PRECISION TEMPERATURE AND VOLTAGE MEASUREMENTS FOR HARSH ENVIRONMENTS The RX1032 instrument combines precision thermocouplebased temperature measurements with a ruggedized enclosure; the RX1000 instrument offers the same precision and resilience for voltage measurements. Both are based on the popular EX1000 series, the most advanced, full-featured data acquisition solutions available on the market today. Both units can offer superior measurement accuracy and repeatability by implementing fully integrated signal conditioning and advanced cold junction compensation (CJC). Military-grade connectors ensure more secure power connections. Thermocouple and voltage sensor inputs are routed through a protected entry to a screw terminal panel to allow for faster and easier connectivity. The rugged construction of the enclosure allows mounting instruments on test pylons, in test cells, or in other harsh environments. • 32-Channel Thermocouple/Voltage Measurement Systems SCALABLE FOR HIGHSPEED SYNCHRONIZED DATA ACQUISITION The RX10XX family supports easy integration and synchronization of multiple devices through industrystandard IEEE-1588 implementation. Multiple instruments can easily be distributed and placed extremely close to the measurement points of interest, reducing the length of analog cable, minimizing errors induced by noisy environments. All measurement data is returned with IEEE1588 timestamp codes; typical accuracies of <200µsec ensure that acquired data is tightly coordinated across the entire object under test. • Advanced CJC Implementation • 1000 Sample/Sec/Channel Scan Rate • Precise Distributed Measurement • Synchronization with IEEE-1588 • Rugged Design • Broad Operating Temperature Range • IP65 Environmental Rating • Ethernet Connectivity via LXI Open • Industry Standard • IVI-COM, IVI-C, LabVIEW® Driver Support APPLICATIONS • Large Scale Engine Test • Harsh Environment • Health Monitoring • HALT / HASS • Rocket Motor Reliability • Wind Tunnel Test COLD JUNCTION COMPENSATION The heart of any truly accurate thermocouple measurement system is the CJC implementation. The RX1032 instrument combines multiple precision thermistors, a significant thermal mass, and careful parts placement to provide worldclass measurement performance.
Open the catalog to page 1PRECISION TEMPERATURE AND VOLTAGE MEASUREMENTS FOR HARSH ENVIRONMENTS General Specifications CHANNELS Channels 32 differential thermocouple or voltage inputs Channel types Thermocouple inputs: j, k, t, e, s, r, b, n (RX1032 only) Sampling rate Temperature resolution Temperature accuracy See thermocouple accuracy table Temperature noise, peak-to-peak PROGRAMMABLE FILTERS Bessel (2 pole) COMMON MODE REJECTION RATIO 4 Hz filter DC: 100db minimum; (50/60)Hz: 140db typical, 120db minimum DC: 100db minimum; (50/60)Hz: 100db typical, 80db minimum Voltage Resolution Voltage Accuracy ± (% of reading...
Open the catalog to page 2PRECISION TEMPERATURE AND VOLTAGE MEASUREMENTS FOR HARSH ENVIRONMENTS RX1032/RX1000 Specifications LXI SPECIFICATIONS Clock specifications Clock oscillator accuracy Synchronization accuracy Reports “synchronized” when < ±200μs of the 1588 master clock Timestamp Accuracy As good as time synchronization down to 50μs IEEE 1588- BASED TRIGGER TIMING Alarm Trigger time accuracy As good as time synchronization down to 50μs Time to trigger delay Receive lan [0-7] event Trigger time accuracy As good as time synchronization down to 50μs Time to trigger delay Future timestamp Past/zero timestamp...
Open the catalog to page 3PRECISION TEMPERATURE AND VOLTAGE MEASUREMENTS FOR HARSH ENVIRONMENTS RX1032 Only Specifications Type E, J, K, N & T Typical Note: Typical is based on 2σ & Max is based on 4σ Conditions • After 1 hour warmup period and temperature stable for 1 hour after any temperature change. • Accuracy does not include accuracy of thermocouple or thermocouple wiring. • Maximum temperature rate of change between any operating temperature is 0.5 ˚C/min. • Guaranteed maximum limits are two times (2x) the typical values. • 7 days, ±5 ˚C from last calibration. • -20 ˚C to +30 ˚C, 1 year from full calibration....
Open the catalog to page 4PRECISION TEMPERATURE AND VOLTAGE MEASUREMENTS FOR HARSH ENVIRONMENTS Ordering Information Models Ruggedized 32-Channel Thermocouple Instrument Ruggedized 32-Channel Voltage Instrument Mounting Angle Bracket with screws Cable Mounting Bracket with screws RX-10XX chassis mounting dimensions (in mm) AMETEK Programmable Power 9250 Brown Deer Road San Diego, CA 92121 +1 858-450-0085 sales.ppd@ametek.com www.powerandtest.com Vibration/shock isolator mounting dimensions (in mm) FIRST TIME
Open the catalog to page 5All AMETEK VTI Instruments catalogs and technical brochures
-
SentinelEX SWITCHING SERIES
10 Pages
-
EX2500A
2 Pages
-
CT-400
5 Pages
-
CT-310A
2 Pages
-
CT-100C
2 Pages
-
SM7000D
2 Pages
-
EMX-7500 SERIES
2 Pages
-
EMX-7000 SERIES
2 Pages
-
RX0224
5 Pages
-
RX0124
8 Pages
-
RX0424
6 Pages
-
SMP1100/1200
1 Pages
-
EX1629-001
7 Pages
-
EX1000 Series
16 Pages
-
EX1200 series
27 Pages
-
EMX-1434
10 Pages