Overview: The DA85 Differential Pressure Manometer is designed for measuring pressure, under pressure, or differential pressure of non-aggressive gases. It features a diaphragm element and an analogue scale, with a compact size of approximately 70 mm.
Specifications:
  • Measurement Range: Customer-specific presetting required, with ranges from 0...200 Pa to 0...6000 Pa.
  • Operating Medium: Air or non-aggressive gases.
  • Measuring Principle: Diaphragm measuring system.
  • Temperature Range: Ambient: -10...+50 °C, Storage: -25...+60 °C.
  • Conformity: Complies with RoHS directive 2011/65/EU.
Physical Characteristics:
  • Housing: Made of ABS.
  • Dimensions: Diameter 70 mm, depth 50 mm.
  • Weight: Approximately 90 g.
  • Protection Class: IP65 according to EN 60529.
  • Display: Pointer display with a scale length of 300° (approx. 130 mm).
  • Pressure Connections: Hose connections 5 mm ø.
Accuracy and Error Limits:
  • Zero Drift: ± 0.75 %.
  • Linearity and Hysteresis: < 400 Pa ± 3 %; > 400 Pa ± 2 %.
  • Temperature Drift: Zero point: ± 0.3 % / 10 K, Measuring range: ± 0.2 % / 10 K.
Performance:
  • Overload Protection: 0.2 bar.
  • Static Pressure: Max. 0.2 bar.
  • Zero Point Calibration: Adjustable by REED contact, no cyclic calibration required.
  • Reaction Time: Direct.
Accessories: Includes a mounting set with screws, bleeder connections, and a 2 m plastic hose (4 x 1.5 mm). Optional limit value pointer in red or green.
See more

Catalog excerpts

data sheet DA85-1

DIFFERENTIAL PRESSURE MANOMETER DA85 □ Diaphragm element □ Smallest measuring range 0...200 Pa □ Analogue scale □ Compact size of about ca. 70 mm Manometer with diaphragm element for indication of pressure, under pressure or differential pressure of non-aggressive gas. Overload protection: Static pressure: Zero point calibration: Reaction time: PERFORMANCE: 0.2 bar max. 0.2 bar Adjustable by REEDcontact, no cyclic zero-point calibration required direct Zero drift: Sum of linearity and hysteresis: Temperature drift zero point: Temperature drift measuring range: Operating mode: Measuring medium: Measuring principle: Measuring unit: Smallest measuring range: Largest measuring range: Measuring range: Measuring range selection: Ambient temperature: Storage temperature: ACCURACY/ERROR LIMIT: ± 0.75 % < 400 Pa ± 3 %; > 400 Pa ± 2 % ± 0.3 % / 10 K ± 0.2 % / 10 K TECHNICAL SPECIFICATIONS: Measuring mode Air or non-aggressive gases diaphragm measuring system Pa 0...200 Pa (2 mbar) 0...6000 Pa (60 mbar) 0...200 Pa (2 mbar), 0...400 Pa (4 mbar), 0...1000 Pa (10 mbar), 0...2000 Pa (20 mbar), 0...4000 Pa (40 mbar), 0...6000 Pa (60 mbar) factory preset -10...+50 °C -25...+60 °C Housing: Dimensions: Weight: Protection class: Display: Scaling: Pressure connections: PHYSICALLY: ABS diameter 70 mm, depth 50 mm approx. 90 g IP65 according EN 60529 Pointer display with scale Scale length 300° = approx. 130 mm Hose connections 5 mm 0 CONFORMITY: RoHS: according RoHS-directive 2011/65/EU Measurement range: Customer-specific presetting required! Probe measuring range ± on request Mounting set with screws, bleeder Connections and 2 m plastic hose (4 x 1.5 mm) 25111 1. Limit value pointer red or green - Please choose! 25201 Arthur Grillo GmbH • Am Sandbach 7 • 40878 Ratingen • phone 0 21 02 - 47 10 22 • fax 0 21 02 - 47 58 82 www.grillo-messgeraete.de/en • Email: [email protected] • web shop: www.sensor-store.de

 Open the catalog to page 1

All Arthur Grillo GmbH catalogs and technical brochures

  1. SPECIAL

    18  Pages

  2. TEMPERATURE

    20  Pages

  3. Display

    14  Pages

  4. HUMIDITY

    21  Pages

  5. PRESSURE

    35  Pages

  6. WR 200-F

    2  Pages

  7. SHT11/15

    1  Page

  8. Orifice plate

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.