video corpo

FitoClima 300 CB - Cold bend testing chamber
12Pages

{{requestButtons}}

Catalog excerpts

FitoClima 300 CB - Cold bend testing chamber - 1

TESTING TESTA CT/TT 300 CR ENVIRONMENTAL CHAMBER FOR COLD BEND / COLD IMPACT TESTING OF WIRES & CABLES

Open the catalog to page 1
FitoClima 300 CB - Cold bend testing chamber - 2

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS ARALAB is a company specialised in designing, developing, manufacturing and servicing of high quality climatic chambers and controlled environment rooms. Since 1985 we have been perfecting ways to create and control temperature, humidity, light, air flow and many other environmental conditions. Only the highest quality components are used to manufacture our chambers so customers can have the best equipment for their research and testing purposes. Control the environment, Your own climate. TESTA temperature and humidity testing chambers offer...

Open the catalog to page 2
FitoClima 300 CB - Cold bend testing chamber - 3

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS TEMPERATURE AND HUMIDITY CONTROL RANGES TESTA TT CHAMBERS – TEMPERATURE ONLY TESTA TT CHAMBERS TEMPERATURE RANGE HUMIDITY RANGE TESTA CT CHAMBERS – TEMPERATURE AND HUMIDITY TESTA CT CHAMBERS TEMPERATURE RANGE HUMIDITY RANGE TEMPERATURE AND HUMIDITY CONTROL RANGES TESTA CT TESTING CHAMBERS Performance in CLIMATIC testing range | only TESTA CT chambers TEMPERATURE RANGE TEMPERATURE UNIFORMITY HUMIDITY RANGE Performance in TEMPERATURE testing | TESTA TT and TESTA CT chambers TEMPERATURE RANGE TEMPERATURE RATE OF CHANGE HEATING (2a) (2b) TEMPERATURE...

Open the catalog to page 3
FitoClima 300 CB - Cold bend testing chamber - 4

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS DIMENSIONS AND DRAWINGS TESTA TT / TESTA CT 300 EXTERNAL DIMENSIONS (HxWxD) (mm) INTERNAL DIMENSIONS (HxWxD) (mm) MINIMUM CLEARANCE MAINTENANCE AREA 2660 Standard refrigeration system is air cooled Services hub installation needs: Supply power ECP75: 400VAC, 50Hz, 50A / 3-Phase + Neutral + Ground ½” demineralized water supply Electrical protection: Circuit breaker 3 x 63A + N with 300mA differential 50mm water drain at floor level 3. Electrical cabinet installation needs: Water cooled option: 230VAC, 50Hz, 16A / Single Phase + Neutral + Ground...

Open the catalog to page 4
FitoClima 300 CB - Cold bend testing chamber - 5

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS EQUIPMENT DESCRIPTION TEMPERATURE TEMPERATURE PRECISION (in the interior of the chamber, at 5 cm from walls, floor and top) • In Time ≤ ± 0,5ºC • In Space ≤ ± 1,5ºC TEMPERATURE SENSORS • One (1) PT 100 Class A, located in air treatment tunnel • One (1) PT 100 Class A, movable sensor for flexible placing inside chamber HEATING • By tubular stainless steel electric heaters located in the air treatment tunnel COOLING • Air cooled hermetic scroll compressor group (low noise and high efficiency) with enforced ventilation and without CFC’s....

Open the catalog to page 5
FitoClima 300 CB - Cold bend testing chamber - 6

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS • Automatic stop function in case of water failure, with indication on the controller; High / Low Temperature alarms; High / Low humidity alarms. Interior: AISI 304 hermetical welded, vapour tight, stainless steel Exterior: Zinc mild steel with epoxy coating finish (color RAL 7035) Insulation: Rock Wool Interior illumination: Halogen lamp 12V (only available with optional window) Door: Double silicone joints and anti-condensation heating frames (optional window) Reinforced Floor Plate • Air Flow: Forced through 1 ventilator/fan. • Air Renovation:...

Open the catalog to page 6
FitoClima 300 CB - Cold bend testing chamber - 7

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS PROGRAMABLE WINDING MECHANISM WITH 6 MANDRELS ENVIRONMENTAL CHAMBER FOR COLD BEND / COLD IMPACT TESTING OF WIRES & CABLES

Open the catalog to page 7
FitoClima 300 CB - Cold bend testing chamber - 8

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS CONTROLLER CLIMA PLUS Programmable PLC exclusively developed for ARALAB chambers Easy to use coloured Touch-Screen Display Interface Resolution of 0,1ºC for Temperature and 0,1% for Relative Humidity High performance temperature and humidity control with value correction in all ranges Capability for creating 50 programs of 50 segments each Internal non volatile memory for storing test data Automatic restart of tests due to power failure, without losing data and restarting test where it was interrupted Real-time monitoring of all functions and...

Open the catalog to page 8
FitoClima 300 CB - Cold bend testing chamber - 9

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS FITOLOG SOFTWARE The FitoLog software pack is a set of applications designed to facilitate the managing, monitoring and recording of programs and data from the TESTA chambers. It consists of 3 applications: FitoLog, FitoLogView and FitoProgram. FITOLOG Records and displays in real time all data and details related to the set-points, running variables and equipment behaviour. It also retrieves information about the active components of the chamber, running processes, errors, alarms and allows the configuration of periodic or alarm triggered remote...

Open the catalog to page 9
FitoClima 300 CB - Cold bend testing chamber - 10

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS COMMON ACCESSORIES PLEASE CONSULT ARALAB FOR OTHER ITEMS Software pack FitoLog and FitoLogView Observation anti-condensation windows in multi layered glass Water demineralizer (for TESTA CT Chambers) Integrated water tank Air dryer Additional entry ports Calibration certificate from accredited external laboratory Faster heating / cooling temperature change rates up to 10k/minute WINDOW OPTION The observation window is composed of a multilayered glass with optimum levels of thermal insulation. The interior and exterior glasses have a heating...

Open the catalog to page 10
FitoClima 300 CB - Cold bend testing chamber - 11

TESTA - TEMPERATURE & ENVIRONMENTAL CHAMBERS INSTALLATION REQUIREMENTS INSTALLATION SITE The place should be easily accessible, according to equipment dimensions and weight. It should have good air circulation and a room temperature between 10º and 26ºC. The floor should be leveled and a minimum distance of 50cm from the walls and other equipment must be kept. ELECTRICAL SUPPLY Near the equipment with the specified requirements. HUMIDIFICATION CIRCUIT AND DEMINERALIZED WATER (FOR TESTA CT MODELS) The humidification circuit works exclusively with distilled or demineralized water. For this...

Open the catalog to page 11
FitoClima 300 CB - Cold bend testing chamber - 12

Let’s meet! aralab@aralab.pt www.aralab.pt T: +351 219 154 960 Control the environment Your own climate

Open the catalog to page 12

All Aralab catalogs and technical brochures