![AOIP - logo](https://img.directindustry.com/images_di/logo-p/L7451.gif)
Group: ASGARD
Catalog excerpts
![CALYS 80 IS - 1](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_1m.jpg)
ATEX field documenting calibrator with high accuracy 0.01% CALYS 80 IS is an ATEX documenting multifunction calibrator within CALYS IS range.
Open the catalog to page 1![CALYS 80 IS - 2](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_2m.jpg)
Description CALYS 80 IS is an ATEX documenting multifunction calibrator within CALYS IS range. It is the perfect tool for advanced process maintenance and use on test bench in hazardous and nonhazardous areas. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument. Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, CALYS IS series complies with both 21...
Open the catalog to page 2![CALYS 80 IS - 3](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_3m.jpg)
Specifications Specifications and performances in temperature @23°C ±5°C Uncertainty is given in % of reading (CALYS 80 IS display) + fixed value. Resistive probes: Measurement and simulation Sensor Range (Input and Output) Connections: 2, 3 and 4 wires Rtd simulation excitation current: from 0.1 to 4 mA without incremental error Rtd measurement excitation current: 0.2 mA Rtd cable compensation: up to 100 m (for each wire) Rtd cable compensation error (Pt100): ± 0.005°C/ of total wire Maximum load resistance: 600 Ω @ 20 mA Thermocouples: Measurement and simulation Type
Open the catalog to page 3![CALYS 80 IS - 4](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_4m.jpg)
Display units: °C, °F et K Resolution: 0.01°C / 0.01°F Temperature scale: ITS90 and IPTS68 Reference junction compensation: internal automatic from -10°C to +55°C or external Rj compensation drift: ± 0.002°C/°C (from -10 °C to +45 °C) - Class A Pt100 Input impedance: > 10 MΩ Specifications and performances in pressure @23°C ±5°C Pressure Pressure media: AISI 316 SS compatible fluids (water, gas, and oil) Temperature compensation: Automatic with built-in calibration matrix Engineering units: mbar, bar, hPa, kPa, Mpa, kg/cm2, kg/m2, psi, mmH2O, cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, FTH2O,...
Open the catalog to page 4![CALYS 80 IS - 5](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_5m.jpg)
Accuracy: ± 0.025% F.S. Overpressure: 125% F.S. Port: male, 1/4" BSP Connection wire length: 2 m Specifications and performances in process @23°C ±5°C DC voltage: Measurement and simulation Range Input impedance: < 10 MΩ for ranges up to 2 V > 500 kΩ for ranges up to 20 V Output impedance (emf output): < 0.5 Ω with a maximum current of 0.5 mA Output noise (at 300 Hz): < 2 μV pp for ranges up to 200 mV < 10 μV pp for ranges up to 2 V < 80 μV pp for ranges up to 20 V DC current: Measurement and simulation With or without loop supply Range
Open the catalog to page 5![CALYS 80 IS - 6](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_6m.jpg)
Input impedance: < 20 Ω at 1 mA Maximum load resistance: 600 Ω at 21 mA Loop supply: 12 V ± 5% Resistance: Measurement and simulation Type Connections: 2, 3 and 4 wires Source resistance effects: ± 1 μV error for 1000 Ω source resistance Simulation excitation current: from 0.1 to 4 mA without incremental error Measurement excitation current: 0.2 mA Maximum load resistance: 600 Ω @ 20 mA Frequency, pulse: Measurement and simulation Type Further features Scaling in measurement and simulation modes Setup of zero and span programmable within -399999 and +999999 Scaling allows process signals to...
Open the catalog to page 6![CALYS 80 IS - 8](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_8m.jpg)
Safety specifications Protections Thermal fuse High voltage suppressor Resistor-diode voltage limiter class II 1G Ex ia IIC T4 (-20°C from T Ambient to +50°C)
Open the catalog to page 8![CALYS 80 IS - 9](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_9m.jpg)
CALYS80 IS On-site ATEX documenting multifunction calibrator with high accuracy 0.01% Delivered in standard with: • User manual • Battery charger • Protection rubber holster • Factory test report Internal pressure sensors - AISI3165S - ± 0.025% FS: (2 IPM000100G -100 / 100 mbar (gauge) - Res. 0.001 mbar IPM000500G -500 / 500 mbar (gauge) - Res. 0.01 mbar External pressure sensors - AISI316SS - ± 0.025% FS: EPM000100G -100 / 100 mbar (gauge) - Res. 0.001 mbar EPM000500G -500 / 500 mbar (gauge) - Res. 0.01 mbar
Open the catalog to page 9![CALYS 80 IS - 10](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_10m.jpg)
External hand pump: F3280013 From -0.8 to 2 bar pressure pump From -0.95 to 40 bar pressure pump From 0 to 350 bar oil / water pump External 700 bar oil / water pump External 700 bar oil / water pump with high pressure hose External 1000 bar oil/water pump with high pressure hose 700 bar oil/water compact bench Electrical signal test leads Vinyl carrying case with accessories folder MiniDin Connectors Kit Tc J, K, T, S Female connector MiniDin Connectors Kit Tc J, K, T, S Male connector Software: DATACAL Calibration software for CALYS 75 / 100 / 150 Supplied with USB cable COFRAC...
Open the catalog to page 10![CALYS 80 IS - 11](https://img.directindustry.com/pdf/repository_di/7451/calys-80-is-524021_11m.jpg)
With all relevant data points where the device has been tested Packing information: Size Weight without packing
Open the catalog to page 11All AOIP catalogs and technical brochures
-
DRAGO 934
5 Pages
-
CALYS 1200
10 Pages
-
CALYS 1000
10 Pages
-
RAYOMATIC 16
7 Pages
-
RLV
5 Pages
-
EPM
10 Pages
-
UNIGAS 3000
7 Pages
-
RN 5306
5 Pages
-
LKS 1000 Leakshooter camera
6 Pages
-
TTI 10
6 Pages
-
RN 5360
5 Pages
-
MEGOHM 200
7 Pages
-
MF 28
6 Pages
-
IND 45
6 Pages
-
CALYS 1500
12 Pages
-
RAYOMATIC USB 2.2
6 Pages
-
RAYOMATIC PE
6 Pages
-
IRTEC P IVT
5 Pages
-
RAYOMATIC 16 2.2
6 Pages
-
RAYOMATIC NET
6 Pages
-
RAYOMATIC TC
6 Pages
-
ICO 51
6 Pages
-
LKS 1000
5 Pages
-
MILLIK
7 Pages
-
TM 6602 / 6612 / 6630
8 Pages
-
TC 6621 / 6622
10 Pages
-
SN 8310
7 Pages
-
CALYS 60 IS
9 Pages
-
CP 6632
7 Pages
-
THERMYS 150
10 Pages
-
CALYS 120 IS
12 Pages
-
ZX
5 Pages
-
OM 10
6 Pages
-
OM 21
8 Pages
-
ISOPALM 200
7 Pages
-
CRIT 100
4 Pages
-
IRtec P
11 Pages
-
FD 5 / 10 / 15
12 Pages
-
VNE20
5 Pages
-
OM 16
8 Pages
-
SATURN CYCLOP 878
4 Pages
-
CALYS 80P
10 Pages
-
P200 PRO
8 Pages
-
CALYS 150
8 Pages
-
motor starter
4 Pages
-
Frontdaq 10 isolated channels
4 Pages
-
Data Acquisition Systems DATALOG
12 Pages
-
Accuracy Thermometer PHP 602
2 Pages
Archived catalogs
-
2016 RAYOMATIC USB
6 Pages