Catalog excerpts

Autocorrelators Model Overview pulseCheck NX High-Resolution Multitalent page 3 pulseCheck SM 2000 Long-Range Autocorrelator page 9
Open the catalog to page 1
... Benefits & Technology Optics and Detector Technology A wide variety of professional applications can be covered with various detector types and optics sets. Photomultiplier (PMT) detectors are highly sensitive and are ideal for pulse measurements at low pulse energies. Photodiode detectors, on the other hand, are perfect for measuring lasers with high peak power. In combination with highly efficient optics, our detectors pave the way for measurements over an wide wavelength spectrum from 200 nm to 12 µm. Collinear and Non-Collinear Measurements Our autocorrelators support fast switching...
Open the catalog to page 2
pulseCheck NX The Modular Autocorrelator Pulse Measurement Perfection with the Multitalent from APE The pulseCheck platform grows with your tasks. Suitable for characterizing virtually all ultrafast pulse lasers, APE's pulseCheck autocorrelator covers a wide range of wavelengths and pulse widths. This flexibility is achieved by using interchangeable optics sets, typically consisting of a nonlinear crystal and a dedicated detector module. Exchangeable Optics Sets for wavelength coverage from 200 nm to 12 µm Wide range of sensitivity levels covered with PMT, PD, and TPA High temporal...
Open the catalog to page 3
Overview Specifications Pulse Width Refresh Rate Depending on scan range, 120 ps/sec Intensity Resolution Delay Resolution Detector Types Electronics Software Connectivity see page 14 pulsecheck SM 2000 for details High Resolution & Strong Processing Performance The advanced system-on-chip based architecture enables measurements with low latency, a high time resolution down to 50 attoseconds and an intensity dynamic range of 18 bit. This allows full access to even the tiniest autocorrelation features.
Open the catalog to page 4
... Flexibility for your Experiments Maximum Functionality through Modular Design APE meets the growing need for maximum functionality and flexibility with the modular concept on which the pulseCheck autocorrelator series is based. Thus, Optics sets to upgrade the wavelength range or FROG capability can be added any time. From Ultrashort to Longer Pulses Various pulseCheck configurations can be adapted to the individual needs of pulse duration measurement. The measurement of extra long pulse durations up to 500 ps is possible with pulseCheck SM. Pulse Width Measurement Range t High...
Open the catalog to page 5
Optics Sets A variety of different and exchangeable optics sets cover a wide wavelength range: from UV at 200 nm to mid IR at 12 µm. Photodiode (PD) ▪ Typ. Sensitivity: 1 W2 ▪ Rep. Rate: > 10 Hz ▪ Measurement Mode: Collinear and Non-collinear λ[nm] Photomultiplier (PMT) ▪ Typ. Sensitivity: up to 10-6 W2 ▪ Rep. Rate: > 250 kHz ▪ Measurement Mode: Collinear and Non-collinear λ[nm] Two Photon Absorption (TPA) ▪ Typ. Sensitivity: < 0.1 W2 (NIR/IR) ▪ < 1 W2 (VIS) ▪ < 100 W2 (UV) ▪ Rep. Rate: > 10 Hz ▪ Measurement Mode: Collinear Intensity λ[nm] * For cross-correlation: wavelength range depends...
Open the catalog to page 6
Specifications Measurable Pulse Width Range* 5 fs - 15 ps with pulseCheck NX 50; 10 fs - 40 ps with pulseCheck NX 150 Wavelength Range Exchangeable Detector (Optics Sets) Dealy Sampling Resolution Down to 50 attoseconds (NX 50), 200 attoseconds (NX 150) Refresh Rate Typically 1 ... 10-6 W2 depending on Optics Set** Intensity Resolution Pulse Repetition Rate Type of Measurement Mode PMT, PD : non-collinear intensity, collinear interferometric; TPA: hybrid non-collinear intensity, collinear interferometric; SHG Tuning for Phase Matching PMT/PD: automatic; TPA: not applicable Trigger Mode...
Open the catalog to page 7
Options ▪ Various Optics Sets ▪ Fiber coupling ▪ Polarization rotator ▪ FROG Compliance ▪ CE certified ▪ NIST tracable calibration with HeNe laser ▪ Fully compliant with DIN 58175-2:2021-04 (Autocorrelation measurement methods) Dimensions and Power Dimensions 244 x 201 x 309 mm (pulseCheck NX 50) 244 x 201 x 395 mm (pulseCheck NX 150) See appendix for technical drawings
Open the catalog to page 8
pulseCheck SM 2000 Long-Range Autocorrelator Pulse Width Measurement up to 500 ps With the revised version of the long-range delay technology in pulseCheck SM, the new autocorrelator offers a scanning range up to three times larger and a measuring speed up to five times faster than the previous long-range models. The pulseCheck SM includes a high performance controller that is integrated into the device. The pulseCheck SM is suitable for particularly long pulse durations of up to 500 ps. At the other end, short pulses down to 20 fs can be measured. A resolution of 1 fs and a sampling rate...
Open the catalog to page 9
Specifications Measurable Pulse Width* 20 fs ... 500 ps (20 fs - 150 ps directly measured, 150 ps - 500ps half ACF measured) Wavelength Range Optics Sets Delay Resolution Delay Linearity Scan Speed Sampling Rate Typically 1 ... 10-6 W2 depending on Optics Set Recommended Repetition Rate Type of Measurement Mode PMT, PD : non-collinear intensity, collinear interferometric; TPA: hybrid collinear intensity Trigger Mode Input Polarization Linear horizontal, vertical available as option Input Beam Coupling Free-space (6 mm aperture); Option: fiber coupling FC/PC, FC/APC, SMA 1 W (e.g. oscillator...
Open the catalog to page 10
Appendix Technical Drawings pulseCheck NX Multitalent for any task Beam distance Beam distance Beam input (turnable mirror) Beam input (turnable mirror) Focus alignment Focus alignment Beam distance Long-range autocorrelator Beam input (turnable mirror) Plauener Str. 163-165|Haus N|13053 Berlin|Germany T: +49 30 986 011-30 E: sales@ape-berlin.de APE Angewandte Physik & Elektronik GmbH © APE GmbH | Rev. 4.1.1 | Specifications are subject to change without notice.
Open the catalog to page 11All A.P.E catalogs and technical brochures
-