Overview
The CDF60 is a compact, handheld calibrator designed for checking the calibration of leak and flow ATEQ instruments, as well as other pressure measuring devices. It is ideal for both field and laboratory use, offering high accuracy and repeatability in a portable format.

Applications
  • Leak Detection: Calibration checking on leak and flow testers.
  • Flow Control: Suitable for other types of pressure measuring instruments.

Key Features
  • Pocket-sized with a user-friendly display.
  • Equipped with a Li-ion rechargeable battery.
  • USB interface for connectivity.
  • High accuracy in measurements.

Technical Specifications
  • Measurement Ranges: Atmospheric pressure and temperature indicators.
  • Flow Measurement Ranges: 4 ml/min and 40 ml/min with respective accuracies of 2% Rdg + 0.01 ml/min and 2% Rdg + 0.1 ml/min.
  • Operating Temperature: +10°C to +45°C; Storage Temperature: 0°C to +60°C.
  • Power Supply: Converts 120 to 240 V AC to 24 V DC.

Additional Features
  • Automatic sleep mode and battery level check.
  • Print out capability via USB or RS232.
  • LCD display with 4 x 20 characters.
  • 1-year warranty.

Accessories
  • Carrying case, needle valve, soft case, protective holster.
  • USB cable for PC connection.

Compliance
The CDF60 meets air quality standards as per ISO 8573-1 and can be calibrated by an ISO/IEC 17025 laboratory for enhanced accuracy.
See more

Catalog excerpts

Leak/Flow calibrator CDF 60-1

COMPACT HANHELD LEAK/FLOW CALIBRATOR The CDF60 calibrator checks the calibration of leak and flow ATEQ instruments as well as any other instrument. This light, compact and user friendly calibrator is essential for field or laboratory checks, when accuracy and repeatability cannot be compromised and large instrument are too bulky or too expensive. With the CDF60, you can now easily check and follow the requirements of ISO or other quality standards. Highlights POCKET SIZE AND USER FRIENDLY DISPLAY LI-ION RECHARGEABLE BATTERY → USB INTERFACE → HIGH ACCURACY → → Applications LEAK DETECTION Calibration checking on Leak and Flow testers FLOW CONTROL Checking of other types of pressure measuring instr

 Open the catalog to page 1
Leak/Flow calibrator CDF 60-2

COMPACT HANHELD LEAK/FLOW CALIBRATOR Measurement ranges Main features FLOW MEASUREMENT Range Rdg = Reading value Accuracy, in ambiant conditions, defined as non-linearity, hysteresis, repeatability, temperature effect and long term drift. * Specification in option (calibrated by our ISO/IEC 17025 laboratory) • Atmospheric pressure and temperature indicators • Automatic sleep mode • Print out (USB or RS232) • Battery level check • LCD display (4 x 20 characters) • 1 year warranty • Accessories: Carrying case, needle valve, soft case, protective holster... Technical specifications Air supply Temperature...