SX

SX

SX

Product catalog summary
Introduction
The SX Level Transmitter offers superior signal stability, reducing the need for frequent recalibrations and preventing overflow or run-dry situations. It maintains 0.2% accuracy over a year and features a new field calibration option for easy adjustments without special tools.

Features
- Driftless transmitters reduce tune-ups and prevent overflows.
- Quick Disconnect Receptacles and One-Touch Field Calibration.

Specifications
Performance
- Level Measurement Ranges: SX5 (0-30” to 0-140”), SX6 (0-140.1” to 0-415”), SX7 (0-415.1” to 0-830”), SX8 (0-830.1” to 0-1385”).
- Accuracy: ±0.20% of URL, Repeatability: ±0.075% of URL, Hysteresis: ±0.075% of URL, Linearity: ±0.05% of URL.
- Temperature Limits: Process (-18°C to 130°C), Ambient (-9°C to 49°C).
- Power Signal: 12-40 vdc, Output: 4-20mA dc.

Materials/Construction
- Housing: 304 and 316 stainless steel.
- Wetted Parts: 316L stainless steel.
- Ratings: NEMA 4X, IP66, IP67.

Agency Approvals
- Intrinsically safe for Class 1, Div. 1, Groups A-D.
- CE Compliant, 3-A compliant.

Order Information
- Span Range: Options from 0-30” to 0-1385”.
- Fittings: Various options including Anderson, Cherry Burrell, King Gage, and Tank Mate.
- Diaphragm: SS-316L Electropolished or Hastelloy “C”.
- Mounting: Horizontal, Vertical, or other specified angles.
- Sensor Wiring: Quick Disconnect Receptacle options.

Accessories
- Gaskets and Sensor Calibration Adaptors available for different fittings.
- Cord Sets: Molded with various cable lengths.

Contact Information
Anderson Instrument Company, 156 Auriesville Road, Fultonville, NY 12072. Phone: 800-833-0081, Email: [email protected].
See more

Catalog excerpts

SX-1

SENSORS FOR FOOD AND LIFE SCIENCES. LIFE SCIENCES “SX” Level Transmitter Introduction This line of sensors provide superior signal stability. The field tested performance of the “SX” transmitter means the elimination of expensive, repetitive re-calibrations along with protection against costly overflow/run-dry situations that can occur with less stable level transmitters. The 0.2% accuracy experienced at installation can now be expected a year later! “Driftless” transmitters mean: We’ve also incorporated a new field calibration feature that provides “one-touch” sensor zeroing as well as simple range calibration, without the need for special tools and fixtures, pressure sources, or removal of the transmitter from the tank. This translates to real cost savings during installation or when adjusting for new application requirements. Special Features Include: Of course the “SX” Transmitter is available in all standard fitting and range combinations you’ll need to fit any sanitary tank application. All variations are CE compliant, meet Intrinsically Safe requirements, 3-A Sanitary Standards, and are backed by a 2-year warranty. Complete specifications and ordering information are available on the reverse. For more information please visit our Web Site at www.anderson-negele.com, or contact your local Authorized Anderson-Negele Distributor. ·· Fewer tune-ups ·· No more overflows ·· More reliable inventories ·· Quick Disconnect Receptacles with optional Field Wiring Connectors

 Open the catalog to page 1
SX-2

LIFE SCIENCES Specifications Specifications PERFORMANCE Level Measurement Ranges: (Factory calibrated at no charge and easily field calibrated within the parameters listed below) Minimum Maximum (URL*) Proof Pressure SX5 Series 0-30 " w.c.** 0-140 " w.c. 10 psig Ranges to accomodate vacuum available on request Loop Resistance: 1550 ohms (max.) at 40 vdc, 750 ohms (max) at 24 vdc Cable Recommended: 2 conductor, stranded, 18-24 AWG, shielded with ground 0.17 - 0.26" Cable Sheath OD for use with field wiring connector. Anderson molded cord set recommended for best EMI and water protection. Receptacle: 5-pin...

 Open the catalog to page 2
SX-3

Order Information LIFE SCIENCES Order Information SPAN RANGE 5 0-30” min to 0-140” max 6 0-140.1” to 0-415” max 7 0-415.1” to 0-830” max 8 0-830.1” to 0-1385” max FITTING 089 Anderson Flush Mount Long - with USP VI gasket (71060-A3, A5, A7, A9) 088 Anderson Flush Mount Short - with USP VI gasket (71060-A4, A6, A8) 091 Cherry Burrell Flush Mount Long (insulated tank) 090 Cherry Burrell Flush Mount Short (uninsulated tank) 092 King Gage Flush Mount Long (1777-3) 093 King Gage Flush Mount Medium (1777-1, -6 Standard) 094 King Gage Flush Mount Short (1777-2 non-insulated) 097 Tank Mate Flush Mount...

 Open the catalog to page 3
SX-4

LIFE SCIENCES "A" 3-7/64" Sensor Fittings and Dimensions "A" 3-7/64" ANDERSON SHELL TYPE ANDERSON SHELL TYPE NON INSULATED NON INSULATED INSULATED INSULATED ANDERSON SHELLANDERSON SHELL"A" TYPE TYPE KING SHELL TYPE "A" KING SHELL TYPE "A" NON INSULATED 2-3/16 NON INSULATED 2-3/16 KING SHELL TYPE STANDARD"A" 6-9/16 STANDARD 6-9/16 LONG KING SHELL TYPE "A" "A" LONGTYPE 8-13/16 NON INSULATED KING SHELL2-3/16 8-13/16 NON INSULATED 2-3/16 NON INSULATED 2-3/16 STANDARD 6-9/16 STANDARD 6-9/16 6-9/16 STANDARD LONG 8-13/16 LONG 8-13/16 ANDERSON SHELL TYPE NON INSULATED NON INSULATEDNON INSULATED 2-3/16...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.