ChlorLog

ChlorLog

ChlorLog

Product catalog summary
Introduction
The ChlorLog Compact Residual Chlorine Data Logger is designed to gather data on residual chlorine concentrations in remote parts of water distribution systems. This device ensures minimum disinfectant levels are maintained throughout the system by providing accurate data over extended periods.

Features
  • Continuous measurement of residual chlorine with options for Free, Combined, or Total Chlorine.
  • LCD display with a range of 0-5.00 ppm and 0.01 ppm resolution.
  • Flowcell with integral 0.25 LPM flow control.
  • Reagent-free amperometric membraned sensor.
  • Operates for 4-6 months on internal replaceable batteries.
  • Programmable data storage interval from 1 second to 12 hours.
  • Capacity for 64,000 data points.
  • Includes free software for data download and graphing, with a USB download cable.

Measurement Specifications
  • Measurement Range: 0.00 – 5.00 ppm.
  • Accuracy: ± 0.02 below 1 ppm, ±0.05 above 1 ppm.
  • Repeatability: ± 0.02 ppm.
  • Zero Stability: ± 0.02 ppm non-cumulative.
  • Data Log Frequency: Programmable from 1 second to 12 hours.
  • Power: Three internal ½ AA Lithium cells.
  • Environmental Rating: Nema 6 (IP-68) Submersible.

Sample Requirements
  • Sample Inlet Pressure: 1-50 PSI (0.1-4 bar).
  • Sample Temperature: 1-50° C.
  • Sample pH: 5-10.
  • Suspended Solids: Sample filtered to less than 100 microns.
  • Sample Flowrate: Fixed 250 ml/min.

Ordering Information
  • 00-1819 A22/62 ChlorLog for Free Chlorine.
  • 00-1820 A22/63 ChlorLog for Combined Chlorine.
  • 00-1821 A22/79 ChlorLog for Total Chlorine.
Each ChlorLog comes with a carrying case, spare membranes, electrolyte, O-ring kit, USB cable, calibration tool, and inlet and outlet tube assemblies with valved quick-disconnect fittings. EasyLog software is available for download from www.lascarelectronics.com.
See more

Catalog excerpts

ChlorLog-1

Gathering data on residual chlorine concentrations in remote parts of water distribution systems is increasingly important. Accurate data over extended periods provides the information needed to insure at least minimum disinfectant levels in all parts of a system. Until now, temporary chlorine residual measurement systems required the installation of standard chlorine monitors with high sample flow requirements, power connections, and separate data recording hardware. ATI has developed the “ChlorLog” to provide a compact, stand-alone measurement and logging system that can be used virtually anywhere. Sensor, electronics, data logger, and flowcell are combined in one assembly. All that’s needed is sample connection. Even sample flow control is included as part of the flow assembly. Compact Residual Chlorine Data Log

 Open the catalog to page 1
ChlorLog-2

CHLORLOG FEATURES: • Continuous measurement of residual chlorine • Sensor options for Free Chlorine, Combined Chlorine, or Total Chlorine. • LCD display range of 0-5.00 ppm with 0.01 ppm resolution. • Flowcell with integral 0.25 LPM flow control (4 GPH) • Sample inlet pressure 1-60 PSI (0.1-4 bar) • Reagent free amperometric membraned sensor. • 4-6 Months operation on internal replaceable batteries. • Programmable data storage interval from 1 second to once every 12 hours. • 64,000 data points maximum. • Free software for data download and graphing. USB download cable supplied with ChlorLog....

 Open the catalog to page 2

All Analytical Technology catalogs and technical brochures

  1. C12-17

    2  Pages

  2. F12/D

    4  Pages

  3. Model Q46H/83

    4  Pages

  4. E12-15 IR

    2  Pages

  5. Metrinet

    4  Pages

  6. Model Q45N

    4  Pages

  7. Q45H/64

    2  Pages

  8. Q45C2

    2  Pages

  9. Q22PX

    2  Pages

  10. Model Q45P/R

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.