UV Online Analyzer for COD BOD TOC TSS NO3 STAC2 - 1

STAC2: On-line UV-Vis Alert Analyser for on-line water chemical quality monitoring of BOD / COD / TOC / TSS and Nitrates ◼ Fingerprint capability: scanning from 190 up to 800nm with unique 1 nm bandwidth ◼ Versatile sampling: 4 channels and 10 parameters available per channel ◼ Data connectivity: both local and cloud based data collection by means of a WiFi connection ◼ Real-time visualization of the water’s fingerprint: BOD / COD / TOC / Nitrates / Pesticides / Aromatic Hydrocarbons ◼ Secomam UV-Vis technology since 1966! ◼ Robust and rugged design with high-tech data handling Areas of application: ◼ Wastewater treatment plant (follow-up purification performance, input, discharge, etc.) ◼ Drinking water plant (pumping station, tank management, etc.) ◼ Natural water ◼ Process water ◼ River monitoring ◼ Aquaculture STAC2 Solution for Water Quality Monitoring: Water is largely transparent to ultraviolet and visible light. The STAC2 analyser measures the absorbance of ultraviolet light, at a broad range of wavelengths. Organic content of the sample stream absorbs UV light, and thus the measurement of UV absorbance provides an adequate indication of the organic concentration in the sample. It thereby creates a unique fingerprint of the water. The STAC2 absorbance signal can be correlated using different chemical methods such as BOD/COD/TOC and nitrates by applying a mathematical adjustment factor or gradient to the measured value. If the water source changes, the correlation gradient may also change. The STAC2 has the flexibility to use 4 different streams and to combine this with up to 10 different models per sample stream. As UV absorbance may be affected by the turbidity content of water, the STAC2 analyser does compensate for this. Monitoring your real-time data was never been so simple! Datasheet : STAC2 Update : February 2020 FicheTechnique_SCAF_v1 AQUALABO 90 Rue du Professeur Paul Milliez 94500 CHAMPIGNY SUR MARNE Tel.: +33 (0)1.55.09.10.10- Fax : +33 (0)1.55.09.10.39 info@aquala

Open the catalog to page 1
UV Online Analyzer for COD BOD TOC TSS NO3 STAC2 - 2

DATASHEET STAC2 STAC2 Full SpectrumTechnology and Data Handling: The STAC2 scans and measures the entire absorption spectrum and creates a unique fingerprint from the sample. STAC technology is used by many customers worldwide as a vital decision making tool in their water quality monitoring. The STAC2 is the only analyser in the world that can measure and analyze a multitude of organic substances without using reagents. The STAC2 analyser is a technologically advanced instrument, able to be connected to an Ethernet or WIFI network. There is no screen on the instrument itself, and the human...

Open the catalog to page 2
UV Online Analyzer for COD BOD TOC TSS NO3 STAC2 - 3

STAC2 unique detector: Xenon lamp with scanning capability from 190 to 800nm with 1nm bandwidth. STAC2 unique sampling: • Automatic and sequential sampling circuit rinsing and cuvette cleaning • Automatic blank (reference sample) scanning to calculate and compare the absorbance spectra. • Programmable automatic sampling from 5 minutes to 6 hours • Pinching valves 0 8 mm for TSS reading without filtration • 4 different sample streams. STAC2 datamining & connectivity: The STAC2 analyser is a technologically advanced instrument that connects to an Ethernet or WIFI network. The Web platform...

Open the catalog to page 3
UV Online Analyzer for COD BOD TOC TSS NO3 STAC2 - 4

STAC2 STAC2 analyser specifications: Sample streams Datasheet : STAC2 Update : February 2020 AQUALABO 90 Rue du Professeur Paul Milliez 94500 CHAMPIGNY SUR MARNE Tel.: +33 (0)1.55.09.10.10- Fax : +33 (0)1.55.09.10.39 info@aqualabo.fr - www.aqualabo.fr

Open the catalog to page 4
UV Online Analyzer for COD BOD TOC TSS NO3 STAC2 - 5

The S200 displays the concentrations of the parameters measured by STAC2 and the models used to determine these concentrations. It also allows to recover data from 2 parameters on 4-20 mA outputs STAC2 & Secomam History: The STAC2 is the successor to STAC1 which was developed by Secomam in Ales, France. The success story started back in the 90’s as the “Station Alert Compact' (probably the 1st online UV analyser in the world!) in combination with the unique portable UV-Vis analyser (Pastel UV) and UV Pro modelling software. This unique set of instruments and software was way ahead of its...