1. Catalogs
  2. AMETEK Process Instruments
  3. 919 Hot Wet Gas Analyzer

919 Hot Wet Gas Analyzer

919 Hot Wet Gas Analyzer

919 Hot Wet Gas Analyzer

Product catalog summary
Key Benefits
  • High reliability with reduced maintenance due to no moving parts.
  • No need for water correction factors.
  • Automated zero and span gas calibration.
  • Provides serial interface with plant Distributed Control System (DCS).
  • Four temperature control zones for sample lines, probe, sample conditioning, and oven.
  • Compatibility with 900 ADA and series 9xx analyzers.
Applications
  • Continuous emissions monitoring in sulfur plants, smelters, coal, oil, gas-fired power plants, industrial boilers, and process heaters.
Key Markets
  • Sulfur recovery.
  • Wet scrubbers.
  • Sulfuric acid production.
Technical Overview

The 919 Hot/Wet Gas Analyzer is designed for challenging process and emissions monitoring applications. It uses high-resolution UV technology with a resolution better than 0.02 nm, ensuring measurement stability and low noise performance. The analyzer is fully extractive and heated, simplifying gas concentration calculations without the need for sample drying.

Performance Specifications
  • Methodology: Dual wavelength, high resolution, nondispersive UV.
  • Accuracy: ±1% full scale of standard ranges.
  • Repeatability: Better than 0.5% full scale.
  • Response time: Typically less than 30 seconds to T90.
  • Linearity: Better than 1% of full scale.
  • Sample transport: Air aspiration with a typical flow of 3 to 5 L/min.
  • Temperature control: Independent control of up to four zones.
  • Power: 120 VAC or 240 VAC, 600 W for analyzer only.
  • Communications: Analog and digital interfaces including RS232 and RS422 with Modbus protocol.
  • Physical dimensions: 1117.6 x 1553.6 x 306 mm.
  • Weight: Estimated minimum 160 kg.
Approvals and Certifications
  • NEC/CEC Class I, Division 2, Groups C & D.
  • ATEX II 2 G Ex db eb pxb IIB T3 Gb.
  • IECEx Ex d e px IIB T3 Gb.
  • Russian Ex Proof Certification; 1ExpydIIBT3.
  • GOST: 1ExdIIBT3.
Species Measurable
  • Sulfur dioxide (SO2): 250 ppm to 100%.
  • Nitric oxide (NO): 300 ppm to 100%.
  • Nitrogen dioxide (NO2): 300 ppm to 100%.
  • Hydrogen sulfide (H2S): 125 ppm to 100%.
  • Ammonia (NH3): 500 ppm to 100%.

For more information or to request a quote, visit ametekpi.com.

See more

Catalog excerpts

919 Hot Wet Gas Analyzer-1

AMETEK* PROCESS INSTRUMENTS PRODUCT DATA SHEET 919 Hot/Wet Gas Analyzer Hot-wet ultraviolet (UV) based photometric analyzer The 919 was designed to meet the requirements of the most challenging process and emissions monitoring applications in a cost-effective fashion. It is a complete system with a sample extraction and transport designed to maintain sample integrity. High resolution UV technology Resolution better than 0.02 nm is provided by high intensity line source lamps. These sources emit at a fixed wavelength providing great measurement stability, and emit low total power removing the potential for sample photolysis. The dual beam configuration, combined with the reference measurement, ensures low noise performance with minimal baseline and span drift. UV measurements do not suffer from water (H2O) and carbon dioxide (CO2) interference as these species are transparent in the UV. This greatly simplifies sample handling. Unparalleled linearity The high resolution enables unparalleled linearity over a wide dynamic range which, in turn, leads to simple, robust data analysis. Simple, accurate calculation of gas concentrations The 919 is a fully extractive, heated wet-basis analyzer. The sample cell and all components in contact with the sample are heated above the dew points of all gases in the sample stream. This results in a simpler and more accurate calculation of gas concentrations requiring no corrections for condensed and dissolved components. It also results in a simpler analytical system as there is no need for sample drying. Housing options for the analyzer unit include a cabinet or walk-in shelter built to client specifications. An optional oxygen (O2) sensor can be included. n KEY BENEFITS • High reliability and reduced maintenance -no moving parts • No water correction factors • Automated zero and span gas calibration • Provides serial interface with plant DCS • Four temperature control zones for sample lines, probe, sample conditioning and oven • Fully compatible with 900 ADA and series 9xx analyzers ^applications • Continuous emissions monitoring (CEM) applications in sulfur plants, smelters, coal, oil, gas-fired power plants, industrial boilers and process heaters BS KEY MARKETS • Sulfur recovery • Wet scrubbers • Sulfuric acid

 Open the catalog to page 1
919 Hot Wet Gas Analyzer-2

PRODUCT DATA SHEET PERFORMANCE SPECIFICATIONS Methodology SALES, SERVICE & MANUFACTURING WORLDWIDE SALES AND SERVICE LOCATIONS Canada - Alberta 2876 Sunridge Way NE Calgary AB T1Y 7H9 Tel: +1 403 235 8400 Fax: +1 403 248 3550 AMETEK* PROCESS INSTRUMENTS © 2019, by AMETEK, Inc. All rights reserved. Printed in the U.S.A. F-0178 Rev 6 (0419) One of a family of innovative process analyzer solutions from AMETEK Process Instruments. Specifications subject to change without notice.