CFPP

CFPP

CFPP

Product catalog summary
Overview: The Cold Filter Plugging Point Process Analyzer CFPP-4 by BARTEC BENKE is designed for the automatic determination of the cold filter plugging point (CFPP) of diesel and domestic fuels. This is crucial for ensuring fuel flow in cold temperatures, preventing clogging in fuel systems.
Application: The CFPP-4 allows diesel producers to optimize cold flow additives, enabling the use of winter-grade diesel below the cloud point. It features both stepped and linear cooling procedures and uses an identical test mesh filter as laboratory methods.
Special Features: The analyzer includes a plexiglass/glass measuring cell for visible function cycles, easy cell removal, and no paraffin adhesion due to preheated sample flushing. It offers exact cycle reconstruction as per ASTM D6371, integrated failure diagnosis, and multiple communication interfaces including Modbus and Ethernet.
Norms and Standards: The CFPP-4 complies with ASTM D6371, DIN EN 116, IP 309, and EN 16329 standards.
Technical Data: The analyzer operates within a measuring range of -35 to 15°C, with a cycle time of 25 to 90 minutes depending on CFPP temperature. It requires a nominal voltage of 230 VAC and has a maximum power consumption of approximately 700 W. The protection class is IP 54, suitable for ambient temperatures of 5 to 35°C.
Utilities and Connections: The instrument requires air consumption for purging and operation, with specific pressure and quality requirements. It includes various signal outputs and inputs for alarm, validation, and stream selection, and supports remote access via Ethernet.
Explosion Protection: The CFPP-4 is marked for ATEX, NEC 500, NEC 505, and CEC Sec. 18 standards, ensuring safety in explosive environments.
Additional Offerings: BARTEC GROUP also provides fast loop systems, sample conditioning systems, validation systems, and more, making them a comprehensive partner for system solutions in the oil and gas industry.
See more

Catalog excerpts

CFPP-1

Process Analyzer Cold Filter Plugging Point Process Analyzer CFPP-4 Credible Solutions for the Oil and Gas Industry

 Open the catalog to page 1
CFPP-2

Process Analyzer Cold Filter Plugging Point Process Analyzer CFPP-4 To remain competitive, today’s refiners must employ all optimization and product control techniques available. The use of online physical property analyzers is one of the key features to reach those objectives because they measure important quality properties in the process directly. The cold filter plugging point (CFPP) is the lowest temperature at which diesel and domestic heating fuels will pass through a filter in a given time when cooled under certain conditions. CFPP is supposed to give an estimate for the lowest temperature...

 Open the catalog to page 2
CFPP-3

Special Features:  Visible function cycles by using a measuring cell made of plexiglass/glass  Optimized assembly – easy removal of complete cell  No paraffin-adhesions on test mesh filter by flushing with preheated sample  No correlative measurement, but exact reconstruction of cycles as described in ASTM D6371  Identical test mesh filter as used in laboratory method  Possibility to shorten cycle time by: - Switching between summer and winter setting - Reading cloud point value (if available)  Integrated failure diagnosis and self monitoring  Available communication interfaces: - Modbus/RTU,...

 Open the catalog to page 3
CFPP-4

Cold Filter Plugging Point Process Analyzer CFPP-4 man* BENKE EXPLOSION PROTECTION Marking ATEX: II 2 G IIC T4 Gb NEC 500: Class I, Div. 2, Groups B, C, D, T4 NEC 505: Class I, Zone 1, AEx IIB+H2 T4 CEC Sec. 18: Class I, Zone 1, Ex IIB+H2 T4 Signal outputs and inputs Analog outputs Cold Filter Plugging Point (others on request) Digital outputs Alarm, Ready / Valid Digital inputs Stream Selection, Validation Request, Reset TECHNICAL DATA Technology Method Measuring range Repeatability Reproducibility Measuring cycle Electrical data of signal outputs and inputs Analog outputs plugging...

 Open the catalog to page 4
CFPP-5

Cold Filter Plugging Point Process Analyzer CFPP-4

 Open the catalog to page 5
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.