Onsite micro-ohmmeter up to 10 A Designed for field use as well indoor as outdoor, OM 16 micro-ohmmeter performs 4-wire measurements of inductive and non-inductive resistances with continuous, pulse or AC measuring current up to 10 A. Offering a high accuracy of 0.05 % and resolution of 0.1 µΩ, it works over a large selectable range of resistances from 5 mΩ up to 2.5 KΩ.

 Open the catalog to page 1
OM 16-2

Description Designed for field use as well indoor as outdoor, OM 16 micro-ohmmeter performs 4-wire measurements of inductive and non-inductive resistances with continuous, pulse or AC measuring current up to 10 A. Offering a high accuracy of 0.05 % and resolution of 0.1 µΩ, it works over a large selectable range of resistances from 5 mΩ up to 2.5 KΩ. All parameters are user-programmable, either directly through the instrument interface or via Log OM software delivered in option: measuring current, range, unit, alarmes, automatic average function, absolute autozero. Before every measurement, EMFs...

 Open the catalog to page 2
OM 16-3

continuity, contact resistances, non-twisted cables… Automatic compensation of thermal E.M.F. Temperature compensation Metal temperature compensation

 Open the catalog to page 3
OM 16-4

Specifications Resistance measurement Measurement range Measuring current Voltage drop Automatic or manual selection of measurement range Accuracy given in % of reading + fixed value at 23°C ± 5°C Possible excess over the normal rating: 5 mΩ range: + 20 % 25 mΩ range: + 20 % Maximum voltage between the terminals in an open circuit: 7V. Further features Resistance types Inductive resistances: Coils, transformers, motor windings... Non-inductive resistances: Earth bonding, coating, contact resistances... Measurement current DC current from 1 mA to 10 A Continuous ou pulsed Measurement trigger conditions...

 Open the catalog to page 4
OM 16-5

Précision : ± 0,5°C (Tref = Température de reference programmée, à laquelle la mesure est ramenée) Temperature coefficient beyond operating range <10% accuracy/°C (from 0 to18°C and from 28 to 50°C) 2 programmable thresholds with visuel and sound signal Digital calibration without internal adjustment General specifications Size Batteries (option) Type: Ni/Mh 8.5 Ah Charging time: 5 h Battery life: 5,000 measurements at 10 A Communication ports Storage capacity 1,000 measurements identified by numbers Memory reading directly on the display, via software or printer Environmental specifications...

 Open the catalog to page 5
OM 16-7

Models and accessories Instrument: OM16 On-site micro-ohmmeter Clips and probes: Please note that 2 clips are needed per OM 16. AMT005 Long handspike, per unit Needle diameter: 3 mm, length without handle: 83 mm, total length: 215 mm Cable length: 5 m Large kelvin clip, per unit Opening diameter: 25 mm, cable length: 5 m Triggered handspike, per unit Needle diameter: 3 mm, length without handle 83 mm, total length: 215 mm Cable length: 5 m Triggering cable to be connected by RS 232 to OM 16 Small handspike, per unit Needle diameter: 3 mm, total length: 125 mm Cable length: 5 m Small kelvin clip,...

 Open the catalog to page 7
OM 16-8

COFRAC certificate of calibration Packing information: Shipping size