FPA

FPA

Product catalog summary
Electrical Data
The FPA-4 Flash Point Process Analyzer operates at a nominal voltage of 230 V AC ± 10%, with a maximum power consumption of approximately 500 W. It is classified under protection class IP 54 (NEMA 13). The operational ambient temperature ranges from +5 to +40 °C, while storage conditions allow for 0 to +60 °C. Humidity levels for operation are between 5 to 80% relative humidity, non-corrosive, and for storage, 5 to 85% relative humidity, non-corrosive.
Sample Requirements
The sample quality must be filtered to 50 µm, free of suspended water, bubble-free, with sulfur content less than 2000 ppm, and free of heavy metals and phosphate. Consumption is approximately 2 to 3 l/h at the sample inlet, with pressure requirements of 2 to 5 bar. The temperature at the inlet should be a minimum of 15 °C below the flash point and a maximum of +80 °C, with a temperature change not exceeding 1 K/min.
Utilities
Instrument air consumption is 8 Nm³/h during purging and approximately 1 Nm³/h during operation, with an inlet pressure of 2 to 7 bar. The coolant consumption varies depending on the flash point temperature, with sample as coolant requiring 30 to 60 l/h or plant cooling water needing 10 to 40 l/h. The temperature range for coolant is +5 to +40 °C, with an inlet pressure of 2 to 5 bar.
Signal Outputs and Inputs
The analyzer provides analog outputs for flash point temperature and digital outputs for alarm and readiness. It accepts digital inputs for stream selection, validation request, and reset. The electrical data for signal outputs and inputs include analog outputs of max. 8 (4 to 20 mA; 1000 Ω) and digital outputs of 24 V DC; max. 0.5 A.
Control Unit
The central control unit is an industrial PC running Windows Embedded Standard 7® with PACS control software. User interfaces include a TFT display with touch function and a virtual keyboard.
Connections and Dimensions
Tube fittings are available in Swagelok® 6 mm/12 mm/18 mm, with other fittings on request. The analyzer weighs approximately 200 kg and measures 1140 x 2000 x 710 mm, requiring 200 mm of space on each side.
Optional Interfaces
Optional interfaces include MODBUS/RTU via RS485 or RS422, MODBUS/TCP via FOC, and remote access via Ethernet.
Technical Data
The technology used is continuous measurement using catalytic combustion, correlating with standards such as ASTM D56, ASTM D93, and DIN EN ISO 2719. The measuring range is 25 to 180 °C, with repeatability and reproducibility adhering to DIN EN/ASTM standards.
Important Notice
The FPA-4 is subject to continuous product improvement, and specifications may change without notice. For technical clarification, contact the manufacturer.
See more

Catalog excerpts

FPA-1

Utilities Instrument air Consumption Purge: 8 Nm3/h while purging (~12 min) Operation: approx. 1 Nm3/h Pressure at inlet 2 to 7 bar (29 to 101.5 psi) Qualityhumidity class 2 or better acc. to ISO 8573.1 Coolant depends on flash point temperature Consumption sample as coolant: 30 to 60 l/h or plant cooling water: 10 to 40 l/h Temperature +5 to +40 °C (+41 to +104 °F) Certification TR CU available Other approvals and certificates, see www.bartec.de Technical data Technology continuous measurement using catalytic combustion Method correlates with: Repeatability s DIN EN/ASTM e.g. kerosene typ. 0.1 °C (approx. 0.2 °F) Reproducibility s DIN EN/ASTM Measuring cyclecontinuous Product streams 2 x sample, 1 x validation (additional hardware required) Signal outputs and inputs Analog outputs flash point temperature (others on request) Digital outputsAlarm, Ready/Valid Digital inputs Stream Selection, Validation Request, Reset Electrical data of signal outputs and inputs Analog outputs max. 8 (4 to 20 mA; 1000 Q) active isolated on request Analog inputs4 to 20 mA; 160 Q Digital outputs 24 V DC; max. 0.5 A Digital inputs high: 15 to 28 V DC; low: 0 to 4 V DC Auxiliary power 24 V DC; max. 0.8 A supply output Control unit_ Central control unit Industrial PC Operating system Windows Embedded Standard 7® Control software PACS Electrical data Nominal voltage 230 V AC ± 10 %, 1 phase; 50 Hz; other ratings on request Maximum approx. 500 W power consumption Protection class IP 54 (NEMA 13) Ambient temperature operation: +5 to +40 °C (+41 to +104 °F) storage: 0 to +60 °C (+32 to +140 °F) Ambient humidity operation: 5 to 80 % relative humidity, non-corrosive storage: 5 to 85 % relative humidity, non-corrosive Sample Quality filtered 50 gm, free of suspended water, bubble-free, sulfur < 2000 ppm, free of heavy metals, free of phosphate (s 37 cSt at inlet temperature) Consumption approx. 2 to 3 l/h (at sample inlet) Temperature at inlet min. 15 °C below Flash Point, max. +80 °C, temperature change maximum 1 K/min, For cooling with product: max. +40 °C For using an inductive ring-type initiator ("min. contact") on the flow meter: max. +60 °C User interfaces Display TFT display with touch function 1024 x 768 pixel Keyboard virtual keyboard, controlled via TFT display with touch function Connections Tube fittings Swagelok® 6 mm/12 mm/18 mm other fittings on request Vent/Drain open to atmosphere Dimensions (W x H x D) approx. 1140 x 2000 x 710 mm Space requirements right: 200 mm/left: 200 mm Optional interfaces_ Analog outputs on request MODBUS interface MODBUS/RTU via RS485 or RS422 or FOC is, MODBUS/TCP via FOC is Remote access via Ethernet (VDSL or FOC is) Important notice FPA-4 is subject to continuous product improvement, specifications are preliminary and may be subject to change without notice. If your technical data do not comply with existing data, please contact us for technical clarification. BARTEI BENKE Phone: +49 40 72703-0 [email protected] www.bartec-benke.de