MODEL1019-P-Datasheet

MODEL1019-P-Datasheet

Product catalog summary
Technical Specifications
  • Alarms: Two user-configurable O2 alarms with options for latching, override during Span & Zero Calibration, and Sensor Fail Alarm.
  • Gas Connections: Available in 1/8" or 1/4” compression tube fittings or 1/8” female NPT.
  • Menu Controls: Features membrane switch with menu-driven selection for analysis range, display resolution, signal filter strength, temperature compensation, calibration, alarm setpoints, and system functions.
Optional Accessories/Equipment
  • AST-222-OP sensor for analysis up to 100% Oxygen.
  • RS232 or RS485 Communication.
Advanced Sensor Technology
  • Proprietary electrolyte formulation and anode material.
  • Electro-etched smooth cathode.
  • Recovery from air to 0.1% in under 30 seconds.
  • Extended operating temperature range of -10°C (-4°F).
  • Z-Series sensor compatibility with 0-100% CO2.
  • Fast sensor response with no routine maintenance needed.
Key Analyzer Features
  • Model: 1019-P % Oxygen Analyzer.
  • Available in rack and panel mount configurations.
  • ISO 9001:2015 Quality System certified.
  • Power: 100-240 VAC Universal AC/DC 5V Power Adapter, can be powered via USB.
  • User-selectable range from 0.5% to 25%.
  • Output locking on selected range.
  • Isolated 4-20mA and 0-1 VDC signal output standard.
  • Fine-tuned temperature and barometric pressure compensation.
  • Sensor Fail Alarm.
Performance Specifications
  • Response Time: 90% of step change in less than 30 seconds.
  • Repeatability: +/-1% over all ranges.
  • Recommended Sample Flow Rate: 1-2 SCFH, vent to atmosphere.
  • Sample System: Includes SS 316 sensor housing with compressing tube fittings gas connectors.
  • Sensor Model: AST-222 for inert gases, AST-222-Z for gases containing more than 0.5% CO2.
  • Expected Sensor Life: 36 months at 25ºC and 1 atm.
  • Operating Temperature Range: 5ºC to 45ºC.
  • Enclosure: Painted SS 7” x 4.5” x 4” panel mount.
  • Warranty: 12 months for both analyzer and sensor.
  • Measurement Range: User-selectable range from 0-25%.
  • Display: Bright 2-Line LED.
  • Output Lock: 4-20 mA locked on user-selectable range.
  • Accuracy: Less than 2% of FS at constant operating conditions.
  • Linearity: +/-1% after calibration.
  • Temperature Compensation: Fine-tuned on low % range for enhanced stability.
  • Barometric Pressure Compensation: Optional.
  • Calibration Method: Certified Span Gas or air, recommended calibration interval of 30 days.
Contact Information
See more

Catalog excerpts

MODEL1019-P-Datasheet-1

APPLIED SENSING TECHNOLOGIES Advanced Sensor Technology Proprietary Electrolyte Formulation and Anode Meterial Electro-etched Smooth Cathode Recovery From Air to 0.1% in under 30 Seconds Extended Operating Temperature Range -10°C (-4°F) Z-Series sensor Compatibility with 0-100% CO2 Fast Sensor Response, no Routine Sensor Maintenance Needed MODEL 1019-P % Oxygen Analyzer Rack and Panel Mount Configurations User Selectable Range From 0.5% to 25% Output Locking on Selected Range Isolated 4-20mA and 0-1 VDC Signal Output Standard Fine Tuned Temperature Compensation Barometric Pressure Compensation Sensor Fail Alarm 100-240 VAC Universal AC/DC 5V Power Adoptor Can be Powered via USB Analyzer Technical Specifications * Measurement Range: User selectable range from 0-25% • Display: Bright 2-Line LED • Output Lock: 4-20 mA locked on user selectable range. • Accuracy: < 2% of FS at constant operating conditions • Linearity: +/-1% after calibration • Temperature Compensation: Fine tune on low % range for enhanced stability • Barometric Pressure Compensation: Optional • Calibration Method: Certified Span Gas or air, Recommended Calibration interval 30 days • Alarms: Two user configurable O2 alarms with option for latching, Override Alarms during Span & Zero Calibration, Sensor Fail Alarm • Gas Connections: 1/8" or 1/4” compression tube fittings or 1/8” female NPT • Menu Controls: Membrane Switch, Menu driven selection of Analysis Range, Display Resolution, Signal Filter Strength, Fine Tune Temp Compensation, Calibration, Alarm Setpoints and System Functions • Response Time: 90% of Step Change in <30 sec • Repeatability: +/-1% over all ranges • Recommended Sample Flow Rate: 1-2 SCFH, vent to atmosphere • Sample System Included: SS 316 Sensor Housing with Compressing Tube Fittings Gas Connectors • Sensor Model: AST-222 inert gases, AST-222-Z for gases containing more than 0.5% CO2 • Expected Sensor Life: 36 months @ 25ºC and 1 atm • Operating Temp: Range: 5ºC to 45ºC • Enclosure: Painted SS 7” x 4.5” x 4” Panel Mount • Warranty: 12 months analyzer,12 months sensor Optional Accessories/Equipment AST-222-OP sensor for analysis up to 100% Oxygen RS232 or RS485 Communication *Specification validated during desing, subject to change without notice ISO 9001:2015 Quality System (909) 517 003 7 info@app liedsensing.com www.appliedsensin

 Open the catalog to page 1

All AST - Applied Sensing Technologies catalogs and technical brochures

  1. SRX-CA32-RTS

    1  Page

  2. SRX-CA32-4

    1  Page

  3. SRX-CA60-4

    1  Page

  4. SRZ-CA24-4

    1  Page

  5. SRX-CTE2

    1  Page

  6. SRX-CA60-RTS

    1  Page

  7. SRX-CT22D

    1  Page

  8. SRX-CT33

    1  Page

  9. SRX-CT23

    1  Page

  10. SRX-CT21

    1  Page

  11. SRZ-MA333

    1  Page

  12. SRZ-MA333-M

    1  Page

  13. SRZ-MA100-M

    1  Page

  14. SRX-MA100-M

    1  Page

  15. SRZ-MA100-4

    1  Page

  16. SRX-MGP-41

    1  Page

  17. SRX-MA333-M

    1  Page

  18. MODEL 1015PW

    1  Page

  19. MODEL 1019P

    1  Page

  20. MODEL1015-PW

    1  Page

  21. MODEL 1015

    1  Page

  22. SRX-MTL-2C

    1  Page

  23. Model 1012

    1  Page

  24. SRX-CT17A

    1  Page

  25. SRX-CT21A

    1  Page

  26. SRX-CT22A

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.