Catalog excerpts

CALYS 120 IS-1

ATEX documenting multifunction calibrator with high accuracy 0.006% CALYS 120 IS is the top instrument of our range of ATEX documenting multifunction calibrators.

 Open the catalog to page 1
CALYS 120 IS-2

Description CALYS 120 IS is the top instrument of our range of ATEX documenting multifunction calibrators. It is the perfect tool for advanced process maintenance and use on test bench in hazardous and non-hazardous areas. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument. CALYS 120 IS does not only work as a simulator (IN / OUT) but also as a dual channel thermometer (IN / IN) to perform comparison calibration. Providing extended...

 Open the catalog to page 2
CALYS 120 IS-3

ATEX compliance: ATEX Ex II 1G EEx ia IIC T4 –20°C Tamb +50°C) X Switch test, leak test IN / IN: CALYS 120 IS has a dual channel input for comparison calibration

 Open the catalog to page 3
CALYS 120 IS-4

Specifications Specifications and performances in temperature @23°C ±5°C Uncertainty is given in % of reading (CALYS 120 IS display) + fixed value. Resistive probes: Measurement and simulation Sensor Range (Input and Output) Connections: 2, 3 and 4 wires Rtd simulation excitation current: from 0.1 to 4 mA without incremental error Rtd measurement excitation current: 0.2 mA Rtd cable compensation: up to 100 m (for each wire) Rtd cable compensation error (Pt100): ± 0.005°C/ of total wire Maximum load resistance: 600 Ω @ 20 mA Thermocouples: Measurement and simulation Type

 Open the catalog to page 4
CALYS 120 IS-5

Display units: °C, °F et K Resolution: 0.01°C / 0.01°F Temperature scale: ITS90 and IPTS68 Reference junction compensation: internal automatic from -10°C to +55°C or external Rj compensation drift: ± 0.002°C/°C (from -10 °C to +45 °C) - Class A Pt100 Input impedance: > 10 MΩ Specifications and performances in pressure @23°C ±5°C Pressure Pressure media: AISI 316 SS compatible fluids (water, gas, and oil) Temperature compensation: Automatic with built-in calibration matrix Engineering units: mbar, bar, hPa, kPa, Mpa, kg/cm2, kg/m2, psi, mmH2O, cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, FTH2O, mmHg,...

 Open the catalog to page 5
CALYS 120 IS-6

Accuracy: ± 0.025% F.S. Overpressure: 125% F.S. Port: male, 1/4" BSP Connection wire length: 2 m Specifications and performances in process @23°C ±5°C DC voltage: Measurement and simulation Range Input impedance: < 10 MΩ for ranges up to 2 V > 500 kΩ for ranges up to 20 V Output impedance (emf output): < 0.5 Ω with a maximum current of 0.5 mA Output noise (at 300 Hz): < 2 μV pp for ranges up to 200 mV < 10 μV pp for ranges up to 2 V < 80 μV pp for ranges up to 20 V DC current: Measurement and simulation With or without loop supply Range

 Open the catalog to page 6
CALYS 120 IS-7

Input impedance: < 20 Ω at 1 mA Maximum load resistance: 600 Ω at 21 mA Loop supply: 12 V ± 5% Resistance: Measurement and simulation Type Connections: 2, 3 and 4 wires Source resistance effects: ± 1 μV error for 1000 Ω source resistance Simulation excitation current: from 0.1 to 4 mA without incremental error Measurement excitation current: 0.2 mA Maximum load resistance: 600 Ω @ 20 mA Frequency, pulse: Measurement and simulation Type Further features Scaling in measurement and simulation modes Setup of zero and span programmable within -399999 and +999999 Scaling allows process signals to be...

 Open the catalog to page 7
CALYS 120 IS-8

Ramps generation Autoramp and Autostep capability with Start, End, and Step programmable parameters Single and continuous cycle with Start, End, Rises, Soaks, and Falls programmable parameters The signal value setting uses a unique in-line single-digit setting mode or a direct numeric entry Direct keypad access to n.10 programmable memory stored values Transmitter function CALYS IS can be used as a transmitter. Any input signal (electric or pressure) can be converted into a 4-20 mA output. The galvanic insulation between the input and output channels allows this function o be used on the prcess...

 Open the catalog to page 8
CALYS 120 IS-9

Resistor-diode voltage limiter Class class II 1G Ex ia IIC T4 (-20°C from T Ambient to +50°C)

 Open the catalog to page 9
CALYS 120 IS-10

CALYS120 IS On-site ATEX documenting multifunction calibrator with high accuracy Delivered in standard with: • User manual • Battery charger • Protection rubber holster • Factory test report Internal pressure sensors - AISI316SS - ± 0.025% FS: (2 IPM000100G -100 / 100 mbar (gauge) - Res. 0.001 mbar IPM000500G -500 / 500 mbar (gauge) - Res. 0.01 mbar External pressure sensors - AISI316SS - ± 0.025% FS: EPM000100G -100 / 100 mbar (gauge) - Res. 0.001 mbar EPM000500G -500 / 500 mbar (gauge) - Res. 0.01 mbar

 Open the catalog to page 10
CALYS 120 IS-11

External hand pump: F3280013 From -0.8 to 2 bar pressure pump From -0.95 to 40 bar pressure pump From 0 to 350 bar oil / water pump External 700 bar oil / water pump External 700 bar oil / water pump with high pressure hose External 1000 bar oil/water pump with high pressure hose 700 bar oil/water compact bench Electrical signal test leads Vinyl carrying case with accessories folder MiniDin Connectors Kit Tc J, K, T, S Female connector MiniDin Connectors Kit Tc J, K, T, S Male connector Software: DATACAL Calibration software for CALYS 75 / 100 / 150 Supplied with USB cable COFRAC certificate...

 Open the catalog to page 11
CALYS 120 IS-12

With all relevant data points where the device has been tested Packing information: Size Weight without packing 1.4 g

 Open the catalog to page 12
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.