BOD Measurement System BD 600 Accurate, automatic & direct control of your wastewater samples Biochemical Oxygen Demand (BOD) Respirometric BOD measurement using BD 600 BOD – biochemical oxygen demand – is an expression for the quantity of oxygen required for biological degradation of organic matter in a waste water sample. BOD measurement is therefore used as a basis for the detection of biologically degradable organic matter in water. The difference between BOD and chemical oxygen demand (COD) is that COD additionally registers biologically non-degradable organic matter. The AQUALYTIC® sensor system BD 600 is a 6 sample system that allows precise measurements of BOD based on the manometric principle. Manometric respirometers relate oxygen uptake to the change in pressure caused by oxygen consumption while maintaining a constant volume. Due to the modern integral pressure sensors, it is not necessary to use mercury for the measurements. BOD measurement is an important measurement of the effects of domestic and industrial waste water on sewage plants and outflow points. ■ Large, illuminated & brilliant graphic display ■ Remote control ■ User-selectable time span from 1 to 28 days ■ Graphical representation of measured values ■ Free individual programming of each of the six samples ■ USB & SD Card data transfer ■ Inductive stirring system, 100 - 240 V / 50 - 60 Hz Regularly updated information: www.aqualytic.de
Open the catalog to page 1Measuring ranges and sample volumes The BOD level of a sample depends on the quantity of organic matter present, which can vary considerably. The BOD measuring system BD 600 is therefore calibrated for the various sample volumes and the corresponding measuring ranges listed in the table below. The overall measuring range of the system is 0 – 4000 mg/l. For all measuring ranges, BOD is shown directly in mg/l. Range mg/l BOD Pressure sensor BD 600 Principle Respirometric methods provide direct measurements of the oxygen consumed by microorganisms from an air or oxygenenriched environment in a...
Open the catalog to page 2BOD-Measurement Evaluation of measurements The BD 600 measuring system records a measurement once every hour, independent of the length of the measuring period. This way the quality of the series of measurement can be evaluated in an early stage. Current values and stored values may be called up at any time. Stored value can be displayed numerically or graphically. The table/graph below illustrates an example of BOD5 evaluation. The development of BOD over a period of five days is easily seen. Day Display 1. day Automatic start function Variations in sample temperature prior to testing...
Open the catalog to page 3BOD-Measurement Technical data Meas. principle Manometric; mercury-free; electronic pressure sensor Applications BOD5, BOD7, OECD 301 F ... Display Measurement period User-selectable, between 1 and 28 days Auto result storage Up to 672 results, depending on measurement period Storage – hourly (1 day) interval every 2 hours (2 days) – – daily (3-28 days) Automatic start function – After temperature equalisation of samples – Can be switched off Power supply 3 alkaline-manganese batteries (“Baby” cells/size “C”) or via power supply unit using y-cable togehter with stirring unit USB host port...
Open the catalog to page 4All AQUALYTIC catalogs and technical brochures
-
Turbidity meter AL400T-WL
1 Pages
-
General Catalogue
143 Pages
-
Reagents
22 Pages
-
Jar Tester
2 Pages
-
checkit-comparator
12 Pages
-
Powder Dispenser PD250
2 Pages
-
COD Set-Ups COD VARIO
2 Pages
-
AL200
4 Pages
-
AL400 & AL410
4 Pages
-
SD 310 Oxi
2 Pages
-
AL15
2 Pages
-
Comparator 2000+
20 Pages
-
CHECKIT ® Comparator
12 Pages
-
Laboratory Refrigerator
2 Pages
-
Incubator
2 Pages
-
Turbidity meter AL250T-IR
1 Pages
-
Thermoreactor AL125
1 Pages
-
Photometer AL450 series
4 Pages
-
Photometer AL100
4 Pages
-
MicroDirect
2 Pages
-
Hand-Held Meters AL10 Series
2 Pages
-
AQUALYTIC Test-Kits
2 Pages
-
Minikit
2 Pages
Archived catalogs
-
General catalog
136 Pages