1. Catalogs
  2. AMETEK MOCON
  3. OX-TRAN Model 2/10
video corpo

OX-TRAN Model 2/10
1 /2Pages

OX-TRAN Model 2/10

OX-TRAN Model 2/10
1 /2Pages

Catalog excerpts

OX-TRAN Model 2/10-1

OX-TRAN ® Model 2/10 A Cost Effective System for Oxygen Transmission Rate Testing of Flat Films and Finished Packages The OX-TRAN® Model 2/10 is the latest value system from mocon using a standard calibration-dependent, electrochemical (non-coulometric*) oxygen sensor which reliably measures oxygen barrier in a range of 0.1 - 144,000 cc/m 2 - day at precise temperatures and relative humidities. *For instruments with coulometric sensors visit www.mocon.com Features • Standard electrochemical (non-coulometric) oxygen sensor • Includes 2 N.I.S.T. traceable calibration films (which are critical with calibration dependent sensors to assure accuracy of results) • Utilizes N.I.S.T. recommended two pressure method of relative humidity generation. In addition, relative humidity sensors are located at the film to verify precise RH • Oxygen transmission rate sensor is flow and pressure compensated for improved accuracy • User friendly Windows ® based software utilizing WinPerm™ and/or PermNet™ • Equipped with two 50 cm 2 cells • Easily expandable up to 20 cells with one sensor for each two cells • Instrument is ETL listed, conforms to UL Standard 61010-1 is certified to CSA 61010-1 and complies with CE Product Safety EN 61010-1 ASTM F-2622 and CFR21 Part 11 Compliant Over 40 Years of Permeation Exp

 Open the catalog to page 1
OX-TRAN Model 2/10-2

Specifications for the OX-TRAN® Model 2/10 Technical O2 TR Test Range Low Test Range: 0.1 cc/m 2 - day with 100% oxygen and carrier gas flow of 10cc/min Upper Test Range: 144,000 cc/m2 - day with 21% oxygen and carrier gas flow of 100cc/min Sensor Standard Calibration-Dependent Electrochemical Oxygen Sensor Test Temperature Range 10 - 40 ° C Films and Packages (optional package environmental chamber available) Precise RH Control (both sides of film) Dry, 35 - 90% Test Sample Sizes: Films: 50 cm2 Flow Carrier gas flow of 10 cc/min or 10Occ/min Resolution 0.01 cc/m2 - day Repeatability ± 0.05 cc/m2...

 Open the catalog to page 2

All AMETEK MOCON catalogs and technical brochures

  1. AQUATRAN 3/40

    4  Pages

  2. BevAlert®

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.