DTI-1000 and STS-100 reference sensors
10Pages

{{requestButtons}}

Catalog excerpts

DTI-1000 and STS-100 reference sensors - 1

DTI-IOOOo STS-050, -100,-102 Digital Temperature Indicator and STS Reference Sensors \METEK’ SENSORS, TEST & CALIBRATION

Open the catalog to page 1
DTI-1000 and STS-100 reference sensors - 2

What to expectWide Temperature Range Improve your Accuracy The DTI-1000 with a STS-probe is a fully traceable thermometer recommended as the reference instrument to verify the true temperature in any type of temperature calibrator, liquid bath, or dry-block calibrator. DTI-1000: -200 to 750°C STS-050: -50 to 400°C STS-100: -150 to 650°C STS-102: -50 to 155°C Use the DTI-1000 and the STS probes as your working temperature reference in any calibration application or use the set-up directly in custody transfer applications where high accuracy (low uncertainty) means money. DTI-1000 accuracy up...

Open the catalog to page 2
DTI-1000 and STS-100 reference sensors - 3

Intuitive and fast navigation Menu / Enter The MENU/ENTER key includes intelligent prompts that guide the user through setup and operation. This key allows the user to specify measuring units, resolution, sensor On / The ON/OFF key automatically initiates a selftest routine and the display indicates tested Display Display with 2 * 20 characters, which continuously read out the two inputs. parameter results plus the date of the last calibration. Reset / Select The MODE key allows the user to specify The RESET/SELECT key allows the user to enter peakholds and to change settings in the MENU...

Open the catalog to page 3
DTI-1000 and STS-100 reference sensors - 4

...and even more True Ohm Measurement The DTI-1000 has been designed to employ state-of-the-art DC measuring techniques in combination with powerful microprocessor technology. To achieve high accuracy, the measuring principle used by the DTI-1000 is True Ohm Measurement thus eliminating the EMF from cables, sockets, and sensors. True Ohm Measurement is a proven method to achieve accurate compensation for errors induced by thermal effects. The resistance is measured through the 4-wire system at 1 mA, after which the instrument takes a reading without any applied current; this second reading...

Open the catalog to page 4
DTI-1000 and STS-100 reference sensors - 5

*for accuracies in °C and °Fplease see table Battery Life Typically 15 hours Shipping Weight incl. standard accessories and carrying case Termination____Goldplated LEMO/4mm banana test sockets

Open the catalog to page 5
DTI-1000 and STS-100 reference sensors - 6

STS Reference Sensors System accuracy for DTI-1000 & STS STS-050 sensor -45 to 400 °C / -49 to 752 °F . . . . . . ±0.060 °C / 0.108 °F 1,2 ±0.070°C / 0.126 °F 1,3 STS-100 sensor -50 to 250 °C / -58 to 482 °F . . . . . . ±0.030 °C / 0.054 °F 1,2 ±0.045 °C / 0.081°F 1,3 -50 to 320 °C / -58 to 608 °F . . . . . . ±0.040 °C / 0.072 °F 1,2 ±0.050 °C / 0.090 °F 1,3 -50 to 400 °C / -58 to 752 °F . . . . . . ±0.060 °C / 0.108 °F 1,2 ±0.070 °C / 0.126 °F 1,3 -50 to 650 °C / -58 to 1202 °F . . . . . ±0.060 °C / 0.108 °F 1,2 ±0.080 °C / 0.144 °F 1,3 -90 to 650 °C / -130 to 1202 °F . . . . ±0.080 °C /...

Open the catalog to page 6
DTI-1000 and STS-100 reference sensors - 7

Functional Temperature Range STS-050 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-50 to 400°C STS-100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -150 to 650°C STS-102 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-50 to 155°C Accuracy, 12 months use STS-050 Hysteresis3) @0°C/32°F . . . . . . . . . . . . . . . . . . . . . . . 0.01°C / 0.02°F Long Term Stabilty4) @0°C/32°F . . . . . . . . typ. 0.014°C / 0.025°F Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.005°C /...

Open the catalog to page 7
DTI-1000 and STS-100 reference sensors - 8

DTI-1000 indicatorDTI1000A F C ^ Base Model Number Calibration Certificate Traceable Certificate to international standards (Standard) C Carrying Case, aluminium DTI-1000 A temperature indicator with standard traceable certificate and carrying case Standard Delivery DTI-1000 ■ DTI-1000 Indicator ■ USB containing manual and JOFRACAL software ■ AmeTrim to adjust the DTI-1000 ■ RS232 cable, 9 pin adapter ■ 8 batteries ■ Calibration certificate, traceable to int. standards ■ Mains adapter AMETEK SENSORS, TEST & CALIBRATION

Open the catalog to page 8
DTI-1000 and STS-100 reference sensors - 9

Base Model Number STS050 Pt100 reference probe w/ handle, -50 to 400°C / -58 to 752°F Base Model Number STS102 Diameter of the sensor A Overall diameter 4 mm / 0.16 in Shape and length 250 Straight, 250 mm / 9.8 in, incl. cardboard box 350 Straight, 350 mm / 13.8 in, incl. cardboard box Shape and length 030 Short sensor 30 mm / 1.18 in, incl. plastic case Cable length and termination Cable 1 m / 3.3 ft + LEMO connector S Cable length and termination Cable 1 m / 3.3 ft plus LEMO connector G 4mm STS-050 A reference sensor, 250 mm straight sensor, 1m cable plus LEMO connector, Base Model...

Open the catalog to page 9
DTI-1000 and STS-100 reference sensors - 10

Accessories - DTI-1000 124716 .. .4 x 1.5 Volt rechargeable batteries (requires 8 batteries) 124718 ... Charger for rechargeable batteries - 115/230VAC 124944 ... Carrying case 124315 ... Removeable trolley for carrying case 125002... Edgeport Converter with 4 pcs of RS232 ports 124720... Mains adapter 9VDC/800mA - 230VAC/115VAC EN ISO/IEC 17025 Laboratory accreditation AMETEK Sensors, Test & Calibration has two EN ISO/IEC 17025 accredited laboratories that issues accredited certifirates in accordance with international standards. Laboratory accreditation is a reliable indicator of...

Open the catalog to page 10

All Ametek STC catalogs and technical brochures

  1. PTC-125 Cooler

    12 Pages

  2. PTC Series

    13 Pages

  3. CPF

    4 Pages

  4. 30 series

    4 Pages

  5. JofraCloud

    6 Pages

Archived catalogs

  1. HPC40 Series

    7 Pages