OPM200
9Pages

{{requestButtons}}

Catalog excerpts

OPM200 - 1

Optoelectronics & Engineering Services Fast Optical Power Monitor HIGHLIGHTS • Maximum input power • Noise equivalent power (NEPRMS) • Small signal rise time (10-90%) • Wide selection of photodiodes • Fast signal analysis • Component testing • Quality control • OEM (eg: laser noise monitoring) MEASUREMENT PRINCIPLE The OPM200 amplifier employs high quality photodiodes and precision transimpedance input stages to provide for low offset and high linearity throughout the full dynamic range. The OPM200 series is insensitive to electromagnetic interference by design, an important factor when working in „dirty“ industrial environments. These units are provided in OEM-style enclosures. The case wings provide for mounting on standard 25mm and 1“ optical table tops and for OEM applications. These optical power monitors are particularly useful for the measurement of rapidly changing optical signals. These units have a particularly high bandwidth of 10MHz allowing monitoring of high frequency noise components of many light sources. The output is a voltage linearly proportional to input power. The fast response time at high signal-noise-ratio makes the OPM200 series particularly useful in systems control feedback loops. Artifex Engineering Dortmunder Str. 16-18 26723 Emden, Germany mail: info@afx-eng.com http://www.afx-eng.com

Open the catalog to page 1
OPM200 - 2

ABSOLUTE MAXIMUM RATINGS Average Current Temperature Range ORDERING INFORMATION Full order code: Options ptions Case style OEM style with gullwing mounts Number of channels (n): 1 to 4 For example, a 2 channel, OEM style unit with a silicon (Si) photodiode input would be ordered as: OPM200GS2 For customized systems, please contact us.

Open the catalog to page 2
OPM200 - 3

SPECIFICATIONS Si, InGaAs PARAMETER INPUT Wavelength range extended InGaAs InGaAs Si UV-Si Power ranges (full scale) Polarization Dependant Loss (PDL) Fibre type Noise equivalent power (NEPRMS) single mode, multi-mode (φ core ≤ 62.5 µm; NA ≤ 0.275) Linear analogue Vout = scale x Pin FC, FSMA, free beam Receptacles OUTPUT Function Output scale Output range (full scale) Rise / Fall time (10% - 90%) Output impedance POWER SUPPLY Type Wall plug (supplied) Dimensions DIMENSIONS APPLICATIONS EXAMPLES Typical applications may be found in laser power control, fluorescence lifetime measurements,...

Open the catalog to page 3
OPM200 - 4

CASE STYLES (Lab style)

Open the catalog to page 4
OPM200 - 6

THE FRONT AND BACK PANELS The front panel contains the input and output BNC connectors. The upper row of BNC connectors is for the current inputs. The channel count begins at the left. If the current source is a photodiode then the cathode is to be connected to the centre pin of the BNC. If a current source other than a photodiode is to be used, then the current drain is to be connected to the centre pin. The lower row of BNC connectors is for the amplifier outputs. The output is 0-4.5V, linearly proportional to the current input. The channel count begins at the left. The back panel...

Open the catalog to page 6
OPM200 - 7

INSTRUCTIONS FOR MEASUREMENT The OPM200 optial power monitor comprises one to four independent measurement channels, depending on the model chosen. Each channel consists of a receptacled optical input and a BNC-output. These two user interfaces are arranged in logical groups (see „The Front and Back Panels“). To make a measurement, proceed as follows: 1. Turn the unit on. For the most accurate measurement, please allow for a 15 minute warm up before using. 2. Connect an optical source to the desired channel via the corresponding receptacle. Ensure that the lug of the fibre connector is...

Open the catalog to page 7
OPM200 - 8

DAMAGE The unit may be damaged by exceeding the maximum average input power. Please read „Absolute Maximum Ratings“ for these maximum values before working with the instrument. Use only the power supply and power supply cable provided with the unit. TROUBLESHOOTING In the event that a measurement is not successful, the following possibilities should be analysed: Symptom Possible Errors Ensure the power cord is connected at both ends and switch the system on. Increase input power Input or output connection not correct Output at full scale, independant of input current • Input power too high...

Open the catalog to page 8
OPM200 - 9

NOTICE Artifex Engineering reserves the right to make changes to its products or to discontinue any product or service without notice and advises customers to obtain the latest version of relevant information to verify, before placing orders, that information being relied on is current and complete. All products are sold subject to the terms and conditions of sale supplied at the time of order acknowledgement, including those pertaining to warranty, patent infringement and limitation of liability. Customers are responsible for their applications using Artifex Engineering components. Artifex...

Open the catalog to page 9

All Artifex Engineering e.K. catalogs and technical brochures

  1. LDD100

    8 Pages

  2. LGM100

    4 Pages

  3. LIV100

    29 Pages

  4. LIV110_REV5C

    10 Pages

  5. TZA600

    9 Pages

  6. LDM300

    5 Pages

  7. IRS100 2009

    5 Pages

  8. TZA500

    35 Pages

  9. TZA400

    12 Pages

  10. TZA300

    12 Pages

  11. OPM500

    39 Pages

  12. OPM400

    10 Pages

  13. OPM300

    15 Pages

  14. Filters

    1 Pages

  15. OPM100

    6 Pages