932 Multi Gas Analyzer

932 Multi Gas Analyzer

932 Multi Gas Analyzer

Product catalog summary
Key Benefits:
  • Simultaneous measurement of up to five gas species.
  • Compatible with Heated Acid Gas (HAG) probe sample handling system, eliminating the need for a pump or water removal.
  • Compatible with the 9xx suite of sulfur recovery analyzers.
  • Sample return to process with HAG probe option.
  • Minimal sample conditioning required.
  • Dual-beam, dual-wavelength design for enhanced accuracy.
  • Rated for hazardous areas (Class I, Division 1 and ATEX II 2 G).
Applications:
  • Gas plant amine treater inlet (H2S, CO2).
  • Sour water stripper feeds to SRU (H2S, CO2, NH3).
  • Tail gas treater quench tower inlet (H2, SO2).
  • Tail gas treater absorber and quench tower outlets (H2S, COS, CS2, H2; H2S, SO2, H2).
  • Chemical process applications.
Key Markets:
  • Hydrocarbon processing.
  • Refining.
  • Sulfur recovery units.
Technical Overview:

The 932 Multi-Gas Analyzer is a robust UV photometric gas analyzer designed for explosion-proof environments. It measures up to five UV-absorbing gas species, primarily focusing on hydrogen sulfide (H2S) and sulfur dioxide (SO2), with options for carbonyl sulfide (COS) and carbon disulfide (CS2). The analyzer uses a dual-beam design to minimize drift and improve accuracy, with a hollow cathode lamp for stable performance.

For applications requiring non-UV absorbing species detection, the 932 can be equipped with a thermal conductivity detector (TCD) for hydrogen (H2) or infrared sensors for hydrocarbons or carbon dioxide (CO2).

Condensation Solution:

The analyzer offers a heated cell option to prevent sample condensation, ensuring reliable dew point temperature control.

Performance Specifications:
  • Methodology: Multiple-wavelength, high-resolution, non-dispersive UV/VIS.
  • Accuracy: ±1.0% full-scale for standard ranges; ±2.0% for H2S+NH3 applications.
  • Repeatability: Better than ±0.5% of full-scale range.
  • Linearity: Better than ±1% of reading for H2S.
  • Zero drift: Better than 2% of full-scale range over 24 hours.
  • Speed of response: Typically less than 30 seconds to T90.
  • Maximum sample gas temperature: 165°C (329°F).
  • Outputs: Up to four isolated 4-20 mA, loop or self-powered; digital communication via RS485 Modbus port.
  • Power consumption: 500 W maximum.
  • Approvals: CEC, NEC, ATEX, IECEx, GOST compliant.
See more

Catalog excerpts

932 Multi Gas Analyzer-1

AMETEK A PROCESS INSTRUMENTS ^ PRODUCT DATA SHEET 932 Multi-Gas Analyzer Sophisticated ultraviolet (UV) photometer measures up to five UV-absorbing gas species for process monitoring and control The 932 is a rugged, multi-component UV photometric gas analyzer housed in an explosion-proof package designed for a variety of gas monitoring and process control applications. Hydrogen sulfide (H2S) is frequently the primary measured component, along with sulfur dioxide (SO2). The 932 can also be configured to measure carbonyl sulfide (COS) and/or carbon disulfide (CS2), when end users are required to measure these sulfur species for regulatory or process efficiency reasons. COS and CS2 can poison some catalysts used in sulfur recovery units, driving up operating expenses at the plant. The 932 uses a dual beam design, with one detector used as a reference to reduce drift. Span drift is further minimized through the use of a hollow cathode lamp that is extremely stable over temperature changes and/or time. The hallow cathode lamps also have very narrow emission lines, improving linearity when compared to dispersive or non-dispersive broadband light sources. The optical assembly in the 932 supports the simultaneous generation and detection of multiple wavelengths, enabling users to measure multiple analytes of interest, or even different ranges of a singe analyte, with one analyzer. Many process applications require the detection of species which do not absorb in the UV spectrum. The 932 may optionally be incorporated with a thermal conductivity detector (TCD) for the measurement of hydrogen (H2), or infrared sensors for the measurement of hydrocarbons or carbon dioxide (CO2). Condensation solution For those applications where sample condensation may occur, the 932 analyzer system is available with a heated cell option. The fully integrated heated cell ensures reliable dew point temperature control without running the risk of plugging, contaminating, or flooding the analyzer.

 Open the catalog to page 1
932 Multi Gas Analyzer-2

PRODUCT DATA SHEET PERFORMANCE SPECIFICATIONS Methodology AMETEK* PROCESS INSTRUMENTS © 2018, by AMETEK, Inc. All rights reserved. Printed in the U.S.A. F-0184 Rev 9 (0818) One of a family of innovative process analyzer solutions from AMETEK Process Instruments. Specifications subject to change without notice.