Curing chambers - 600/1200 PHCI - 1

TESTING FITOCLIMA S600 / D1200 PHCI ENVIRONMENTAL TESTING CHAMBERS FOR CURING APPLICATIONS (MORTAR, CEMENT, LIME AND PLASTER)

Open the catalog to page 1
Curing chambers - 600/1200 PHCI - 2

ARALAB is a company specialized in designing, developing, manufacturing and servicing of high quality climatic chambers and controlled environment rooms. Since 1985 we have been perfecting ways to create and control temperature, humidity, light, air flow and many other environmental conditions. Only the highest quality components are used to manufacture our chambers so customers can have the best equipment for their research and testing purposes. Control the environment, your own climate. KEY FEATURES • Wide environmental performance ranges, enabling greater control of climatic conditions •...

Open the catalog to page 2
Curing chambers - 600/1200 PHCI - 3

TECHNICAL SPECIFICATIONS AND PERFORMANCES FITOCLIMA 600 PHCI TEMPERATURE RANGE TEMPERATURE PRECISION TEMPERATURE FLUCTUATION TEMPERATURE UNIFORMITY HUMIDITY RANGE HUMIDITY PRECISION HUMIDITY FLUCTUATION HUMIDITY UNIFORMITY 4 shelves with 530 mm x 630 mm (up to 8 can be fitted) ELECTRICAL SUPPLY FITOCLIMA 1200 PHCI TEMPERATURE RANGE TEMPERATURE PRECISION TEMPERATURE FLUCTUATION TEMPERATURE UNIFORMITY HUMIDITY RANGE HUMIDITY PRECISION HUMIDITY FLUCTUATION HUMIDITY UNIFORMITY 8 shelves with 530 mm x 630 mm (up to 16 can be fitted) ELECTRICAL SUPPLY TEMPERATURE & HUMIDITY WORKING RANGE...

Open the catalog to page 3
Curing chambers - 600/1200 PHCI - 4

DIMENSIONS AND DRAWINGS FITOTERM / FITOCLIMA 600 EXTERNAL DIMENSIONS (HxWxD) (mm) INTERNAL DIMENSIONS (HxWxD) (mm) Electrical cabinet installation needs: Standard refrigeration system is air cooled Services hub installation needs: Supply power: 230VAC, 50Hz, 16A / Single Phase + Neutral + Ground ½” demineralized water supply Electrical protection: Circuit breaker 16A + N with 300mA differential 50mm water drain at floor level ENVIRONMENTAL TESTING CHAMBERS FOR CURING APPLICATIONS (MORTAR, CEMENT, LIME AND PLASTE

Open the catalog to page 4
Curing chambers - 600/1200 PHCI - 5

DIMENSIONS AND DRAWINGS FITOTERM / FITOCLIMA 1.200 EXTERNAL DIMENSIONS (HxWxD) (mm) INTERNAL DIMENSIONS (HxWxD) (mm) Standard refrigeration system is air cooled Services hub installation needs: ½” demineralized water supply 50mm water drain at floor level Electrical cabinet installation needs: Supply power: 230VAC, 50Hz, 16A / Single Phase + Neutral + Ground Electrical protection: Circuit breaker 16A + N with 300mA differential RJ45 communication port ENVIRONMENTAL TESTING CHAMBERS FOR CURING APPLICATIONS (MORTAR, CEMENT, LIME AND PLASTE

Open the catalog to page 5
Curing chambers - 600/1200 PHCI - 6

EQUIPMENT DESCRIPTION CONSTRUCTION Highly resistant stainless steel in the interior and exterior Polyurethane insulation Reinforced stainless steel shelves (up to 40Kg weight load) Front panel with zincor steel and gray epoxy paint Pivoting door(s) with spring lock, magnetic gasket and safety lock 4 or 6 casters (600 or 1200 model, respectively) with built in brakes for effortless mobility and control 35mm diameter side porthole New generation multi-color touch-screen ClimaPlus© controller Open door alarm with configurable time-out function Free slots for connecting and integrating external...

Open the catalog to page 6
Curing chambers - 600/1200 PHCI - 7

CLIMAPLUS CONTROLLER • Programmable Logic Controller exclusively developed by Aralab for FitoClima chambers • Easy to use touch-screen interface • 190mm x 150mm multicolor display • Controls every environmental variable available for any specific FitoClima model (Temperature, Humidity, Lights, Airflow, CO2, and connected external devices) • Friendly program editor for creating 32 programs of 24 segments each, allowing the design of complex and comprehensive climatic simulation programs • Password protection of the controller functions • Content and research protection feature, with...

Open the catalog to page 7
Curing chambers - 600/1200 PHCI - 8

FITOLOG SOFTWARE The FitoLog software pack is a set of applications designed to facilitate the managing, monitoring and recording of programs and data from the FitoClima chambers. It consists of 3 applications: FitoLog, FitoLogView and FitoProgram. FITOLOG Records and displays in real time all data and details related to the set-points, running variables and equipment behaviour. It also retrieves information about the active components of the chamber, running processes, errors, alarms and allows the configuration of periodic or alarm triggered remote notifications (by email or SMS,...

Open the catalog to page 8
Curing chambers - 600/1200 PHCI - 9

ACCESSORIES OPTIONAL ACCESSORIES FitoLog / FitoView software for data monitoring, logging and programs configuration on the PC Ethernet cable for PC connection CO2 Controlling and monitoring system Wall mounting conductivity meter, for water quality control, with assembling accessories Water Demineralizer 20 liter water tank with electric pump and security valve Glass door with double glazing Additional Stainless steel wire shelves Reinforced shelves for heavy samples ACCESSORY FOR ACCELERATED RADIATION AGING 8 fluorescent and UV lamps ACCESSORY FOR CARBONATION CO2 Controlling and...

Open the catalog to page 9
Curing chambers - 600/1200 PHCI - 10

Let’s meet! aralab@aralab.pt www.aralab.pt T: +351 219 154 960 Control the environment Your own climate