Your specialist in innovating humidity & temperature sensors Relative Humidity & Temperature Transmitter Transmitter without LED display Transmitter with LED display AF series Capacitive humidity & temperature transmitter for Duct Mounting 1. Description: Aosong capacitive humidity element based humidity & temperature transmitters are ideal solution for detecting relative humidity and temperature in applications such as HVAC in residential & commercial buildings. The capacitance element based transmitter is one of the most stable transmitters available and no recalibration required, transmitters are temperature compensated and calibrated. Aosong transmitters are problem-free and accurate with long-term reliability by apply unique HydroProbe® digital technology, our transmitters are developed for high-precision measurement of relative humidity & temperature and transmitters are available for wall & duct & outdoor mounting. 2. Product highlights: Good-looking case, digital calibrated, long-term stability, standard industrial output, full range temperature compensated, wide detection range 3. Technical Specification: Power supply Humidity measuring range Humidity sensing element Output for humidity Accuracy of humidity Repeatability Hysteresis http://www.aosong.com Thomas Liu (Sales Manager) Email: thomasliu198518@aliyun.com , sales@aosong.com
Open the catalog to page 1Your specialist in innovating humidity & temperature sensors Sensitivity for humidity Drift rate per year Temperature sensor DS18B20 Accuracy of temperature +-0.3Celsius(at 25Celsius) Output for temperature 0~5V DC or 0~10V DC or 4~20mA Temperature Customer can select temperature range by dial-switches on PCB working range board: 0~50Celsius, -20~80Celsius, -40~60Celsius Storage temperature range -40..80Celsius Case ABS Electrical connection Screw connector( Max1.5mm2) 4. Ordering guide: AF3005 , output 0~5V DC for humidity & temperature both, without LED display AF3005Y, output 0~5V DC for...
Open the catalog to page 2Your specialist in innovating humidity A temperature sensors 1). Connection Diagram for voltage output Temperature- output 2). Connection Diagram for current output Temperature output Temperature error Error of humidity at different temperature Measure values maybe influenced by follow factors: Humidity errors Equilibration time too short, steam, water sprays, dripping water or condensation at the sensor, etc. Temperature errors Equilibration time too short, cold or hot outside wall, sunlights, heating elements, etc. Proper location of humidity sensor is important to ensure accurate...
Open the catalog to page 3Your specialist in innovating humidity & temperature sensors Avoid locating the sensor: * Near heat sources, such as radiant heat from the sun, heat from appliances, or from concealed pipe or chimneys * In areas subject to draft * Behind doors, draperies, or in corners * On walls having excessive vibration * In corrosive environments such as swimming pools or hospital rooms 9. Warnings: Don’t use this item on safety or emergency stop devices and any other application which may cause injury due to the failure of this item. 10. Quality Warranty: Aosong warrants that this product will be free...
Open the catalog to page 4All Aosong Electronics Co. catalogs and technical brochures
-
Data Sheet AHT21
16 Pages
-
Humidity sensor-AS108 AOSONG
4 Pages
-
Dew sensor-HDS10 AOSONG
3 Pages