Catalog excerpts
RX0224 24 CHANNEL HIGH PERFORMANCE CHARGE V I B R AT I O N M E A S U R E M E N T S A P P L I CAT I O N S Gas Turbine measurements Vibration Health Monitoring in Harsh Environments OVERVIEW 24 Channel Charge vibration measurements Provides full charge signal conditioning with built in integrator for velocity output. Single ended or differential (isolated) operation 128 Ksa/Sec with 24 bit ADC per channel for high resolution data and wide dynamic range Precise synchronization of measurements in a distributed environment through IEEE-1588 Rugged design; -20°C to +60°C operating range IP66 rated GigE Ethernet connectivity via LXI industry standard SOFTWARE DRIVERS IVI-COM, IVI-C, LabVIEWTM Specifications contained within this document are subject to change without notice T I M E ©2014 VTI Instruments Corporation. All rights reserved.
Open the catalog to page 1OVERVIEW The RX0224 is part of VTI’s RX0XXX family of rugged instrument with built in sensor signal conditioning. All connections and operations are designed for rugged operation by using Mil-Grade connections for all inputs. The RX0224 is capable of: • 24 channels of charge signal conditioning and simultaneous sampling ADC • Ethernet interface for simplified scalability and distributability of measurements • 24-bit A/D converters per channel with selectable sample rate of up to 128Ksa/s Although the RX0224 has a single-ended charge converter, it behaves as a differential charge input that...
Open the catalog to page 2The unit has an IP66 rating, protecting against dust, spills, humidity, and water jets from every direction. All connections are designed for rugged operation through the incorporation of Mil-Grade connections for all inputs. The unit also has an extended operating temperature range of -20°C to +60°C and a compact, light-weight design. Using thermal conduction cooling techniques, the RX0224 is able to disipate heat without the use of a fan. SELF-TEST DIAGNOSTICS The RX0224 has a built in health monitoring capability. This allows for close monitoring of functions, including: • Input Charge...
Open the catalog to page 3Sensor Type Charge, Piezo-Electric Input Type (Diff or Single-Ended) Both Diff. (Isolated) & Single-Ended Input Ground Isolation LO side Isolated when Input set to Differential LO side grounded when Input set to Single-Ended Input Range ±12.5kpC-pk, ±100kpC-pk ±512 inch/sec (ips) peak when Integrator enabled when Range set to 100kpC and using 10pC/G accelerometer Input High-Pass Filter Charge Measurement Accuracy Signal >0.8% FS 1 to 10 Hz: 2% of reading 10 to 10k Hz: 1% of reading Signal <0.8% FS 1 to 10 Hz: 2% of 0.8% FS 10 to 10k Hz: 1% 0.8% FS Velocity Measurement Accuracy Signal >0.8%...
Open the catalog to page 4Environmental Sealing IP66 Rating (Dust & water jet tight) Operating Temperature: -20°C to +60°C Storage Temperature: -55°C to +80°C Cold Start: -54°C Internal Pressure Vibration & Shock Acoustic Noise Endurance: 160 dB for 30 minutes Continuous: 150 dB 50Hz - 10,000 Hz Resistance to Corrosive Fluids Per RR EIR2553 Fuel, Engine Oil, Hydraulic Fluid Total Exposure: 195 rads 0.65 rads/min EMC EN 61326 Class A, Criteria A. Annex A Service Life Power Connector MIL-C-26482 32-Pin Receptacle, Shell Size 18 Ethernet Connector Isolated Microdoft (10-32 thread) Input Connector
Open the catalog to page 5All AMETEK VTI Instruments catalogs and technical brochures
-
SentinelEX SWITCHING SERIES
10 Pages
-
EX2500A
2 Pages
-
CT-400
5 Pages
-
CT-310A
2 Pages
-
CT-100C
2 Pages
-
SM7000D
2 Pages
-
EMX-7500 SERIES
2 Pages
-
EMX-7000 SERIES
2 Pages
-
RX0124
8 Pages
-
RX0424
6 Pages
-
RX1032/RX1000
5 Pages
-
SMP1100/1200
1 Pages
-
EX1629-001
7 Pages
-
EX1000 Series
16 Pages
-
EX1200 series
27 Pages
-
EMX-1434
10 Pages