1. Catalogs
  2. AMETEK Process Instruments
  3. 5100 and 5100HD Gas Analyzers

5100 and 5100HD Gas Analyzers

5100  and 5100HD Gas Analyzers

5100 and 5100HD Gas Analyzers

Product catalog summary
Key Benefits:
  • Real-time performance monitoring.
  • Low maintenance cost and requirements.
  • Remote monitoring and maintenance access.
  • IP 65/Type 4X enclosure for protection.
  • Fully-integrated sample handling.
Applications:
  • Tariff or custody transfer.
  • Natural gas sweetening.
  • Storage inlet/outlet.
  • Petrochemical production.
Key Markets:
  • Natural gas.
  • Chemicals.
  • Petrochemicals.
  • Refining.
Importance of Moisture Removal:
  • Prevents formation of hydrates and liquid slugs.
  • Maintains BTU value of gas stream.
  • Protects pipelines from corrosion.
  • Reduces bacterial spore activation.
Product Features:
  • Laser-based analyzer for moisture measurement.
  • Instantaneous output during moisture concentration changes.
  • Sealed reference cell for continuous optical verification.
  • Non-contact measurement reduces maintenance.
  • Line lock for constant wavelength verification.
  • Suitable for hazardous environments.
Performance Specifications:
  • Range: 0 to 20 ppmv (other ranges available).
  • Accuracy: ±2% of reading or ±4ppm.
  • Ambient temperature: -20 to 50°C.
  • Relative humidity: 0% to 90%, non-condensing.
  • Sample flow rate: 1 to 10 slpm recommended.
  • Sample cell pressure: 70 to 170 kPa absolute.
  • Response speed: <1 second photometric response.
Outputs and Electrical Requirements:
  • Keypad and display, Fast Ethernet, RS485 serial port.
  • Up to two isolated 4-20 mA analog outputs.
  • Four dry relay contacts.
  • Electrical requirements: 120/240 VAC or 24 VDC.
  • Power requirements: 5100: 105W; 5100HD: 450W.
Physical Specifications:
  • 5100 dimensions: 434.2 x 868.8 x 211.7 mm.
  • 5100HD dimensions: 830 x 674 x 305 mm.
  • Weight: 5100: 25 kg; 5100HD: 60 kg.
  • Enclosure: IP-65, Type 4X.
Approvals and Certifications:
  • Certified to meet multiple ATEX, IECEx, CSA, NEC, and Inmetro standards for hazardous areas.
Contact Information:
  • AMETEK Process Instruments, various global locations.
See more

Catalog excerpts

5100  and 5100HD Gas Analyzers-1

AMETEK PROCESS INSTRUMENTS PRODUCT DATA SHEET 5100 and 5100HD Gas Analyzers Measuring moisture in natural gas and non-absorbing backgrounds Moisture is one of the impurities of most concern found in natural gas. It must be removed to: 1) Prevent the formation of hydrates and liquid slugs that can damage expensive equipment 2) Prevent lowering the British Thermal Unit (BTU) value of the gas stream 3) Protect pipelines from corrosion which occurs when carbon dioxide and hydrogen sulfide combine with water to form carbonic or sulfuric acid 4) Reduce bacterial spore activation In several petrochemical production processes, moisture can also negatively impact final product quality or reduce the operational life of expensive catalysts. The 5100 is a laser-based analyzer designed to measure moisture both in local plants and in remote locations. The analyzer can output results nearly instantaneously following sudden increases or decreases in moisture concentrations during upsets. To further increase the optical performance of the 5100, a sealed reference cell is used for continuous on-line optical system verification and offers high specificity and sensitivity. Non-contact measurement The sample never contacts the key optical components, reducing maintenance. Line lock The wavelength of interest is constantly verified using an integrated reference. Designed for use in difficult environments Can be installed in many environments, including those defined as hazardous. Q KEY BENEFITS • Real-time performance monitoring • Low maintenance cost and requirements • Remote monitoring and maintenance access • IP 65/Type 4X enclosure • Fully-integrated sample handling APPLICATIONS • Tariff or custody transfer • Natural gas sweetening • Storage inlet/outlet • Petrochemical production n KEY MARKETS • Natural gas • Chemicals • Petrochemicals • Refining To find out more or request a quote visit our website ametekpi.com

 Open the catalog to page 1
5100  and 5100HD Gas Analyzers-2

PERFORMANCE SPECIFICATIONS © 2018, by AMETEK, Inc. All rights reserved. Printed in the U.S.A. F-0301 Rev 10 (1018) One of a family of innovative process analyzer solutions from AMETEK Process PROCESS INSTRUMENTS Instruments. Specifications subject to change without notice.

 Open the catalog to page 2

All AMETEK Process Instruments catalogs and technical brochures

  1. CEM O2 TM

    2  Pages

  2. CEM Humox

    2  Pages

  3. WDG-HPIIC

    2  Pages

  4. CG1000

    2  Pages

  5. CMFA-P 2000

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.