Overview: The ProCeas® ATEX/IECEx H2S Gas Analyzer is designed for measuring hydrogen sulfide (H2S) in natural gas, with optional capabilities to measure CO2, CH4, C2H6, and H2O. It utilizes advanced laser technology and patented OFCEAS TDL technology for precise and reliable gas analysis.
Specifications:
  • Measurement Range: H2S: 0-50 ppm (typical), 0-1000 ppm (max); CO2: 0-5% vol (typical), 0-20% vol (max); H2O: 0-50 ppm (optional), 0-500 ppm (max); CH4: 0-100% vol; C2H6: 0-10% vol (typical), 0-20% vol (max).
  • Performance: Linearity is less than 1% of reading, repeatability is 0.5% of reading ±6 ppb for H2S and H2O, response time is under 10 seconds, and drift is negligible.
Typical Stream Composition: The analyzer is suitable for streams with typical compositions of H2S at 10 ppm, CH4 at 80% vol, C2H6 at 3% vol, C3H8 at 1% vol, and CO2 at less than 5% vol.
Features and Benefits:
  • Continuous measurement with high-resolution laser technology.
  • No optical moving parts, reducing maintenance needs.
  • ATEX and IECEx certified for safety in explosive environments.
  • Direct measurement without sample degradation, eliminating the need for scrubbers.
  • Low pressure sampling system and no instrument air consumption.
  • Yearly maintenance, self-calibrating system, and ultra-precise measurements.
  • Fast response time and reduced operational costs.
Technical Data:
  • Power Supply: 110-230 VAC, 50-60 Hz, with a power consumption of 180 VA.
  • Ambient Conditions: Operates between -20 to +55 °C, with IP65 protection class.
  • Physical Info: Weighs less than 80 kg, dimensions are 600 x 510 x 308 mm.
  • Communication: Supports Ethernet, ModBus (TCP/IP, RS), analog, and USB outputs.
  • Sample Flow Rate: Greater than 2.4 slh.
See more

Catalog excerpts

ProCeas explosion-proof in Natural Gas-1

ProCeas® ATEX/IECEx H2S Gas Analyzer in Natural Gas CO2, CH4, C2H6, and H2O optional gases to measure

 Open the catalog to page 1
ProCeas explosion-proof in Natural Gas-2

ontinuous measurement C igh resolution laser technology H atented OFCEAS TDL technology P o optical moving parts A irect measurement without degrading the sample D (no scrubber needed) atented low pressure sampling system H elf-calibrating system (no span gases required) U ery fast response time V educed operational costs R (no gas cylinder, no compressed air, low power consumptions) igh availability of the system Technical data Range Typical stream composition Response time Drift zero / span Analyzer Technique OFCEAS TDL combined with LPS, the principles of measurement are covered by 2 patents...

 Open the catalog to page 2

Archived catalogs

  1. Epsilon 5000

    2  Pages

  2. ProCeas H2S

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.