1. Catalogs
  2. Balun Technology Co., Ltd
  3. BL673 PORTABLE BLACKBODY CALIBRATION SOURCE

BL673 PORTABLE BLACKBODY CALIBRATION SOURCE

BL673 PORTABLE BLACKBODY CALIBRATION SOURCE

Product catalog summary
Features
  • Suitable for most IR Guns
  • Temperature range from -20℃ to +125℃
  • 65 mm aperture diameter with end closed tube
  • Emissivity greater than 0.995
  • Fast heat up and cool down times
  • High temperature stability of 0.15℃
  • Lightweight and portable design
  • Nitrogen purge connection available
  • External reference sensor input
Typical Applications
  • Infrared temperature sensors
  • Infrared thermal imaging systems
  • Spectrographic analyzers
  • Spectral radiometers
  • Heat flux meters
Specifications
  • Temperature range: -20℃ to 125℃
  • Aperture diameter: 65 mm
  • Heated emitter shape: End closed tube
  • Emissivity: > 0.995
  • Display resolution: 0.01℃ (<100℃); 0.1℃ (>100℃)
  • Control stability: 0.1℃
  • Temperature uniformity: 0.15℃
  • Display accuracy: 0.5℃
  • Warm-up time: 20 minutes to 125℃; 30 minutes to -20℃
  • Control method: Digital PID controller
  • Temperature sensor: Precision Platinum RTD
  • Tube depth: 110mm
  • Power supply: 220VAC, 350VA (110VAC optional)
  • Interface: RS485
  • Calibration band: 8um-14um spectral band
  • Dimensions: 300mm(H) x 200mm(W) x 280mm(D)
  • Net weight: Approx. 8.3 kg
Scope of Delivery
  • BL673 Portable blackbody calibration source
  • Power cord (plug type confirmation needed)
  • User manual
Optional Accessories
  • Accredited calibration certificate (P/N: 9000-673)
  • Customized range -23℃ to 155℃ (P/N: BL673-X)
Ordering Model Numbers
  • BL673-110V
  • BL673-220V
Reference Standards
  • IEC/TS 62492-1-2008: Industrial process control devices - Radiation thermometers
  • ASTM E2847-2013: Calibration and accuracy verification of wideband infrared thermometers
  • ASTM E1256-2011a: Test Methods for radiation thermometers (Single waveband type)
  • OIML D24 (1996): Total radiation pyrometers
  • JJG 856-2015: Verification regulation of radiation thermometers
See more

Catalog excerpts

BL673 PORTABLE BLACKBODY CALIBRATION SOURCE-1

BL673 PORTABLE BLACKBODY CALIBRATION SOURCE SPECIFICATION The BL673 is a portable blackbody calibration source utilizing a digital indicating temperature from -20℃ to 125℃ . A precision platinum RTD temperature sensor is embedded in the blackbody emitter, providing high accuracy and repeatability. The BL673 is as easy to use as “point and shoot.” Simply set the desired blackbody temperature from the convenient front panel control buttons, wait a few minutes for equilibrium, and point the gun at the target. The radiated energy from the blackbody is measured by your IR thermometer. Simply compare its readings to the display on the blackbody and record the difference. Ideal for most of IR Guns! Subzero -20℃ to +125℃ 65 mm aperture diameter end closed tube Emissivity is better than 0.995 Fast Heat Up & Cool Down Times 0.15℃ high temperature stability Light weight yet rugged design for portability Nitrogen purge connection available External reference sensor input Infrared temperature sensors Infrared thermal imaging systems Spectrographic analyzers Spectral radiometers Heat flux meter Balun Technology Co., Ltd Add: 1322, F/3, Bldg 1, 33 Zique Rd, Haidian, Beijing100095, China Temperature range Aperture diameter Heated emitter shape Emissivity Display resolution Control stability (10min) Temperature uniformity Display accuracy Warm up time (ambient temperature < 25℃ ) Control method Temperature sensor Tube depth Power supply Interface Calibration band Dimension Net weight -20℃ to 125℃ 65 mm End closed tube > 0.995 0.01℃ (<100℃ );0.1℃ (>100℃ ) 0.1℃ 0.15℃ 0.5℃ 20 minutes from ambient to 125℃ ; 30 minutes from ambient to -20℃ Digital PID controller with readout Precision Platinum RTD 110mm from entrance to target 220VAC, 350VA (110VAC is option) RS485 8um-14um spectral band 300mm(H)x 200mm(W)x 280mm(D) Approx. 8.3 kg BL673 Portable blackbody calibration source Power cord (need to confirm the plug type in advance) User manual Accredited calibration certificate (P/N: 9000-673) Customize range -23℃ to 155℃ , (P/N: BL673-X) ORDERING MODEL NUMBER ■ ■ IEC/TS 62492-1-2008:Industrial process control devices -Radiation thermometers. Part 1: Technical data for radiation thermometers ASTM E2847-2013: Standard practice for calibration and accuracy verification of wideband infrared thermometers ASTM E1256-2011a: Standard Test Methods for radiation thermometers (Single waveband type) OIML D24 (1996): Total radiation pyrometers JJG 856-2015: Verification regulation of ra

 Open the catalog to page 1

All Balun Technology Co., Ltd catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.