements: Includes keys and display for operation.
  • Rear Unit Ports: Contains the main switch, AC socket, handlamp port, and input/output ports for external triggers.
  • Menu Layout: Organized into main and submenus such as Internal Trigger, External Trigger, Line Synchron, and Memory.
  • Keypad Operations: Navigation is done using cursor keys, numeric keys for input, and unit keys for setting measurement units.
Functions in Detail
  • Internal Oscillator: Produces flashes ranging from 30 - 18,000 RPM or 0.5 - 300 Hz.
  • Internal Burst: Generates a series of flashes with selectable numbers and intervals.
  • External Trigger and Line Synchron: Utilizes external emitters for triggering to ensure a stopped image of the object.
  • Readout RPM/Hz/T: Accurately measures external trigger frequency.
  • Phase Shifter: Adjusts the phase of external trigger signals.
  • Slow Motion: Creates slow-motion effects from external triggers.
  • Prescaler: Allows prescaling of external trigger signals.
  • Memory Functions: Stores and recalls readouts, displaying absolute and relative deviations.
RS 232 Serial Interface
Enables PC control through the RS 232 interface, with adjustable data transmission settings.
Inputs and Outputs
Details various input and output ports for connecting external devices.
Technical Data
Provides specifications and operational parameters of the stroboscope.
Safety and Maintenance
  • Only trained personnel should operate the device.
  • High voltages are present; refer to the lamp replacement section before opening.
  • Avoid liquids or metallic objects entering the device.
Specifications
  • Resolution: 0.001% quartz timebase.
  • Measurement Range: 60 – 18000 RPM, 1 – 300 Hz, 3.33 – 1000 ms.
  • Accuracy: 0.001%.
  • File Storage: 25 files (00 - 24).
Readout Display
  • Modes: RPM, Hz, and period (ms).
  • Deviation Display: Shows absolute and relative deviations from a reference value.
  • Memory: Readouts can be stored and recalled even if disconnected from the mains, provided a backup battery is used.
Phase Shifter Function
  • Allows phase shift of a trigger signal from 0 to 359 degrees with 1/10th degree resolution.
  • Adjustable using numeric or cursor keys.
Slow Motion Function
  • Produces slow motion effects with a period range of 0.5 – 30 seconds.
  • Rotation direction can be changed or stopped using cursor keys.
Prescaler Function
  • Prescales an external trigger frequency with a value between 1 and 256.
Memory Management
  • Recall Memory: Retrieve stored readouts using file numbers 00 – 24.
  • Clear Memory: Erase all stored readouts.
RS 232 Serial Interface
  • Allows integration with computers for automated measurements.
  • Data output in ASCII code, adjustable transmission parameters.
  • Baud rates: 600Bd to 9600Bd.
Inputs and Outputs
  • External emitter connection via a seven-pole diode plug with bayonet lock.
Specifications
1. Power Supply: 1/2 +6V DC, max. 315 mA; 3 +12V DC across 150 KΩ for switching contact.
2. Signal Input: 0.5 - 50V, R = 270 KΩ.
3. Output Port: 12 Vss positive, 600 Ohms resistance.
4. Operating Current: 230 V / 50-60 Hz, 115 V on request.
5. Light Intensity: Standard model 700 LUX.
6. Flash Period: 2 - 6 µs approx.
7. Dimensions: W = 246 mm, H = 87 mm, T = 200 mm.
8. Weight: Control section, 3 kg approx.
Procedures
1. External Triggering: Resistance should not exceed 100 Kohms when contacts are closed; short circuit current is less than 50 uA.
2. Serial Interface: Pin connection for data cable includes signal ground, serial input, and output.
Technical Data
1. Control of Flash Sequence: Internal by quartz oscillator, external by periodic signals, and line synchronous.
2. Phase Shift: Independent of frequency, resolution of 1/10th degree.
3. Frequency Range: Internal control 30 - 18000 RPM, 0.5 - 300 Hz; External control 60 - 18000 RPM, 1 - 300 Hz.
4. Readout Display: Selectable in RPM, Hz, and ms.
5. Accuracy Timebase: 10E-5 (0.001% - quartz timebase).
Recommendations
1. Use suitable cables for connecting external devices.
2. Ensure ambient temperature is between 0 – 50 degrees C for operation.
Additional Information
1. Memory Function: 25 storage locations, unaffected by power disconnection.
2. Display: Alphanumeric backlit LCD display.
3. Housing: Light aluminum half-shell construction, height adjustable with carrying handle.
See more

Catalog excerpts

Series 1000 -1

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:22.00pt "Tahoma", sans-serif; } .font1 { font:17.00pt "Times New Roman", serif; } .font2 { font:23.00pt "Times New Roman", serif; } .font3 { font:33.00pt "Times New Roman", serif; } Bamberg + Bormann ELECTRONIC GmbH INSTRUCTION MANUAL MOVISTROB* Sries 1000 Type Digital MOVISTROB 1000 IBE Bamborg i Boi inaiiu Electronic GmbH WlcDcIshcidcstraBe 45 D 59757 Arnsbrrg / Neheim Hiisteo Tel.: 0049 (0)2932 547760 Fax: 0049(0)2932 34675 Internet: http://www.bbe elcclronic.de e-mail: [email protected]

 Open the catalog to page 1
Series 1000 -2

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:16.00pt "Arial", sans-serif; } .font1 { font:12.00pt "Georgia", serif; } .font2 { font:12.00pt "Times New Roman", serif; } .font3 { font:13.00pt "Times New Roman", serif; } .font4 { font:17.00pt "Times New Roman", serif; } .font5 { font:18.00pt "Times New Roman", serif; } Table of contents 1. Iiitnxliiclion 2. Ooerating the strohoscoi* 2.1 Opcrating and display clments on the front board 2.2 Switch and ports on the rear of the unit 2.3 Menus-layout 2.4 Kcypad oprations 2.4.1 Cursor keys 2.4.2 Numeric keys 2A3 Unit...

