Overview: The AVC-010 series oxygen transmitter is designed for monitoring indoor air quality by measuring oxygen levels in various environments such as offices, indoor parking, libraries, hospitals, and shopping malls. Low oxygen levels can lead to hypoxemia and chronic fatigue, affecting organ function and causing toxin accumulation in the bloodstream. The transmitter helps in efficient air quality management by sending data to a monitoring center for analysis or ventilation control.
Specifications:
  • Measurement Principle: Electrochemical
  • Measurement Range: 0 ~ 30%
  • Coverage Area: 80m3
  • Accuracy: Full scale ± 5% at 25℃
  • Resolution: 0.01%
  • Repeatability: Reading ± 5%
  • Response Time: τ90 ≦15 sec
  • Drift: <2% /month
  • Power Consumption: RS485: 0.2W (Max.); 4~20mA/2~10V, RS485: 0.4W (Max.); LoRa / RS485: 1.2W (Max.)
  • Warm-up Time: 1 minute
  • Power Supply: DC 12 ~36V, AC 24V (50/60Hz)
  • Signal Output: DC 4~20mA/2~10V; RS485 (Modbus RTU)/ LoRa (Peer to peer)
  • Radio Frequency: 862~932 MHz
  • Linearity Error: Reading ± 8% (at -10℃~ 50℃)
  • Calibration Interval: Recommended every 6 months
  • Sensor Lifespan: 1 year
  • Pressure Range: 1 atm ± 10%
  • Storage Temperature: -10~+50℃ (14~122℉)
  • Environmental Conditions: -10~+50℃ (14~122℉); 0~95%RH (non-condensing)
  • IP Rating: IP30 (No protection from liquids)
  • Case Material: Fireproof ABS
  • Certifications: CE and FCC
Installation and Usage Recommendations:
  • Install at a height of 1~2M above the floor level.
  • Consider individual coverage area based on layout and enclosed spaces.
  • Install at least one device per room.
  • Calibration is recommended every 6 months, or quarterly in extreme conditions.
  • Sensor lifespan may vary with environmental conditions, such as high ambient temperatures.
Model Variants: AVC-010□□□-M with options for wall-mount installation, RS-485, LoRa, and display configurations.
See more

Catalog excerpts

Aecl Oxygen Transmitter (O2 sensor) AVC-010 Series-1

Oxygen Transmitter AVC-010 series The oxygen level is a very important indication of indoor air quality. It exists generally, in gaseous state at room temperature. When people stays in some places with low oxygen level, such as office, commercial building or basement, it might cause hypoxemia. Chronic fatigue is a major sign that people isn’t getting enough oxygen. Without enough oxygen in the lungs, the organs cannot function properly. As a result, toxins accumulate in the blood stream. AVC-010 Oxygen transmitter can monitor oxygen level and send the data to monitoring center for analysis or for ventilation control. Air quality can be, therefore, efficiently monitored and managed. Easy installation and the recommended height of installation is 1~2M above the floor level. It is perfect for household application, offices, indoor parking, library, hospital, shopping mall and etc.

 Open the catalog to page 1
Aecl Oxygen Transmitter (O2 sensor) AVC-010 Series-2

Measurement principle Measurement range Coverage area Response time Power consumption Warm-up time Power supply DC 12 ~36V, AC 24V (50/60Hz) DC 4~20mA/2~10V; RS485 (Modbus RTU)/ LoRa (Peer to peer) radio frequency: 862~932 MHz(3) Signal output Linearity error Calibration interval (recommended) Sensor lifespan Pressure range Storage temp. IP30 (No protection from liquids) Case material The coverage area should be considered individually to account for the particular layout and any varying or enclosed spaces. It is recommended to install at least one device in each room. The sensor drift will very...

 Open the catalog to page 2

All Autotronic Enterprise Co., Ltd. (Aecl) catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.