The ZE 3000 sensor emits a 4...20 mA signal based on precise magnet angle detection, allowing retrofitting of existing flowmeters for electrical signaling and integration with PLCs for process control. It is pre-programmed for specific DFM flowmeters to ensure accurate measurement.
The DFM model must be known prior to fitting the ZE 3000. The percentage scale is based on H2O at 20°C unless specified otherwise.
The ZE 3000 is mounted on the flowmeter's dovetail guide using screw clamps. Ensure the notch aligns with the halfway mark of the flow capacity. Follow the plug assignment plan for wiring. Maintain minimum spacings of 200 mm for DFM 165-200 and 400 mm for DFM 335-350.
After installation and power supply, press the concealed button for at least 2 seconds to compensate for magnetic influences, ensuring the float is at the bottom position.
Flow data sensor ZE 3000 for DFM 165–350 Features •• controlled by microprocessor to ensure accurate flow measurement and evaluation •• 2-wire technology •• 4...20 mA analogue output •• individual programming adapted to the DFM •• 11-point calibration •• nonvolatile storage of the values •• 0 button to compensate for the surrounding magnetic influences •• factory settable inertia (low-pass filter) 0.1…2.5 sec. according to customer specifications •• Recording accuracy of the magnetic float higher than 0.5% ASV Stübbe GmbH & Co. KG Hollwieser Straße 5 · 32602 Vlotho, Germany Phone: +49 (0) 5733-799-0 · Fax: +49 (0) 5733-799-5000 Email: [email protected] · Internet: www.asv-stuebbe.com We reserve the right to make technical changes. Issue 20
Open the catalog to page 1Flow data sensor ZE 3000 for DFM 165–350 Description •• The ZE 3000 emits a 4 ... 20 mA signal according to industrial standards, generated in the DFM by the precise magnet angle detection of the magnetic float. Existing flowmeters (with magnetic floats) can be easily retrofitted with the ZE 3000 to allow electrical signalling of the flow and, for example, further processing of the values in a PLC for process control or direct display of the flow rate. This unit controlled by micro-processor is programmed by ASV Stübbe for the respective DFM flowmeter. This guarantees accurate flow measurement...
Open the catalog to page 2Flow data sensor ZE 3000 for DFM 165–350 Flowmeter With Float DFM concealed button Once the system is installed and the voltage supply established, press the button (X) for at least 2 seconds to compensate for the magnetic ambient influences. Ensure that the float is in its bottom position during this, i.e. no flow takes place! Note Never press the button during normal/standard operation. ASV Stübbe GmbH & Co. KG Hollwieser Straße 5 · 32602 Vlotho, Germany Phone: +49 (0) 5733-799-0 · Fax: +49 (0) 5733-799-5000 Email: [email protected] · Internet: www.asv-stuebbe.com We reserve the right...
Open the catalog to page 3Basic valve with hand lever ASV Stubbe GmbH & Co. KG Hollwieser Strafe 5 • 32602 Vlotho, Germany Phone: +49 (0) 5733-799-0 • Fax: +49 (0) 5733-799-5000 4 Email: [email protected] • Internet: www.asv-stuebbe.com We reserve the right to make technical changes. Issue 2016.06.14-en Print No. 300742 TR MA DE Rev001