 Open the catalog to page 2
Series 1000 -3

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:13.00pt "Times New Roman", serif; } .font1 { font:17.00pt "Times New Roman", serif; } Brief Description: Introduction Page: 4 Opcrating the stroboscope : General informations about operating and display lments. Page: 5-9 switches and ports, menu tayaut kc\pad op驩rations cet. Internai Trigger : Page: 10-11 Ail triggering signais are interually created within the device Internai Oscillator Page: 10-11 Flash frequency produced by built-in quarts oscillai Internai Burst Page: 11 Products a sries of flashs uilli delined...

 Open the catalog to page 3
Series 1000 -4

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:17.00pt "Times New Roman", serif; } 1. Introduction With stroboscopes you can observe Ihe motion of rapid periodic or semiperiodic processes (stich as rolating or oscillating objects). If an object (such as a rotatiug shaft) is itluminatcd with short flashes of light at exactly the saine point, it will appear to stand still, because the hutnan eye is too slow to detect the individual images. stroboscope also tends ilself to hands-off measurement of r)volutions. Ail functions of the MS 1000 stroboscope aie controlled...

 Open the catalog to page 4
Series 1000 -5

600 - Art.No. 1136, 1137_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:10.00pt "Times New Roman", serif; } 2. Operating the stroboscope 2.1 Operating and display lments on the front board Keys and display are located on the front board (see illustration 2.1, Operating and Display Elements). 2.2 Switch and ports on rear of the unit The following are located on the rear of the unit: Main switch AC cable socket Handlamp port Input port for external trigger: 7-pin diode bayonet socket Impulse output: 3-pin diode bayonet socket 2.3 Menu layout All stroboscope functions...

 Open the catalog to page 5
Series 1000 -6

600 - Art.No. 1136, 1137_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:7.00pt "Arial", sans-serif; } .font1 { font:8.00pt "Arial", sans-serif; } .font2 { font:9.00pt "Arial", sans-serif; } .font3 { font:10.00pt "Arial", sans-serif; } .font4 { font:12.00pt "Arial", sans-serif; } .font5 { font:8.00pt "Times New Roman", serif; } .font6 { font:9.00pt "Times New Roman", serif; } .font7 { font:10.00pt "Times New Roman", serif; } Illustration 2.1 Opration and Display Elments on Front Board MICROPROCESSOR CONTROLLED LC - Display 2 x 20 Characters STO 8 9 HZ 6 RCL RPM...

 Open the catalog to page 6
Series 1000 -7

600 - Art.No. 1136, 1137_V2 table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:5.00pt "Arial", sans-serif; } .font1 { font:10.00pt "Times New Roman", serif; } Bamberg + Bormann ELECTRONIC GmbH 2.4.1 Cursor Keys: < > / A y With the keys on the cursor pad you can move up and down in a menu or raise and lower a numerical value on the LCD display. A Raising a value at the cursor position y Lowering a value at the cursor position < Moving the cursor to the left > Moving the cursor to the right 2.4.2 Numeric Keys: 0 - 9 / . / C The keys on the numeric keypad allow you the...

 Open the catalog to page 7
Series 1000 -8

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:9.00pt "Times New Roman", serif; } .font1 { font:12.00pt "Times New Roman", serif; } .font2 { font:17.00pt "Times New Roman", serif; } .font3 { font:18.00pt "Times New Roman", serif; } ,-r,.tvoiK." LECTBOMC GmbH 2.5 Rccalliiig a submenu or a function Use the cursor keys A / V to inake your sɩlections &om the menu. In Ihe Main menu: - Internai Trigger - Kx tentai Trigger - Line Synchron - Memory - Sriai I / O In the lnicii subinenu: - Internai Oscillator - Internai Burst In the Kxtern and Line menu: -Readout RPM/ Hz...

 Open the catalog to page 8
Series 1000 -9

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:13.00pt "Times New Roman", serif; } .font1 { font:17.00pt "Times New Roman", serif; } Press the ENTER key to recall the Line Synchron subinenu. Key: ENTER Line: V = Down A = Up Readout RPM/Hz/T Now you are in the Line Synchron submenu on the first line of the menu. Press the V key lo recall the Phase Shifter . Now the stroboscope is in the desired Phase Shifter function. Key: V Line: V = Down A = Up Phase Shifter With the ENTER key recall the Phase Shifter . Now the stroboscope is in the desired Phase Shifter function....

 Open the catalog to page 9
Series 1000 -10

table.main {} tr.row {} td.cell {} div.block {} div.paragraph {} .font0 { font:12.00pt "Times New Roman", serif; } .font1 { font:17.00pt "Times New Roman", serif; } 3» Dctailcd description of fonctions 3.1. Inleraal Oscillator lins tiTiclioii produces internai ly-tri ggered tlashes in Ihe 30 - 18.000 RPM or 0.5 -300 Hz range. Whcn Ihe fonction is first rccallcd il is sel foi a frequeucy of 1 Hz, corresponding lo 60 RPM, Other frequencies (RPMs) can be fed in directly with the numeric keys or changed with Ihe cursor keys. ® Kxample 1: setting a tuming frequency of 2000 RPM with the numeric keys...

 Open the catalog to page 10
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.