Overview: The ZUZI UV/VIS double beam spectrophotometer model 4260/50 is an advanced optical equipment designed for precision measurement. It features a double beam system with two detectors for simultaneous sample and reference measurement, enhancing measurement accuracy.
Specifications: The spectrophotometer operates within a wavelength range of 190-1100 nm and offers variable bandwidth options (0.5/1/2/4/5 nm) to balance resolution and precision. It includes a 5-inch LCD display and supports USB and parallel port connections. The device is powered by AC 110/220 V and weighs 22 kg.
Operation Modes: The equipment supports various modes including quantitative analysis, kinetics, scanning, multiple wavelength analysis, and DNA/protein analysis. It is suitable for applications in pharmaceuticals, biochemistry, quality control, research, and clinical laboratories.
Software: The UV/Vis Analyst software, compatible with Windows, provides comprehensive control over the spectrophotometer and enhances data storage capabilities. It supports up to 20 standard solutions for concentration curve establishment and offers multiple curve fitting methods.
Built-in Functions: The spectrophotometer allows for up to 10 wavelengths for multi-component analysis, real-time absorbance vs. time graph display, and post-run data manipulation including re-scaling and curve tracking. DNA and protein concentrations are calculated using specific absorbance ratios.
Technical Features: The device boasts a photometric range of 0-200% T, a photometric accuracy of ±0.3% T, and a stability of 0.0003 A/h at 500 nm. It includes tungsten and deuterium lamps and a standard 10 mm cell holder.
Optional Accessories: A range of accessories is available to extend the spectrophotometer's capabilities, including various cell holders, a Peltier module, and sipper systems.
Contact Information: For further details, contact AUXILAB, S.L. in Spain or ROGO-SAMPAIC in France.
See more

Catalog excerpts

double beam UV/VIS 4260/50 ZUZ-1

double beam UV/VIS 4260/50

 Open the catalog to page 1
double beam UV/VIS 4260/50 ZUZ-2

UV/VIS double beam spectrophotometer [H] OPTICAL EQUIPMENTS ample holder Spectrophotometer model 4260/50 is an advanced equipment with double beam op- Model 4260/50 presents and excellent performance in the measurement range 190- tical system with two detectors that measure sample and reference respectively and 1100 nm with a variable bandwidth (0.5/1/2/4/5 nm) to maximize the balance bet- simultaneously for optimizing precision of measurement. ween resolution, accuracy and precision in each application and to adjust to the This equipment presents different operation modes for performing quantitative...

 Open the catalog to page 2
double beam UV/VIS 4260/50 ZUZ-3

UV/VIS double beam spectrophotometer [H] UV/VIS double beam spectrophotometer Built-in functions in spectrophotometer screen [01] Control of equipment 1| Built-in software allows total control of equipment without the need of software for performing functions such as photometry, quantitative, wavelength scan, kinetics, DNA/Protein analysis and multiple wavelength analysis. 2| All information is displayed in easy-to-read display including the spectrum scan and kinetics and concentration curves. screen [02] Quantitative analysis (standard curves) 1| Up to 10 standard solutions may be used to establish...

 Open the catalog to page 3
double beam UV/VIS 4260/50 ZUZ-4

UV/VIS double beam spectrophotometer [H] OPTICAL EQUIPMENTS UV/VIS double beam spectrophotometer Built-in functions in spectrophotometer screen [04] Wavelength scan 1| The wavelength scan interval can be selected among the following values 0.1, 0.2, 0.5, 1, 2 or 5 nm 2| High, medium and low scan speeds are available; scan speeds vary from 100 to 3000 nm/min. 3| Wavelengths are scanned from high to low so instrument stands-by at high wavelength and this minimizes the degradation of UV sensitive samples. 4| Precise control of filter and lamp changes means that their effects are not seen on the...

 Open the catalog to page 4
double beam UV/VIS 4260/50 ZUZ-5

UV/VIS double beam spectrophotometer [H] UV/VIS double beam spectrophotometer UV/Vis Analyst software (PC-control software) The PC application software UV/Vis Analyst supplied together with pands data processing and storage capacities. spectrophotometer 4260/50 allows total control of instrument through computer and increases its built-in functions as well as exscreen [01] Quantitative analysis (standard curves) 1| Up to 20 standard solutions may be used to establish concentration equation curve. 2| There is a choice of 4 methods for fitting curve through the calibration points: linear fit, linear...

 Open the catalog to page 5
double beam UV/VIS 4260/50 ZUZ-6

UV/VIS double beam spectrophotometer [H] OPTICAL EQUIPMENTS UV/VIS double beam spectrophotometer UV/Vis Analyst software (PC-control software) screen [04] Kinetics 1| This mode may be used for scanning time course or reacting rate calculations. Absorbance vs Time graph is displayed on screen in real time. 2| Wait time, measurement time and time intervals may be entered. 3| Post-run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by entered factor....

 Open the catalog to page 6
double beam UV/VIS 4260/50 ZUZ-7

UV/VIS double beam spectrophotometer [H] UV/VIS double beam spectrophotometer technical spcifications Wavelength range Wavelength accuracy Wavelength repeatability Spectral bandwidth Photometric range Photometric accuracy Optical system Double beam, grating 1200 lines/mm Scan speed High, medium, low (max. 3000 nm/min) Stray light Baseline flatness Standard cell holder 10 mm single cell holder (sample and reference) 4 glass cells and 2 quartz cells included Light source Tungsten and Deuterium lamps (pre-aligned) USB type A for USB memory device (right side) USB type B for PC connection (back side)...

 Open the catalog to page 7
double beam UV/VIS 4260/50 ZUZ-8

UV/VIS double beam spectrophotometer [H] OPTICAL EQUIPMENTS UV/VIS double beam spectrophotometer Optional accessories The following accessories can be used with spectrophotometer 4260/50 to increase its features and application range. Contact us for more information about these accessories: [01] 4- cell holder for up to 50 mm path length cells [02] 4- cell holder for up to 100 mm path length cells [03] Micro cell holder [04] Solid sample holder [05] Test tube holder [06] Water jacketed 4-cell holder for 10 mm path length cells; it requires water bath and circulation system (not included) [07]...

 Open the catalog to page 8
double beam UV/VIS 4260/50 ZUZ-9

AUXILAB, S.L. Polígono Morea Norte Calle D Nº 6 C.P 31191 [Beriain] Navarra .: T: +34 948 310 513 [email protected] www.auxilab.es ROGO-SAMPAIC 23 Rue Ampère, ZI de Vilemilan 91325 Wissous (Francia) T: +33 1 69 53 67 67 [email protected] www.rogosampaic.com

 Open the catalog to page 9
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.